Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses

BACKGROUND: The purpose of this observational case series study was to investigate the role of 25 MHz ultrasound biomicroscopy (UBM) in detecting subluxated lenses and compare it with 50 MHz UBM. METHODS: 45 patients (49 eyes) with suspected subluxation of the lens and 20 normal volunteers (40 eyes)...

Descripción completa

Detalles Bibliográficos
Autores principales: Shi, Mingyu, Ma, Liwei, Zhang, Jinsong, Yan, Qichang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6207873/
https://www.ncbi.nlm.nih.gov/pubmed/30416825
http://dx.doi.org/10.1155/2018/3760280
_version_ 1783366605537804288
author Shi, Mingyu
Ma, Liwei
Zhang, Jinsong
Yan, Qichang
author_facet Shi, Mingyu
Ma, Liwei
Zhang, Jinsong
Yan, Qichang
author_sort Shi, Mingyu
collection PubMed
description BACKGROUND: The purpose of this observational case series study was to investigate the role of 25 MHz ultrasound biomicroscopy (UBM) in detecting subluxated lenses and compare it with 50 MHz UBM. METHODS: 45 patients (49 eyes) with suspected subluxation of the lens and 20 normal volunteers (40 eyes) were included. Different cross-sectional images of the lens position were captured in axial and longitudinal scanning modes using 25 and 50 MHz UBM. The main outcome measurements included the linear distance between the lens equator and ciliary process, the difference value (D-value) between the same cross section of the above bilateral linear distance in the normal and the subluxated subjects, the diagnostic accuracy, and the testing times obtained with 25 and 50 MHz UBM. RESULTS: The position of the lens on axial sections could be clearly shown by using 25 MHz UBM. The D-value of the subluxated eyes was 1-2 mm longer than that of the normal ones. There was a statistically significant difference between 25 and 50 MHz UBM in showing subluxation of the lens, the testing time was significantly faster (2.0 min versus 7.5 min), and the diagnostic accuracy was much higher (98.0% versus 71.4%) with 25 versus 50 MHz UBM. Fifteen eyes with slightly subluxated lens were detected by 25 MHz UBM, and only one eye with slight lens subluxation was detected by 50 MHz UBM. CONCLUSIONS: The results indicated that 25 MHz UBM has a greater diagnostic value than 50 MHz UBM in verifying the status of the lens subluxation and can provide reliable and quantitative imaging evidence for clinical use. This trial is registered with ChiCTR–DOD –15007603.
format Online
Article
Text
id pubmed-6207873
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-62078732018-11-11 Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses Shi, Mingyu Ma, Liwei Zhang, Jinsong Yan, Qichang J Ophthalmol Research Article BACKGROUND: The purpose of this observational case series study was to investigate the role of 25 MHz ultrasound biomicroscopy (UBM) in detecting subluxated lenses and compare it with 50 MHz UBM. METHODS: 45 patients (49 eyes) with suspected subluxation of the lens and 20 normal volunteers (40 eyes) were included. Different cross-sectional images of the lens position were captured in axial and longitudinal scanning modes using 25 and 50 MHz UBM. The main outcome measurements included the linear distance between the lens equator and ciliary process, the difference value (D-value) between the same cross section of the above bilateral linear distance in the normal and the subluxated subjects, the diagnostic accuracy, and the testing times obtained with 25 and 50 MHz UBM. RESULTS: The position of the lens on axial sections could be clearly shown by using 25 MHz UBM. The D-value of the subluxated eyes was 1-2 mm longer than that of the normal ones. There was a statistically significant difference between 25 and 50 MHz UBM in showing subluxation of the lens, the testing time was significantly faster (2.0 min versus 7.5 min), and the diagnostic accuracy was much higher (98.0% versus 71.4%) with 25 versus 50 MHz UBM. Fifteen eyes with slightly subluxated lens were detected by 25 MHz UBM, and only one eye with slight lens subluxation was detected by 50 MHz UBM. CONCLUSIONS: The results indicated that 25 MHz UBM has a greater diagnostic value than 50 MHz UBM in verifying the status of the lens subluxation and can provide reliable and quantitative imaging evidence for clinical use. This trial is registered with ChiCTR–DOD –15007603. Hindawi 2018-10-17 /pmc/articles/PMC6207873/ /pubmed/30416825 http://dx.doi.org/10.1155/2018/3760280 Text en Copyright © 2018 Mingyu Shi et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Shi, Mingyu
Ma, Liwei
Zhang, Jinsong
Yan, Qichang
Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_full Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_fullStr Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_full_unstemmed Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_short Role of 25 MHz Ultrasound Biomicroscopy in the Detection of Subluxated Lenses
title_sort role of 25 mhz ultrasound biomicroscopy in the detection of subluxated lenses
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6207873/
https://www.ncbi.nlm.nih.gov/pubmed/30416825
http://dx.doi.org/10.1155/2018/3760280
work_keys_str_mv AT shimingyu roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses
AT maliwei roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses
AT zhangjinsong roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses
AT yanqichang roleof25mhzultrasoundbiomicroscopyinthedetectionofsubluxatedlenses