Cargando…

Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up

A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD...

Descripción completa

Detalles Bibliográficos
Autores principales: Hori, Yuichi, Nakahara, Shiro, Nishiyama, Naoki, Fukuda, Reiko, Toratani, Noritaka, Sakai, Yoshihiko, Taguchi, Isao
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6174498/
https://www.ncbi.nlm.nih.gov/pubmed/30327706
http://dx.doi.org/10.1002/joa3.12104
_version_ 1783361290237902848
author Hori, Yuichi
Nakahara, Shiro
Nishiyama, Naoki
Fukuda, Reiko
Toratani, Noritaka
Sakai, Yoshihiko
Taguchi, Isao
author_facet Hori, Yuichi
Nakahara, Shiro
Nishiyama, Naoki
Fukuda, Reiko
Toratani, Noritaka
Sakai, Yoshihiko
Taguchi, Isao
author_sort Hori, Yuichi
collection PubMed
description A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD (INCEPTA VR F161). The myopotentials were clearly eliminated by the difference in the band pass filter (PRIZM; 21‐171 Hz, INCEPTA; 20‐85 Hz), but unfortunately, new noise was documented 4 years later. The utility of the bandpass filter for preventing oversensing of myopotentials was observed, but the limitation of its use for long‐term follow‐up was also indicated.
format Online
Article
Text
id pubmed-6174498
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-61744982018-10-16 Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up Hori, Yuichi Nakahara, Shiro Nishiyama, Naoki Fukuda, Reiko Toratani, Noritaka Sakai, Yoshihiko Taguchi, Isao J Arrhythm Case Reports A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD (INCEPTA VR F161). The myopotentials were clearly eliminated by the difference in the band pass filter (PRIZM; 21‐171 Hz, INCEPTA; 20‐85 Hz), but unfortunately, new noise was documented 4 years later. The utility of the bandpass filter for preventing oversensing of myopotentials was observed, but the limitation of its use for long‐term follow‐up was also indicated. John Wiley and Sons Inc. 2018-08-20 /pmc/articles/PMC6174498/ /pubmed/30327706 http://dx.doi.org/10.1002/joa3.12104 Text en © 2018 The Authors. Journal of Arrhythmia published by John Wiley & Sons Australia, Ltd on behalf of the Japanese Heart Rhythm Society. This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.
spellingShingle Case Reports
Hori, Yuichi
Nakahara, Shiro
Nishiyama, Naoki
Fukuda, Reiko
Toratani, Noritaka
Sakai, Yoshihiko
Taguchi, Isao
Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
title Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
title_full Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
title_fullStr Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
title_full_unstemmed Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
title_short Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
title_sort limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
topic Case Reports
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6174498/
https://www.ncbi.nlm.nih.gov/pubmed/30327706
http://dx.doi.org/10.1002/joa3.12104
work_keys_str_mv AT horiyuichi limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup
AT nakaharashiro limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup
AT nishiyamanaoki limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup
AT fukudareiko limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup
AT torataninoritaka limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup
AT sakaiyoshihiko limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup
AT taguchiisao limitationofthebandpassfilterinpreventingoversensingofpectoralmyopotentialsoverthelongtermfollowup