Force measurement metrics for simulated elbow arthroscopy training
BACKGROUND: Elbow arthroscopy is a difficult surgical technique. Objective metrics can be used to improve safe and effective training in elbow arthroscopy. Force exerted on the elbow tissue during arthroscopy can be a measure of safe tissue manipulation. The purpose of this study was to determine th...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6185876/ https://www.ncbi.nlm.nih.gov/pubmed/30315425 http://dx.doi.org/10.1186/s40634-018-0157-1 |
_version_ | 1783362758886031360 |
---|---|
author | Hilgersom, Nick F. J. Horeman-Franse, Tim Bleys, Ronald L. A. W. Eygendaal, Denise van den Bekerom, Michel P. J. Tuijthof, Gabriëlle J. M. |
author_facet | Hilgersom, Nick F. J. Horeman-Franse, Tim Bleys, Ronald L. A. W. Eygendaal, Denise van den Bekerom, Michel P. J. Tuijthof, Gabriëlle J. M. |
author_sort | Hilgersom, Nick F. J. |
collection | PubMed |
description | BACKGROUND: Elbow arthroscopy is a difficult surgical technique. Objective metrics can be used to improve safe and effective training in elbow arthroscopy. Force exerted on the elbow tissue during arthroscopy can be a measure of safe tissue manipulation. The purpose of this study was to determine the force magnitude and force direction used by experts during arthroscopic elbow navigation in cadaveric specimens and assess their applicability in elbow arthroscopy training. METHODS: Two cadaveric elbows were mounted on a Force Measurement Table (FMT) that allowed 3-dimensional measurements (x-, y-, and z-plane) of the forces exerted on the elbow. Five experts in elbow arthroscopy performed arthroscopic navigation once in each of two cadaveric elbows, navigating through the posterior, posterolateral and anterior compartment in a standardized fashion with visualization of three to four anatomic landmarks per compartment. The total absolute force (F(abs)) and force direction exerted (α and β) on the elbow during arthroscopy were recorded. α being the angle in the horizontal plane and β being the angle in the vertical plane. The 10th–90th percentiles of the data were used to set threshold levels for training. RESULTS: The median F(abs) was 24 N (19 N – 30 N), 27 N (20 N – 33 N) and 29 N (23 N – 32 N) for the posterior, posterolateral and anterior compartment, respectively. The median α was - 29° (- 55° – 5°), - 23° (- 56° – -1°) and 4° (- 22° – -18°) for the posterior, posterolateral and anterior compartment, respectively. The median β was - 71° (- 80° – -65°), - 76° (- 86° – -69°) and - 75° (- 81° – -71°) for the posterior, posterolateral and anterior compartment, respectively. CONCLUSION: Expert data on force magnitude and force direction exerted on the elbow during arthroscopic navigation in cadaveric specimens were collected. The proposed maximum allowable force of 30 N (smallest 90th percentile of F(abs)) exerted on the elbow tissue, and the 10th–90th percentile range of the force directions (α and β) for each compartment may be used to provide objective feedback during arthroscopic skills training. |
format | Online Article Text |
id | pubmed-6185876 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-61858762018-10-18 Force measurement metrics for simulated elbow arthroscopy training Hilgersom, Nick F. J. Horeman-Franse, Tim Bleys, Ronald L. A. W. Eygendaal, Denise van den Bekerom, Michel P. J. Tuijthof, Gabriëlle J. M. J Exp Orthop Research BACKGROUND: Elbow arthroscopy is a difficult surgical technique. Objective metrics can be used to improve safe and effective training in elbow arthroscopy. Force exerted on the elbow tissue during arthroscopy can be a measure of safe tissue manipulation. The purpose of this study was to determine the force magnitude and force direction used by experts during arthroscopic elbow navigation in cadaveric specimens and assess their applicability in elbow arthroscopy training. METHODS: Two cadaveric elbows were mounted on a Force Measurement Table (FMT) that allowed 3-dimensional measurements (x-, y-, and z-plane) of the forces exerted on the elbow. Five experts in elbow arthroscopy performed arthroscopic navigation once in each of two cadaveric elbows, navigating through the posterior, posterolateral and anterior compartment in a standardized fashion with visualization of three to four anatomic landmarks per compartment. The total absolute force (F(abs)) and force direction exerted (α and β) on the elbow during arthroscopy were recorded. α being the angle in the horizontal plane and β being the angle in the vertical plane. The 10th–90th percentiles of the data were used to set threshold levels for training. RESULTS: The median F(abs) was 24 N (19 N – 30 N), 27 N (20 N – 33 N) and 29 N (23 N – 32 N) for the posterior, posterolateral and anterior compartment, respectively. The median α was - 29° (- 55° – 5°), - 23° (- 56° – -1°) and 4° (- 22° – -18°) for the posterior, posterolateral and anterior compartment, respectively. The median β was - 71° (- 80° – -65°), - 76° (- 86° – -69°) and - 75° (- 81° – -71°) for the posterior, posterolateral and anterior compartment, respectively. CONCLUSION: Expert data on force magnitude and force direction exerted on the elbow during arthroscopic navigation in cadaveric specimens were collected. The proposed maximum allowable force of 30 N (smallest 90th percentile of F(abs)) exerted on the elbow tissue, and the 10th–90th percentile range of the force directions (α and β) for each compartment may be used to provide objective feedback during arthroscopic skills training. Springer Berlin Heidelberg 2018-10-11 /pmc/articles/PMC6185876/ /pubmed/30315425 http://dx.doi.org/10.1186/s40634-018-0157-1 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Research Hilgersom, Nick F. J. Horeman-Franse, Tim Bleys, Ronald L. A. W. Eygendaal, Denise van den Bekerom, Michel P. J. Tuijthof, Gabriëlle J. M. Force measurement metrics for simulated elbow arthroscopy training |
title | Force measurement metrics for simulated elbow arthroscopy training |
title_full | Force measurement metrics for simulated elbow arthroscopy training |
title_fullStr | Force measurement metrics for simulated elbow arthroscopy training |
title_full_unstemmed | Force measurement metrics for simulated elbow arthroscopy training |
title_short | Force measurement metrics for simulated elbow arthroscopy training |
title_sort | force measurement metrics for simulated elbow arthroscopy training |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6185876/ https://www.ncbi.nlm.nih.gov/pubmed/30315425 http://dx.doi.org/10.1186/s40634-018-0157-1 |
work_keys_str_mv | AT hilgersomnickfj forcemeasurementmetricsforsimulatedelbowarthroscopytraining AT horemanfransetim forcemeasurementmetricsforsimulatedelbowarthroscopytraining AT bleysronaldlaw forcemeasurementmetricsforsimulatedelbowarthroscopytraining AT eygendaaldenise forcemeasurementmetricsforsimulatedelbowarthroscopytraining AT vandenbekerommichelpj forcemeasurementmetricsforsimulatedelbowarthroscopytraining AT tuijthofgabriellejm forcemeasurementmetricsforsimulatedelbowarthroscopytraining AT forcemeasurementmetricsforsimulatedelbowarthroscopytraining |