Cargando…
Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population?
OBJECTIVE: The purposes of this study were to evaluate size-specific dose estimate (SSDE) of low-dose CT (LDCT) in the Korean Lung Cancer Screening (K-LUCAS) project and to determine whether CT protocols from Western countries are appropriate for lung cancer screening in Korea. MATERIALS AND METHODS...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society of Radiology
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6201978/ https://www.ncbi.nlm.nih.gov/pubmed/30386149 http://dx.doi.org/10.3348/kjr.2018.19.6.1179 |
_version_ | 1783365613559742464 |
---|---|
author | Kim, Eun Young Kim, Tae Jung Goo, Jin Mo Kim, Hyae Young Lee, Ji Won Lee, Soojung Lim, Jun-tae Kim, Yeol |
author_facet | Kim, Eun Young Kim, Tae Jung Goo, Jin Mo Kim, Hyae Young Lee, Ji Won Lee, Soojung Lim, Jun-tae Kim, Yeol |
author_sort | Kim, Eun Young |
collection | PubMed |
description | OBJECTIVE: The purposes of this study were to evaluate size-specific dose estimate (SSDE) of low-dose CT (LDCT) in the Korean Lung Cancer Screening (K-LUCAS) project and to determine whether CT protocols from Western countries are appropriate for lung cancer screening in Korea. MATERIALS AND METHODS: For participants (n = 256, four institutions) of K-LUCAS pilot study, volume CT dose index (CTDI(vol)) using a 32-cm diameter reference phantom was compared with SSDE, which was recalculated from CTDI(vol) using size-dependent conversion factor (f-size) based on the body size, as described in the American Association of Physicists in Medicine Report 204. This comparison was subsequently assessed by body mass index (BMI) levels (underweight/normal vs. overweight/obese), and automatic exposure control (AEC) adaptation (yes/no). RESULTS: Size-specific dose estimate was higher than CTDI(vol) (2.22 ± 0.75 mGy vs. 1.67 ± 0.60 mGy, p < 0.001), since the f-size was larger than 1.0 for all participants. The ratio of SSDE to CTDI(vol) was higher in lower BMI groups; 1.26, 1.37, 1.43, and 1.53 in the obese (n = 103), overweight (n = 70), normal (n = 75), and underweight (n = 4), respectively. The ratio of SSDE to CTDI(vol) was greater in standard-sized participants than in large-sized participants independent of AEC adaptation; with AEC, SSDE/CTDI(vol) in large- vs. standard-sized participants: 1.30 ± 0.08 vs. 1.44 ± 0.08 (p < 0.001) and without AEC, 1.32 ± 0.08 vs. 1.42 ± 0.06 (p < 0.001). CONCLUSION: Volume CT dose index based on a reference phantom underestimates radiation exposure of LDCT in standard-sized Korean participants. The optimal radiation dose limit needs to be verified for standard-sized Korean participants. |
format | Online Article Text |
id | pubmed-6201978 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | The Korean Society of Radiology |
record_format | MEDLINE/PubMed |
spelling | pubmed-62019782018-11-01 Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? Kim, Eun Young Kim, Tae Jung Goo, Jin Mo Kim, Hyae Young Lee, Ji Won Lee, Soojung Lim, Jun-tae Kim, Yeol Korean J Radiol Thoracic Imaging OBJECTIVE: The purposes of this study were to evaluate size-specific dose estimate (SSDE) of low-dose CT (LDCT) in the Korean Lung Cancer Screening (K-LUCAS) project and to determine whether CT protocols from Western countries are appropriate for lung cancer screening in Korea. MATERIALS AND METHODS: For participants (n = 256, four institutions) of K-LUCAS pilot study, volume CT dose index (CTDI(vol)) using a 32-cm diameter reference phantom was compared with SSDE, which was recalculated from CTDI(vol) using size-dependent conversion factor (f-size) based on the body size, as described in the American Association of Physicists in Medicine Report 204. This comparison was subsequently assessed by body mass index (BMI) levels (underweight/normal vs. overweight/obese), and automatic exposure control (AEC) adaptation (yes/no). RESULTS: Size-specific dose estimate was higher than CTDI(vol) (2.22 ± 0.75 mGy vs. 1.67 ± 0.60 mGy, p < 0.001), since the f-size was larger than 1.0 for all participants. The ratio of SSDE to CTDI(vol) was higher in lower BMI groups; 1.26, 1.37, 1.43, and 1.53 in the obese (n = 103), overweight (n = 70), normal (n = 75), and underweight (n = 4), respectively. The ratio of SSDE to CTDI(vol) was greater in standard-sized participants than in large-sized participants independent of AEC adaptation; with AEC, SSDE/CTDI(vol) in large- vs. standard-sized participants: 1.30 ± 0.08 vs. 1.44 ± 0.08 (p < 0.001) and without AEC, 1.32 ± 0.08 vs. 1.42 ± 0.06 (p < 0.001). CONCLUSION: Volume CT dose index based on a reference phantom underestimates radiation exposure of LDCT in standard-sized Korean participants. The optimal radiation dose limit needs to be verified for standard-sized Korean participants. The Korean Society of Radiology 2018 2018-10-18 /pmc/articles/PMC6201978/ /pubmed/30386149 http://dx.doi.org/10.3348/kjr.2018.19.6.1179 Text en Copyright © 2018 The Korean Society of Radiology http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Thoracic Imaging Kim, Eun Young Kim, Tae Jung Goo, Jin Mo Kim, Hyae Young Lee, Ji Won Lee, Soojung Lim, Jun-tae Kim, Yeol Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? |
title | Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? |
title_full | Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? |
title_fullStr | Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? |
title_full_unstemmed | Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? |
title_short | Size-Specific Dose Estimation In the Korean Lung Cancer Screening Project: Does a 32-cm Diameter Phantom Represent a Standard-Sized Patient in Korean Population? |
title_sort | size-specific dose estimation in the korean lung cancer screening project: does a 32-cm diameter phantom represent a standard-sized patient in korean population? |
topic | Thoracic Imaging |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6201978/ https://www.ncbi.nlm.nih.gov/pubmed/30386149 http://dx.doi.org/10.3348/kjr.2018.19.6.1179 |
work_keys_str_mv | AT kimeunyoung sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation AT kimtaejung sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation AT goojinmo sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation AT kimhyaeyoung sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation AT leejiwon sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation AT leesoojung sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation AT limjuntae sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation AT kimyeol sizespecificdoseestimationinthekoreanlungcancerscreeningprojectdoesa32cmdiameterphantomrepresentastandardsizedpatientinkoreanpopulation |