Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6174498/ https://www.ncbi.nlm.nih.gov/pubmed/30327706 http://dx.doi.org/10.1002/joa3.12104 |
_version_ | 1783361290237902848 |
---|---|
author | Hori, Yuichi Nakahara, Shiro Nishiyama, Naoki Fukuda, Reiko Toratani, Noritaka Sakai, Yoshihiko Taguchi, Isao |
author_facet | Hori, Yuichi Nakahara, Shiro Nishiyama, Naoki Fukuda, Reiko Toratani, Noritaka Sakai, Yoshihiko Taguchi, Isao |
author_sort | Hori, Yuichi |
collection | PubMed |
description | A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD (INCEPTA VR F161). The myopotentials were clearly eliminated by the difference in the band pass filter (PRIZM; 21‐171 Hz, INCEPTA; 20‐85 Hz), but unfortunately, new noise was documented 4 years later. The utility of the bandpass filter for preventing oversensing of myopotentials was observed, but the limitation of its use for long‐term follow‐up was also indicated. |
format | Online Article Text |
id | pubmed-6174498 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-61744982018-10-16 Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up Hori, Yuichi Nakahara, Shiro Nishiyama, Naoki Fukuda, Reiko Toratani, Noritaka Sakai, Yoshihiko Taguchi, Isao J Arrhythm Case Reports A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD (INCEPTA VR F161). The myopotentials were clearly eliminated by the difference in the band pass filter (PRIZM; 21‐171 Hz, INCEPTA; 20‐85 Hz), but unfortunately, new noise was documented 4 years later. The utility of the bandpass filter for preventing oversensing of myopotentials was observed, but the limitation of its use for long‐term follow‐up was also indicated. John Wiley and Sons Inc. 2018-08-20 /pmc/articles/PMC6174498/ /pubmed/30327706 http://dx.doi.org/10.1002/joa3.12104 Text en © 2018 The Authors. Journal of Arrhythmia published by John Wiley & Sons Australia, Ltd on behalf of the Japanese Heart Rhythm Society. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes. |
spellingShingle | Case Reports Hori, Yuichi Nakahara, Shiro Nishiyama, Naoki Fukuda, Reiko Toratani, Noritaka Sakai, Yoshihiko Taguchi, Isao Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up |
title | Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up |
title_full | Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up |
title_fullStr | Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up |
title_full_unstemmed | Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up |
title_short | Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up |
title_sort | limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up |
topic | Case Reports |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6174498/ https://www.ncbi.nlm.nih.gov/pubmed/30327706 http://dx.doi.org/10.1002/joa3.12104 |
work_keys_str_mv | AT horiyuichi limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup AT nakaharashiro limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup AT nishiyamanaoki limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup AT fukudareiko limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup AT torataninoritaka limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup AT sakaiyoshihiko limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup AT taguchiisao limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup |