Design of a Compact Wireless Multi-Channel High Area-Efficient Stimulator with Arbitrary Channel Configuration

This paper presents the design of a wireless, implantable, multi-channel, programmable stimulator with arbitrary channel combination. A novel channel management module using a switch array is presented, enabling arbitrary channel configuration with a silicon area reduction of 81%. The chip was fabri...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhang, Yuwei, Luo, Deng, Ou, Ting, Yuan, Zhangyi, Huang, Heng, You, Ling, Yue, Yin, Zhang, Milin, Li, Dongmei, Li, Guolin, Yuan, Kexin, Wang, Zhihua
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2017
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6187733/
https://www.ncbi.nlm.nih.gov/pubmed/30393283
http://dx.doi.org/10.3390/mi9010006
Descripción
Sumario:This paper presents the design of a wireless, implantable, multi-channel, programmable stimulator with arbitrary channel combination. A novel channel management module using a switch array is presented, enabling arbitrary channel configuration with a silicon area reduction of 81%. The chip was fabricated in a 0.18- [Formula: see text] m Taiwan semiconductor manufacturing company (TSMC) high voltage (HV) complementary metal–oxide semiconductor (CMOS) technology. A stimulator system was realized using the proposed integrated circuit (IC). A wireless communication link was established between a specified Android-based graphical user interface (GUI) and the proposed device for control of the stimulation pattern and wireless battery charging. The size of the entire system occupies a volume of only 14 mm × 14 mm × 4 mm (without the battery). Experimental results demonstrated a successful independent configuration between different channels, as well as an arbitrary channel combination, as expected.