Cargando…

A biomechanical assessment of hydraulic ankle-foot devices with and without micro-processor control during slope ambulation in trans-femoral amputees

Slope ambulation is a challenge for trans-femoral amputees due to a relative lack of knee function. The assessment of prosthetic ankles on slopes is required for supporting the design, optimisation, and selection of prostheses. This study assessed two hydraulic ankle-foot devices (one of the hydraul...

Descripción completa

Detalles Bibliográficos
Autores principales: Bai, Xuefei, Ewins, David, Crocombe, Andrew David, Xu, Wei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6173401/
https://www.ncbi.nlm.nih.gov/pubmed/30289921
http://dx.doi.org/10.1371/journal.pone.0205093
Descripción
Sumario:Slope ambulation is a challenge for trans-femoral amputees due to a relative lack of knee function. The assessment of prosthetic ankles on slopes is required for supporting the design, optimisation, and selection of prostheses. This study assessed two hydraulic ankle-foot devices (one of the hydraulic ankles is controlled by a micro-processor that allows real-time adjustment in ankle resistance and range of motion) used by trans-femoral amputees in ascending and descending a 5-degree slope walking, against a rigid ankle-foot device. Five experienced and active unilateral trans-femoral amputees performed ascending and descending slope tests with their usual prosthetic knee and socket fitted with a rigid ankle-foot, a hydraulic ankle-foot without a micro-processor, and a hydraulic ankle-foot with a micro-processor optimised for ascending and descending slopes. Peak values in hip, knee and ankle joint angles and moments were collected and the normalcy Trend Symmetry Index of the prosthetic ankle moments (as an indication of bio-mimicry) were calculated and assessment. Particular benefits of the hydraulic ankle-foot devices were better bio-mimicry of ankle resistance moment, greater range of motion, and improved passive prosthetic knee stability according to the greater mid-stance external knee extensor moment (especially in descending slope) compared to the rigid design. The micro-processor controlled device demonstrated optimised ankle angle and moment patterns for ascending and descending slope respectively, and was found to potentially further improve the ankle moment bio-minicry and prosthetic knee stability compared to the hydraulic device without a micro-processor. However the difference between the micro-processor controlled device and the one without a micro-processor does not reach a statistically significant level.