MRS Shimming: An Important Point Which Should not be Ignored

BACKGROUND: Proton magnetic resonance spectroscopy (MRS) is a well-known device for analyzing the biological fluids metabolically. Obtaining accurate and reliable information via MRS needs a homogeneous magnetic field in order to provide well-defined peaks and uniform water suppression. There are lo...

Descripción completa

Detalles Bibliográficos
Autores principales: Zeinali-Rafsanjani, B., Faghihi, R., Mosleh-Shirazi, M.A., Moghadam, Saeedi-M., Lotfi, M., Jalli, R., Sina, S., Mina, L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Journal of Biomedical Physics and Engineering 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6169119/
https://www.ncbi.nlm.nih.gov/pubmed/30320030
_version_ 1783360463411609600
author Zeinali-Rafsanjani, B.
Faghihi, R.
Mosleh-Shirazi, M.A.
Moghadam, Saeedi-M.
Lotfi, M.
Jalli, R.
Sina, S.
Mina, L.
author_facet Zeinali-Rafsanjani, B.
Faghihi, R.
Mosleh-Shirazi, M.A.
Moghadam, Saeedi-M.
Lotfi, M.
Jalli, R.
Sina, S.
Mina, L.
author_sort Zeinali-Rafsanjani, B.
collection PubMed
description BACKGROUND: Proton magnetic resonance spectroscopy (MRS) is a well-known device for analyzing the biological fluids metabolically. Obtaining accurate and reliable information via MRS needs a homogeneous magnetic field in order to provide well-defined peaks and uniform water suppression. There are lots of reasons which can disturb the magnetic field homogeneity which can be corrected by a process known as shimming. This study is intended to recall the importance of shimming and also the significant role of quality control (QC) in achieving an accurate quantification. MATERIAL AND METHOD: An acrylic cylindrical quality control phantom was designed as an analog of brain MRS test phantoms in order to control the accuracy of the obtained signal of a 1.5 T Siemens MRI system which belonged to one of Shiraz hospitals. The signal of NAA, Cho, Cr, the combination of these metabolites and also the distilled water, which was used in this study, was evaluated using separate phantoms. A QC test was performed using Siemens QC phantom and a standard test phantom. RESULTS: The spectrum of our home- made phantom had a significant difference with the expected spectrum. The results of checking the spectrum of metabolites separately also confirmed that there was a systemic problem that affects all the signals originated from all metabolites and even the pure distilled water. The MRS system could not pass QC tests, and peak broadening was common in all spectra. The complex spectrum of standard test phantom was not produced successfully by the MRS system. DISCUSSION: By a simple check of the water peak characteristics, lots of information can be obtained, one of which is the status of shimming that has a considerable effect on the accuracy of the spectrum. Thus, performing an automatic or manual shimming is not a criterion of the spectrum accuracy, and performing a periodic quality control using a test phantom by a specialist is necessary. CONCLUSION: Briefly, the quality control of MRS and all the other clinical device must be taken seriously. Sometimes QC can be the boundary of a right or a wrong decision for the patient.
format Online
Article
Text
id pubmed-6169119
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Journal of Biomedical Physics and Engineering
record_format MEDLINE/PubMed
spelling pubmed-61691192018-10-12 MRS Shimming: An Important Point Which Should not be Ignored Zeinali-Rafsanjani, B. Faghihi, R. Mosleh-Shirazi, M.A. Moghadam, Saeedi-M. Lotfi, M. Jalli, R. Sina, S. Mina, L. J Biomed Phys Eng Original Article BACKGROUND: Proton magnetic resonance spectroscopy (MRS) is a well-known device for analyzing the biological fluids metabolically. Obtaining accurate and reliable information via MRS needs a homogeneous magnetic field in order to provide well-defined peaks and uniform water suppression. There are lots of reasons which can disturb the magnetic field homogeneity which can be corrected by a process known as shimming. This study is intended to recall the importance of shimming and also the significant role of quality control (QC) in achieving an accurate quantification. MATERIAL AND METHOD: An acrylic cylindrical quality control phantom was designed as an analog of brain MRS test phantoms in order to control the accuracy of the obtained signal of a 1.5 T Siemens MRI system which belonged to one of Shiraz hospitals. The signal of NAA, Cho, Cr, the combination of these metabolites and also the distilled water, which was used in this study, was evaluated using separate phantoms. A QC test was performed using Siemens QC phantom and a standard test phantom. RESULTS: The spectrum of our home- made phantom had a significant difference with the expected spectrum. The results of checking the spectrum of metabolites separately also confirmed that there was a systemic problem that affects all the signals originated from all metabolites and even the pure distilled water. The MRS system could not pass QC tests, and peak broadening was common in all spectra. The complex spectrum of standard test phantom was not produced successfully by the MRS system. DISCUSSION: By a simple check of the water peak characteristics, lots of information can be obtained, one of which is the status of shimming that has a considerable effect on the accuracy of the spectrum. Thus, performing an automatic or manual shimming is not a criterion of the spectrum accuracy, and performing a periodic quality control using a test phantom by a specialist is necessary. CONCLUSION: Briefly, the quality control of MRS and all the other clinical device must be taken seriously. Sometimes QC can be the boundary of a right or a wrong decision for the patient. Journal of Biomedical Physics and Engineering 2018-09-01 /pmc/articles/PMC6169119/ /pubmed/30320030 Text en Copyright: © Journal of Biomedical Physics and Engineering http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Zeinali-Rafsanjani, B.
Faghihi, R.
Mosleh-Shirazi, M.A.
Moghadam, Saeedi-M.
Lotfi, M.
Jalli, R.
Sina, S.
Mina, L.
MRS Shimming: An Important Point Which Should not be Ignored
title MRS Shimming: An Important Point Which Should not be Ignored
title_full MRS Shimming: An Important Point Which Should not be Ignored
title_fullStr MRS Shimming: An Important Point Which Should not be Ignored
title_full_unstemmed MRS Shimming: An Important Point Which Should not be Ignored
title_short MRS Shimming: An Important Point Which Should not be Ignored
title_sort mrs shimming: an important point which should not be ignored
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6169119/
https://www.ncbi.nlm.nih.gov/pubmed/30320030
work_keys_str_mv AT zeinalirafsanjanib mrsshimminganimportantpointwhichshouldnotbeignored
AT faghihir mrsshimminganimportantpointwhichshouldnotbeignored
AT moslehshirazima mrsshimminganimportantpointwhichshouldnotbeignored
AT moghadamsaeedim mrsshimminganimportantpointwhichshouldnotbeignored
AT lotfim mrsshimminganimportantpointwhichshouldnotbeignored
AT jallir mrsshimminganimportantpointwhichshouldnotbeignored
AT sinas mrsshimminganimportantpointwhichshouldnotbeignored
AT minal mrsshimminganimportantpointwhichshouldnotbeignored