Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator

PURPOSE: Monte Carlo simulation was carried out for a 6 MV flattening filter-free (FFF) indigenously developed linear accelerator (linac) using the BEAMnrc user-code of the EGSnrc code system. The model was benchmarked against the measurements. A Gaussian distributed electron beam of kinetic energy...

Descripción completa

Detalles Bibliográficos
Autores principales: Mishra, Bibekananda, Mishra, Subhalaxmi, Selvam, T. Palani, Chavan, S. T., Pethe, S. N.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6172861/
https://www.ncbi.nlm.nih.gov/pubmed/30305773
http://dx.doi.org/10.4103/jmp.JMP_58_18
_version_ 1783361027099852800
author Mishra, Bibekananda
Mishra, Subhalaxmi
Selvam, T. Palani
Chavan, S. T.
Pethe, S. N.
author_facet Mishra, Bibekananda
Mishra, Subhalaxmi
Selvam, T. Palani
Chavan, S. T.
Pethe, S. N.
author_sort Mishra, Bibekananda
collection PubMed
description PURPOSE: Monte Carlo simulation was carried out for a 6 MV flattening filter-free (FFF) indigenously developed linear accelerator (linac) using the BEAMnrc user-code of the EGSnrc code system. The model was benchmarked against the measurements. A Gaussian distributed electron beam of kinetic energy 6.2 MeV with full-width half maximum of 1 mm was used in this study. METHODS: The simulation of indigenously developed linac unit has been carried out by using the Monte Carlo-based BEAMnrc user-code of the EGSnrc code system. Using the simulated model, depth and lateral dose profiles were studied using the DOSXYZnrc user-code. The calculated dose data were compared against the measurements using an RFA dosimertic system made by PTW, Germany (water tank MP3-M and 0.125 cm(3) ion chamber). RESULTS: The BEAMDP code was used to analyze photon fluence spectra, mean energy distribution, and electron contamination fluence spectra. Percentage depth dose (PDD) and beam profiles (along both X and Y directions) were calculated for the field sizes 5 cm × 5 cm - 25 cm × 25 cm. The dose difference between the calculated and measured PDD and profile values were under 1%, except for the penumbra region where the maximum deviation was found to be around 3%. CONCLUSIONS: A Monte Carlo model of indigenous FFF linac (6 MV) has been developed and benchmarked against the measured data.
format Online
Article
Text
id pubmed-6172861
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-61728612018-10-10 Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator Mishra, Bibekananda Mishra, Subhalaxmi Selvam, T. Palani Chavan, S. T. Pethe, S. N. J Med Phys Original Article PURPOSE: Monte Carlo simulation was carried out for a 6 MV flattening filter-free (FFF) indigenously developed linear accelerator (linac) using the BEAMnrc user-code of the EGSnrc code system. The model was benchmarked against the measurements. A Gaussian distributed electron beam of kinetic energy 6.2 MeV with full-width half maximum of 1 mm was used in this study. METHODS: The simulation of indigenously developed linac unit has been carried out by using the Monte Carlo-based BEAMnrc user-code of the EGSnrc code system. Using the simulated model, depth and lateral dose profiles were studied using the DOSXYZnrc user-code. The calculated dose data were compared against the measurements using an RFA dosimertic system made by PTW, Germany (water tank MP3-M and 0.125 cm(3) ion chamber). RESULTS: The BEAMDP code was used to analyze photon fluence spectra, mean energy distribution, and electron contamination fluence spectra. Percentage depth dose (PDD) and beam profiles (along both X and Y directions) were calculated for the field sizes 5 cm × 5 cm - 25 cm × 25 cm. The dose difference between the calculated and measured PDD and profile values were under 1%, except for the penumbra region where the maximum deviation was found to be around 3%. CONCLUSIONS: A Monte Carlo model of indigenous FFF linac (6 MV) has been developed and benchmarked against the measured data. Medknow Publications & Media Pvt Ltd 2018 /pmc/articles/PMC6172861/ /pubmed/30305773 http://dx.doi.org/10.4103/jmp.JMP_58_18 Text en Copyright: © 2018 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Mishra, Bibekananda
Mishra, Subhalaxmi
Selvam, T. Palani
Chavan, S. T.
Pethe, S. N.
Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator
title Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator
title_full Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator
title_fullStr Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator
title_full_unstemmed Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator
title_short Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator
title_sort comparison of measured and monte carlo calculated dose distributions from indigenously developed 6 mv flattening filter free medical linear accelerator
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6172861/
https://www.ncbi.nlm.nih.gov/pubmed/30305773
http://dx.doi.org/10.4103/jmp.JMP_58_18
work_keys_str_mv AT mishrabibekananda comparisonofmeasuredandmontecarlocalculateddosedistributionsfromindigenouslydeveloped6mvflatteningfilterfreemedicallinearaccelerator
AT mishrasubhalaxmi comparisonofmeasuredandmontecarlocalculateddosedistributionsfromindigenouslydeveloped6mvflatteningfilterfreemedicallinearaccelerator
AT selvamtpalani comparisonofmeasuredandmontecarlocalculateddosedistributionsfromindigenouslydeveloped6mvflatteningfilterfreemedicallinearaccelerator
AT chavanst comparisonofmeasuredandmontecarlocalculateddosedistributionsfromindigenouslydeveloped6mvflatteningfilterfreemedicallinearaccelerator
AT pethesn comparisonofmeasuredandmontecarlocalculateddosedistributionsfromindigenouslydeveloped6mvflatteningfilterfreemedicallinearaccelerator