Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams
Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were dig...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6197593/ https://www.ncbi.nlm.nih.gov/pubmed/30364574 http://dx.doi.org/10.1016/j.heliyon.2018.e00864 |
_version_ | 1783364799785074688 |
---|---|
author | Tagiling, Nashrulhaq Ab Rashid, Raizulnasuha Azhan, Siti Nur Amirah Dollah, Norhayati Geso, Moshi Rahman, Wan Nordiana |
author_facet | Tagiling, Nashrulhaq Ab Rashid, Raizulnasuha Azhan, Siti Nur Amirah Dollah, Norhayati Geso, Moshi Rahman, Wan Nordiana |
author_sort | Tagiling, Nashrulhaq |
collection | PubMed |
description | Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were digitised using Epson(®) Expression(®) 10000XL flatbed scanner and image analyses of net optical density (netOD) were conducted using five scanning parameters i.e. film type, resolution, image bit depth, colour to grayscale transformation and image inversion. The results showed that increasing spatial resolution and deepening colour depth did not improve film sensitivity, while grayscale scanning caused sensitivity reduction below than that detected in the Red-channel. It is also evident that invert and colour negative film type selection negated netOD values, hence unsuitable for scanning RCFs. In conclusion, choosing appropriate scanning parameters are important to maintain preciseness and reproducibility in films dosimetry. |
format | Online Article Text |
id | pubmed-6197593 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-61975932018-10-24 Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams Tagiling, Nashrulhaq Ab Rashid, Raizulnasuha Azhan, Siti Nur Amirah Dollah, Norhayati Geso, Moshi Rahman, Wan Nordiana Heliyon Article Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were digitised using Epson(®) Expression(®) 10000XL flatbed scanner and image analyses of net optical density (netOD) were conducted using five scanning parameters i.e. film type, resolution, image bit depth, colour to grayscale transformation and image inversion. The results showed that increasing spatial resolution and deepening colour depth did not improve film sensitivity, while grayscale scanning caused sensitivity reduction below than that detected in the Red-channel. It is also evident that invert and colour negative film type selection negated netOD values, hence unsuitable for scanning RCFs. In conclusion, choosing appropriate scanning parameters are important to maintain preciseness and reproducibility in films dosimetry. Elsevier 2018-10-17 /pmc/articles/PMC6197593/ /pubmed/30364574 http://dx.doi.org/10.1016/j.heliyon.2018.e00864 Text en © 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). |
spellingShingle | Article Tagiling, Nashrulhaq Ab Rashid, Raizulnasuha Azhan, Siti Nur Amirah Dollah, Norhayati Geso, Moshi Rahman, Wan Nordiana Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams |
title | Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams |
title_full | Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams |
title_fullStr | Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams |
title_full_unstemmed | Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams |
title_short | Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams |
title_sort | effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6197593/ https://www.ncbi.nlm.nih.gov/pubmed/30364574 http://dx.doi.org/10.1016/j.heliyon.2018.e00864 |
work_keys_str_mv | AT tagilingnashrulhaq effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams AT abrashidraizulnasuha effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams AT azhansitinuramirah effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams AT dollahnorhayati effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams AT gesomoshi effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams AT rahmanwannordiana effectofscanningparametersondoseresponseofradiochromicfilmsirradiatedwithphotonandelectronbeams |