The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses
ISO 14243 is the governing standard for wear testing of knee prostheses, but there is controversy over the correct direction of anterior-posterior (AP) displacement and loading and the correct direction of tibial rotation (TR) angles and torque. This study aimed to analyze how altering the direction...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6205607/ https://www.ncbi.nlm.nih.gov/pubmed/30372460 http://dx.doi.org/10.1371/journal.pone.0206496 |
_version_ | 1783366212259938304 |
---|---|
author | Wang, Xiao-Hong Zhang, Wei Song, Da-Yong Li, Hui Dong, Xiang Zhang, Min Zhao, Feng Jin, Zhong-Min Cheng, Cheng-Kung |
author_facet | Wang, Xiao-Hong Zhang, Wei Song, Da-Yong Li, Hui Dong, Xiang Zhang, Min Zhao, Feng Jin, Zhong-Min Cheng, Cheng-Kung |
author_sort | Wang, Xiao-Hong |
collection | PubMed |
description | ISO 14243 is the governing standard for wear testing of knee prostheses, but there is controversy over the correct direction of anterior-posterior (AP) displacement and loading and the correct direction of tibial rotation (TR) angles and torque. This study aimed to analyze how altering the direction of AP and TR affected wear on the tibial insert. Modifications to the conditions specified in ISO 14243–1 and ISO 14243–3 were also proposed. As such, five loading conditions were applied to FEA models of a knee prosthesis: (1) Modified ISO 14243–3 with positive AP displacement and TR angle, (2) ISO 14243–3:2004 with negative AP displacement and positive TR angle, (3) ISO 14243–3:2014 with positive AP displacement and negative TR angle, (4) Modified ISO 14243–1 with positive AP load and TR torque, and (5) ISO 14243–1:2009 with negative AP load and positive TR torque. This study found that changing the input directions for AP and TR according to ISO 14243–1 and 14243–3 had an influence on the wear rate and wear contours on the tibial insert model. However, the extent of wear varies depending on the design features of the tibial insert and shape of the input curves. For displacement control according to ISO 14243–3, changing the direction of AP displacement had a marked influence on the wear rate (272.77%), but changing the direction of TR angle had a much lower impact (2.17%). For load control according to ISO 14243–1, reversing the AP load (ISO 14243–1:2009) only increased the wear rate by 6.73% in comparison to the modified ISO 14243–1 conditions. The clinical relevance of this study is that the results demonstrate that tibial wear is affected by the direction of application of AP and TR. Incorrect application of the loading conditions during the design stage may lead to an ineffective preclinical evaluation and could subsequently influence implant longevity in clinical use. |
format | Online Article Text |
id | pubmed-6205607 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-62056072018-11-19 The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses Wang, Xiao-Hong Zhang, Wei Song, Da-Yong Li, Hui Dong, Xiang Zhang, Min Zhao, Feng Jin, Zhong-Min Cheng, Cheng-Kung PLoS One Research Article ISO 14243 is the governing standard for wear testing of knee prostheses, but there is controversy over the correct direction of anterior-posterior (AP) displacement and loading and the correct direction of tibial rotation (TR) angles and torque. This study aimed to analyze how altering the direction of AP and TR affected wear on the tibial insert. Modifications to the conditions specified in ISO 14243–1 and ISO 14243–3 were also proposed. As such, five loading conditions were applied to FEA models of a knee prosthesis: (1) Modified ISO 14243–3 with positive AP displacement and TR angle, (2) ISO 14243–3:2004 with negative AP displacement and positive TR angle, (3) ISO 14243–3:2014 with positive AP displacement and negative TR angle, (4) Modified ISO 14243–1 with positive AP load and TR torque, and (5) ISO 14243–1:2009 with negative AP load and positive TR torque. This study found that changing the input directions for AP and TR according to ISO 14243–1 and 14243–3 had an influence on the wear rate and wear contours on the tibial insert model. However, the extent of wear varies depending on the design features of the tibial insert and shape of the input curves. For displacement control according to ISO 14243–3, changing the direction of AP displacement had a marked influence on the wear rate (272.77%), but changing the direction of TR angle had a much lower impact (2.17%). For load control according to ISO 14243–1, reversing the AP load (ISO 14243–1:2009) only increased the wear rate by 6.73% in comparison to the modified ISO 14243–1 conditions. The clinical relevance of this study is that the results demonstrate that tibial wear is affected by the direction of application of AP and TR. Incorrect application of the loading conditions during the design stage may lead to an ineffective preclinical evaluation and could subsequently influence implant longevity in clinical use. Public Library of Science 2018-10-29 /pmc/articles/PMC6205607/ /pubmed/30372460 http://dx.doi.org/10.1371/journal.pone.0206496 Text en © 2018 Wang et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Wang, Xiao-Hong Zhang, Wei Song, Da-Yong Li, Hui Dong, Xiang Zhang, Min Zhao, Feng Jin, Zhong-Min Cheng, Cheng-Kung The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses |
title | The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses |
title_full | The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses |
title_fullStr | The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses |
title_full_unstemmed | The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses |
title_short | The impact of variations in input directions according to ISO 14243 on wearing of knee prostheses |
title_sort | impact of variations in input directions according to iso 14243 on wearing of knee prostheses |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6205607/ https://www.ncbi.nlm.nih.gov/pubmed/30372460 http://dx.doi.org/10.1371/journal.pone.0206496 |
work_keys_str_mv | AT wangxiaohong theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT zhangwei theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT songdayong theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT lihui theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT dongxiang theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT zhangmin theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT zhaofeng theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT jinzhongmin theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT chengchengkung theimpactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT wangxiaohong impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT zhangwei impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT songdayong impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT lihui impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT dongxiang impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT zhangmin impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT zhaofeng impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT jinzhongmin impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses AT chengchengkung impactofvariationsininputdirectionsaccordingtoiso14243onwearingofkneeprostheses |