The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors

Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT). Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion. The aim of this study was to estimate the impact of deviations in positionin...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Jia-Cheng, Chen, Yi-Wei, Chuang, Keh-Shih, Hsu, Fang-Yuh, Chou, Fong-In, Hsu, Shih-Ming, Yen, Sang-Hue, Wu, Yuan-Hung
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6188766/
https://www.ncbi.nlm.nih.gov/pubmed/30364042
http://dx.doi.org/10.1155/2018/5826174
_version_ 1783363240067072000
author Lee, Jia-Cheng
Chen, Yi-Wei
Chuang, Keh-Shih
Hsu, Fang-Yuh
Chou, Fong-In
Hsu, Shih-Ming
Yen, Sang-Hue
Wu, Yuan-Hung
author_facet Lee, Jia-Cheng
Chen, Yi-Wei
Chuang, Keh-Shih
Hsu, Fang-Yuh
Chou, Fong-In
Hsu, Shih-Ming
Yen, Sang-Hue
Wu, Yuan-Hung
author_sort Lee, Jia-Cheng
collection PubMed
description Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT). Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion. The aim of this study was to estimate the impact of deviations in positioning on the dose administered by BNCT for brain tumors at the Tsing Hua open-pool reactor (THOR). For these studies, a simulated head model was generated based on computed tomography (CT) images of a patient with a brain tumor. A cylindrical brain tumor 3 cm in diameter and 5 cm in length was modeled at distances of 6.5 cm and 2.5 cm from the posterior scalp of this head model (T(6.5 cm) and T(2.5 cm), respectively). Radiation doses associated with positioning errors were evaluated for each distance, including left and right shifts, superior and inferior shifts, shifts from the central axis of the beam aperture, and outward shifts from the surface of the beam aperture. Rotational and tilting effects were also evaluated. The dose prescription was 20 Gray-equivalent (Gy-Eq) to 80 % of the tumor. The treatment planning system, NCTPlan, was used to perform dose calculations. The average decreases in mean tumor dose for T(6.5 cm) for the 1 cm, 2 cm, and 3 cm lateral shifts composed by left, right, superior, and inferior sides, were approximately 1 %, 6 %, and 11 %, respectively, compared to the dose administered to the initial tumor position. The decreases in mean tumor dose for T(6.5 cm) were approximately 5 %, 11 %, and 15 % for the 1 cm, 2 cm, and 3 cm outward shifts, respectively. For a superficial tumor at T(2.5cm), no significant decrease in average mean tumor dose was observed following lateral shifts of 1 cm. Rotational and tilting up to 15° did not result in significant difference to the tumor dose. Dose differences to the normal tissues as a result of the shifts in positioning were also minimal. Taken together, these data demonstrate that the mean dose administered to tumors at greater depths is potentially more vulnerable to deviations in positioning, and greater shift distances resulted in reduced mean tumor doses at the THOR. Moreover, these data provide an estimation of dose differences that are caused by setup error or intra-fractional motion during BNCT, and these may facilitate more accurate predictions of actual patient dose in future treatments.
format Online
Article
Text
id pubmed-6188766
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-61887662018-10-25 The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors Lee, Jia-Cheng Chen, Yi-Wei Chuang, Keh-Shih Hsu, Fang-Yuh Chou, Fong-In Hsu, Shih-Ming Yen, Sang-Hue Wu, Yuan-Hung Biomed Res Int Research Article Unlike conventional photon radiotherapy, sophisticated patient positioning tools are not available for boron neutron capture therapy (BNCT). Thus, BNCT remains vulnerable to setup errors and intra-fractional patient motion. The aim of this study was to estimate the impact of deviations in positioning on the dose administered by BNCT for brain tumors at the Tsing Hua open-pool reactor (THOR). For these studies, a simulated head model was generated based on computed tomography (CT) images of a patient with a brain tumor. A cylindrical brain tumor 3 cm in diameter and 5 cm in length was modeled at distances of 6.5 cm and 2.5 cm from the posterior scalp of this head model (T(6.5 cm) and T(2.5 cm), respectively). Radiation doses associated with positioning errors were evaluated for each distance, including left and right shifts, superior and inferior shifts, shifts from the central axis of the beam aperture, and outward shifts from the surface of the beam aperture. Rotational and tilting effects were also evaluated. The dose prescription was 20 Gray-equivalent (Gy-Eq) to 80 % of the tumor. The treatment planning system, NCTPlan, was used to perform dose calculations. The average decreases in mean tumor dose for T(6.5 cm) for the 1 cm, 2 cm, and 3 cm lateral shifts composed by left, right, superior, and inferior sides, were approximately 1 %, 6 %, and 11 %, respectively, compared to the dose administered to the initial tumor position. The decreases in mean tumor dose for T(6.5 cm) were approximately 5 %, 11 %, and 15 % for the 1 cm, 2 cm, and 3 cm outward shifts, respectively. For a superficial tumor at T(2.5cm), no significant decrease in average mean tumor dose was observed following lateral shifts of 1 cm. Rotational and tilting up to 15° did not result in significant difference to the tumor dose. Dose differences to the normal tissues as a result of the shifts in positioning were also minimal. Taken together, these data demonstrate that the mean dose administered to tumors at greater depths is potentially more vulnerable to deviations in positioning, and greater shift distances resulted in reduced mean tumor doses at the THOR. Moreover, these data provide an estimation of dose differences that are caused by setup error or intra-fractional motion during BNCT, and these may facilitate more accurate predictions of actual patient dose in future treatments. Hindawi 2018-10-01 /pmc/articles/PMC6188766/ /pubmed/30364042 http://dx.doi.org/10.1155/2018/5826174 Text en Copyright © 2018 Jia-Cheng Lee et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Lee, Jia-Cheng
Chen, Yi-Wei
Chuang, Keh-Shih
Hsu, Fang-Yuh
Chou, Fong-In
Hsu, Shih-Ming
Yen, Sang-Hue
Wu, Yuan-Hung
The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors
title The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors
title_full The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors
title_fullStr The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors
title_full_unstemmed The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors
title_short The Dosimetric Impact of Shifts in Patient Positioning during Boron Neutron Capture Therapy for Brain Tumors
title_sort dosimetric impact of shifts in patient positioning during boron neutron capture therapy for brain tumors
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6188766/
https://www.ncbi.nlm.nih.gov/pubmed/30364042
http://dx.doi.org/10.1155/2018/5826174
work_keys_str_mv AT leejiacheng thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT chenyiwei thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT chuangkehshih thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT hsufangyuh thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT choufongin thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT hsushihming thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT yensanghue thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT wuyuanhung thedosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT leejiacheng dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT chenyiwei dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT chuangkehshih dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT hsufangyuh dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT choufongin dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT hsushihming dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT yensanghue dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors
AT wuyuanhung dosimetricimpactofshiftsinpatientpositioningduringboronneutroncapturetherapyforbraintumors