The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors
Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT). Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion. The aim of this study was to estimate the impact of deviations in positionin...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6188766/ https://www.ncbi.nlm.nih.gov/pubmed/30364042 http://dx.doi.org/10.1155/2018/5826174 |
_version_ | 1783363240067072000 |
---|---|
author | Lee, Jia-Cheng Chen, Yi-Wei Chuang, Keh-Shih Hsu, Fang-Yuh Chou, Fong-In Hsu, Shih-Ming Yen, Sang-Hue Wu, Yuan-Hung |
author_facet | Lee, Jia-Cheng Chen, Yi-Wei Chuang, Keh-Shih Hsu, Fang-Yuh Chou, Fong-In Hsu, Shih-Ming Yen, Sang-Hue Wu, Yuan-Hung |
author_sort | Lee, Jia-Cheng |
collection | PubMed |
description | Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT). Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion. The aim of this study was to estimate the impact of deviations in positioning on the dose administered by BNCT for brain tumors at the Tsing Hua open-pool reactor (THOR). For these studies, a simulated head model was generated based on computed tomography (CT) images of a patient with a brain tumor. A cylindrical brain tumor 3 cm in diameter and 5 cm in length was modeled at distances of 6.5 cm and 2.5 cm from the posterior scalp of this head model (T(6.5 cm) and T(2.5 cm), respectively). Radiation doses associated with positioning errors were evaluated for each distance, including left and right shifts, superior and inferior shifts, shifts from the central axis of the beam aperture, and outward shifts from the surface of the beam aperture. Rotational and tilting effects were also evaluated. The dose prescription was 20 Gray-equivalent (Gy-Eq) to 80 % of the tumor. The treatment planning system, NCTPlan, was used to perform dose calculations. The average decreases in mean tumor dose for T(6.5 cm) for the 1 cm, 2 cm, and 3 cm lateral shifts composed by left, right, superior, and inferior sides, were approximately 1 %, 6 %, and 11 %, respectively, compared to the dose administered to the initial tumor position. The decreases in mean tumor dose for T(6.5 cm) were approximately 5 %, 11 %, and 15 % for the 1 cm, 2 cm, and 3 cm outward shifts, respectively. For a superficial tumor at T(2.5cm), no significant decrease in average mean tumor dose was observed following lateral shifts of 1 cm. Rotational and tilting up to 15° did not result in significant difference to the tumor dose. Dose differences to the normal tissues as a result of the shifts in positioning were also minimal. Taken together, these data demonstrate that the mean dose administered to tumors at greater depths is potentially more vulnerable to deviations in positioning, and greater shift distances resulted in reduced mean tumor doses at the THOR. Moreover, these data provide an estimation of dose differences that are caused by setup error or intra-fractional motion during BNCT, and these may facilitate more accurate predictions of actual patient dose in future treatments. |
format | Online Article Text |
id | pubmed-6188766 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-61887662018-10-25 The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors Lee, Jia-Cheng Chen, Yi-Wei Chuang, Keh-Shih Hsu, Fang-Yuh Chou, Fong-In Hsu, Shih-Ming Yen, Sang-Hue Wu, Yuan-Hung Biomed Res Int Research Article Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT). Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion. The aim of this study was to estimate the impact of deviations in positioning on the dose administered by BNCT for brain tumors at the Tsing Hua open-pool reactor (THOR). For these studies, a simulated head model was generated based on computed tomography (CT) images of a patient with a brain tumor. A cylindrical brain tumor 3 cm in diameter and 5 cm in length was modeled at distances of 6.5 cm and 2.5 cm from the posterior scalp of this head model (T(6.5 cm) and T(2.5 cm), respectively). Radiation doses associated with positioning errors were evaluated for each distance, including left and right shifts, superior and inferior shifts, shifts from the central axis of the beam aperture, and outward shifts from the surface of the beam aperture. Rotational and tilting effects were also evaluated. The dose prescription was 20 Gray-equivalent (Gy-Eq) to 80 % of the tumor. The treatment planning system, NCTPlan, was used to perform dose calculations. The average decreases in mean tumor dose for T(6.5 cm) for the 1 cm, 2 cm, and 3 cm lateral shifts composed by left, right, superior, and inferior sides, were approximately 1 %, 6 %, and 11 %, respectively, compared to the dose administered to the initial tumor position. The decreases in mean tumor dose for T(6.5 cm) were approximately 5 %, 11 %, and 15 % for the 1 cm, 2 cm, and 3 cm outward shifts, respectively. For a superficial tumor at T(2.5cm), no significant decrease in average mean tumor dose was observed following lateral shifts of 1 cm. Rotational and tilting up to 15° did not result in significant difference to the tumor dose. Dose differences to the normal tissues as a result of the shifts in positioning were also minimal. Taken together, these data demonstrate that the mean dose administered to tumors at greater depths is potentially more vulnerable to deviations in positioning, and greater shift distances resulted in reduced mean tumor doses at the THOR. Moreover, these data provide an estimation of dose differences that are caused by setup error or intra-fractional motion during BNCT, and these may facilitate more accurate predictions of actual patient dose in future treatments. Hindawi 2018-10-01 /pmc/articles/PMC6188766/ /pubmed/30364042 http://dx.doi.org/10.1155/2018/5826174 Text en Copyright © 2018 Jia-Cheng Lee et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Lee, Jia-Cheng Chen, Yi-Wei Chuang, Keh-Shih Hsu, Fang-Yuh Chou, Fong-In Hsu, Shih-Ming Yen, Sang-Hue Wu, Yuan-Hung The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors |
title | The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors |
title_full | The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors |
title_fullStr | The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors |
title_full_unstemmed | The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors |
title_short | The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors |
title_sort | dosimetric impact of shifts in patient positioning during boron neutron capture therapy for brain tumors |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6188766/ https://www.ncbi.nlm.nih.gov/pubmed/30364042 http://dx.doi.org/10.1155/2018/5826174 |
work_keys_str_mv | AT leejiacheng thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT chenyiwei thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT chuangkehshih thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT hsufangyuh thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT choufongin thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT hsushihming thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT yensanghue thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT wuyuanhung thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT leejiacheng dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT chenyiwei dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT chuangkehshih dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT hsufangyuh dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT choufongin dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT hsushihming dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT yensanghue dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors AT wuyuanhung dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors |