Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams

Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were dig...

Descripción completa

Detalles Bibliográficos
Autores principales: Tagiling, Nashrulhaq, Ab Rashid, Raizulnasuha, Azhan, Siti Nur Amirah, Dollah, Norhayati, Geso, Moshi, Rahman, Wan Nordiana
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6197593/
https://www.ncbi.nlm.nih.gov/pubmed/30364574
http://dx.doi.org/10.1016/j.heliyon.2018.e00864
_version_ 1783364799785074688
author Tagiling, Nashrulhaq
Ab Rashid, Raizulnasuha
Azhan, Siti Nur Amirah
Dollah, Norhayati
Geso, Moshi
Rahman, Wan Nordiana
author_facet Tagiling, Nashrulhaq
Ab Rashid, Raizulnasuha
Azhan, Siti Nur Amirah
Dollah, Norhayati
Geso, Moshi
Rahman, Wan Nordiana
author_sort Tagiling, Nashrulhaq
collection PubMed
description Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were digitised using Epson(®) Expression(®) 10000XL flatbed scanner and image analyses of net optical density (netOD) were conducted using five scanning parameters i.e. film type, resolution, image bit depth, colour to grayscale transformation and image inversion. The results showed that increasing spatial resolution and deepening colour depth did not improve film sensitivity, while grayscale scanning caused sensitivity reduction below than that detected in the Red-channel. It is also evident that invert and colour negative film type selection negated netOD values, hence unsuitable for scanning RCFs. In conclusion, choosing appropriate scanning parameters are important to maintain preciseness and reproducibility in films dosimetry.
format Online
Article
Text
id pubmed-6197593
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-61975932018-10-24 Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams Tagiling, Nashrulhaq Ab Rashid, Raizulnasuha Azhan, Siti Nur Amirah Dollah, Norhayati Geso, Moshi Rahman, Wan Nordiana Heliyon Article Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were digitised using Epson(®) Expression(®) 10000XL flatbed scanner and image analyses of net optical density (netOD) were conducted using five scanning parameters i.e. film type, resolution, image bit depth, colour to grayscale transformation and image inversion. The results showed that increasing spatial resolution and deepening colour depth did not improve film sensitivity, while grayscale scanning caused sensitivity reduction below than that detected in the Red-channel. It is also evident that invert and colour negative film type selection negated netOD values, hence unsuitable for scanning RCFs. In conclusion, choosing appropriate scanning parameters are important to maintain preciseness and reproducibility in films dosimetry. Elsevier 2018-10-17 /pmc/articles/PMC6197593/ /pubmed/30364574 http://dx.doi.org/10.1016/j.heliyon.2018.e00864 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Article
Tagiling, Nashrulhaq
Ab Rashid, Raizulnasuha
Azhan, Siti Nur Amirah
Dollah, Norhayati
Geso, Moshi
Rahman, Wan Nordiana
Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
title Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
title_full Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
title_fullStr Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
title_full_unstemmed Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
title_short Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
title_sort effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6197593/
https://www.ncbi.nlm.nih.gov/pubmed/30364574
http://dx.doi.org/10.1016/j.heliyon.2018.e00864
work_keys_str_mv AT tagilingnashrulhaq effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams
AT abrashidraizulnasuha effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams
AT azhansitinuramirah effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams
AT dollahnorhayati effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams
AT gesomoshi effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams
AT rahmanwannordiana effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams