Cargando…

Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis

Angular stability locking plates are commonly used in proximal humeral fractures, especially in old patients with osteoporotic bones. These implants show good results in short and midterm follow-up, but complications are not uncommon. Here we present a case report to describe a case of early failure...

Descripción completa

Detalles Bibliográficos
Autores principales: Logroscino, Giandomenico, Donati, Fabrizio, Saracco, Michela, Pilloni, Luciano, Piconi, Corrado
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6178128/
https://www.ncbi.nlm.nih.gov/pubmed/30310842
http://dx.doi.org/10.1016/j.tcr.2018.09.005
_version_ 1783361909576171520
author Logroscino, Giandomenico
Donati, Fabrizio
Saracco, Michela
Pilloni, Luciano
Piconi, Corrado
author_facet Logroscino, Giandomenico
Donati, Fabrizio
Saracco, Michela
Pilloni, Luciano
Piconi, Corrado
author_sort Logroscino, Giandomenico
collection PubMed
description Angular stability locking plates are commonly used in proximal humeral fractures, especially in old patients with osteoporotic bones. These implants show good results in short and midterm follow-up, but complications are not uncommon. Here we present a case report to describe a case of early failure of a proximal angular stability plate implanted in a 72 years old female with a highly unstable two part proximal humeral fracture. An optical and metallurgic analysis was also conducted to study the modalities of failure. We concluded that anatomic reduction and restoration of the medial cortical support are crucial in order to minimize the mechanical load on the bone–metal interface and to prevent mechanical failure, in particular when the fracture rim drop in the weak point of the plate that was found in the proximal two symmetric holes (Holes E).
format Online
Article
Text
id pubmed-6178128
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-61781282018-10-11 Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis Logroscino, Giandomenico Donati, Fabrizio Saracco, Michela Pilloni, Luciano Piconi, Corrado Trauma Case Rep Article Angular stability locking plates are commonly used in proximal humeral fractures, especially in old patients with osteoporotic bones. These implants show good results in short and midterm follow-up, but complications are not uncommon. Here we present a case report to describe a case of early failure of a proximal angular stability plate implanted in a 72 years old female with a highly unstable two part proximal humeral fracture. An optical and metallurgic analysis was also conducted to study the modalities of failure. We concluded that anatomic reduction and restoration of the medial cortical support are crucial in order to minimize the mechanical load on the bone–metal interface and to prevent mechanical failure, in particular when the fracture rim drop in the weak point of the plate that was found in the proximal two symmetric holes (Holes E). Elsevier 2018-10-03 /pmc/articles/PMC6178128/ /pubmed/30310842 http://dx.doi.org/10.1016/j.tcr.2018.09.005 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Article
Logroscino, Giandomenico
Donati, Fabrizio
Saracco, Michela
Pilloni, Luciano
Piconi, Corrado
Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis
title Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis
title_full Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis
title_fullStr Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis
title_full_unstemmed Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis
title_short Early failure of a locked titanium plate in a proximal humeral fracture: Case report and metallurgic analysis
title_sort early failure of a locked titanium plate in a proximal humeral fracture: case report and metallurgic analysis
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6178128/
https://www.ncbi.nlm.nih.gov/pubmed/30310842
http://dx.doi.org/10.1016/j.tcr.2018.09.005
work_keys_str_mv AT logroscinogiandomenico earlyfailureofalockedtitaniumplateinaproximalhumeralfracturecasereportandmetallurgicanalysis
AT donatifabrizio earlyfailureofalockedtitaniumplateinaproximalhumeralfracturecasereportandmetallurgicanalysis
AT saraccomichela earlyfailureofalockedtitaniumplateinaproximalhumeralfracturecasereportandmetallurgicanalysis
AT pilloniluciano earlyfailureofalockedtitaniumplateinaproximalhumeralfracturecasereportandmetallurgicanalysis
AT piconicorrado earlyfailureofalockedtitaniumplateinaproximalhumeralfracturecasereportandmetallurgicanalysis