Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy

Purpose  This study aimed to evaluate the effectiveness of four-dimensional (4D) robust optimization for proton pencil-beam scanning (PBS) treatment of lung tumors. Patients and methods  In seven patients with lung cancer, proton beam therapy was planned using 4D robust optimization over 4D computed...

Descripción completa

Detalles Bibliográficos
Autores principales: Cummings, David, Tang, Shikui, Ichter, William, Wang, Peng, Sturgeon, Jared D, Lee, Andrew K, Chang, Chang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Cureus 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6200439/
https://www.ncbi.nlm.nih.gov/pubmed/30402360
http://dx.doi.org/10.7759/cureus.3192
_version_ 1783365325998260224
author Cummings, David
Tang, Shikui
Ichter, William
Wang, Peng
Sturgeon, Jared D
Lee, Andrew K
Chang, Chang
author_facet Cummings, David
Tang, Shikui
Ichter, William
Wang, Peng
Sturgeon, Jared D
Lee, Andrew K
Chang, Chang
author_sort Cummings, David
collection PubMed
description Purpose  This study aimed to evaluate the effectiveness of four-dimensional (4D) robust optimization for proton pencil-beam scanning (PBS) treatment of lung tumors. Patients and methods  In seven patients with lung cancer, proton beam therapy was planned using 4D robust optimization over 4D computed tomography (CT) data sets. The gross target volume (GTV) was contoured based on individual breathing phases, and a 5-mm expansion was used to generate the clinical target volume (CTV) for each phase. The 4D optimization was conducted directly on the 4D CT data set. The robust optimization settings included a CT Hounsfield unit (HU) uncertainty of 4% and a setup uncertainty of 5 mm to obtain the CTV. Additional target dose objectives such as those for the internal target volume (ITV) as well as the organ-at-risk (OAR) dose requirements were placed on the average CT. For comparison, three-dimensional (3D) robust optimization was also performed on the average CT. An additional verification 4D CT was performed to verify plan robustness against inter-fractional variations. Results  Target coverages were generally higher for 4D optimized plans. The difference was most pronounced for ITV V70Gy when evaluating individual breathing phases. The 4D optimized plans were shown to be able to maintain the ITV coverage at full prescription, while 3D optimized plans could not. More importantly, this difference in ITV V70Gy between the 4D and 3D optimized plans was also consistently observed when evaluating the verification 4D CT, indicating that the 4D optimized plans were more robust against inter-fractional variations. Less difference was seen between the 4D and 3D optimized plans in the lungs criteria: V5Gy and V20Gy. Conclusion  The proton PBS treatment plans optimized directly on the 4D CT were shown to be more robust when compared to those optimized on a regular 3D CT. Robust 4D optimization can improve the target coverage for the proton PBS lung treatments.
format Online
Article
Text
id pubmed-6200439
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Cureus
record_format MEDLINE/PubMed
spelling pubmed-62004392018-11-06 Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy Cummings, David Tang, Shikui Ichter, William Wang, Peng Sturgeon, Jared D Lee, Andrew K Chang, Chang Cureus Medical Physics Purpose  This study aimed to evaluate the effectiveness of four-dimensional (4D) robust optimization for proton pencil-beam scanning (PBS) treatment of lung tumors. Patients and methods  In seven patients with lung cancer, proton beam therapy was planned using 4D robust optimization over 4D computed tomography (CT) data sets. The gross target volume (GTV) was contoured based on individual breathing phases, and a 5-mm expansion was used to generate the clinical target volume (CTV) for each phase. The 4D optimization was conducted directly on the 4D CT data set. The robust optimization settings included a CT Hounsfield unit (HU) uncertainty of 4% and a setup uncertainty of 5 mm to obtain the CTV. Additional target dose objectives such as those for the internal target volume (ITV) as well as the organ-at-risk (OAR) dose requirements were placed on the average CT. For comparison, three-dimensional (3D) robust optimization was also performed on the average CT. An additional verification 4D CT was performed to verify plan robustness against inter-fractional variations. Results  Target coverages were generally higher for 4D optimized plans. The difference was most pronounced for ITV V70Gy when evaluating individual breathing phases. The 4D optimized plans were shown to be able to maintain the ITV coverage at full prescription, while 3D optimized plans could not. More importantly, this difference in ITV V70Gy between the 4D and 3D optimized plans was also consistently observed when evaluating the verification 4D CT, indicating that the 4D optimized plans were more robust against inter-fractional variations. Less difference was seen between the 4D and 3D optimized plans in the lungs criteria: V5Gy and V20Gy. Conclusion  The proton PBS treatment plans optimized directly on the 4D CT were shown to be more robust when compared to those optimized on a regular 3D CT. Robust 4D optimization can improve the target coverage for the proton PBS lung treatments. Cureus 2018-08-23 /pmc/articles/PMC6200439/ /pubmed/30402360 http://dx.doi.org/10.7759/cureus.3192 Text en Copyright © 2018, Cummings et al. http://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Medical Physics
Cummings, David
Tang, Shikui
Ichter, William
Wang, Peng
Sturgeon, Jared D
Lee, Andrew K
Chang, Chang
Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy
title Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy
title_full Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy
title_fullStr Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy
title_full_unstemmed Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy
title_short Four-dimensional Plan Optimization for the Treatment of Lung Tumors Using Pencil-beam Scanning Proton Radiotherapy
title_sort four-dimensional plan optimization for the treatment of lung tumors using pencil-beam scanning proton radiotherapy
topic Medical Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6200439/
https://www.ncbi.nlm.nih.gov/pubmed/30402360
http://dx.doi.org/10.7759/cureus.3192
work_keys_str_mv AT cummingsdavid fourdimensionalplanoptimizationforthetreatmentoflungtumorsusingpencilbeamscanningprotonradiotherapy
AT tangshikui fourdimensionalplanoptimizationforthetreatmentoflungtumorsusingpencilbeamscanningprotonradiotherapy
AT ichterwilliam fourdimensionalplanoptimizationforthetreatmentoflungtumorsusingpencilbeamscanningprotonradiotherapy
AT wangpeng fourdimensionalplanoptimizationforthetreatmentoflungtumorsusingpencilbeamscanningprotonradiotherapy
AT sturgeonjaredd fourdimensionalplanoptimizationforthetreatmentoflungtumorsusingpencilbeamscanningprotonradiotherapy
AT leeandrewk fourdimensionalplanoptimizationforthetreatmentoflungtumorsusingpencilbeamscanningprotonradiotherapy
AT changchang fourdimensionalplanoptimizationforthetreatmentoflungtumorsusingpencilbeamscanningprotonradiotherapy