Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation

PURPOSE: To compare the agreement and efficiency of a swept source-optical coherence tomography biometer, IOLMaster 700 (IOLM700), and a low-coherence optical reflectometry biometer, LENSTAR LS 900 (LS900), when acquiring biometry measurements during cataract evaluation. METHODS: A retrospective cha...

Descripción completa

Detalles Bibliográficos
Autores principales: Passi, Samuel F, Thompson, Atalie C, Gupta, Preeya K
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6219312/
https://www.ncbi.nlm.nih.gov/pubmed/30464385
http://dx.doi.org/10.2147/OPTH.S182898
_version_ 1783368635807432704
author Passi, Samuel F
Thompson, Atalie C
Gupta, Preeya K
author_facet Passi, Samuel F
Thompson, Atalie C
Gupta, Preeya K
author_sort Passi, Samuel F
collection PubMed
description PURPOSE: To compare the agreement and efficiency of a swept source-optical coherence tomography biometer, IOLMaster 700 (IOLM700), and a low-coherence optical reflectometry biometer, LENSTAR LS 900 (LS900), when acquiring biometry measurements during cataract evaluation. METHODS: A retrospective chart review of biometry measurements that were performed in 64 eyes of 32 patients on the same day using the LS900 and the IOLM700. The total image acquisition time per subject was compared between the two machines using a Wilcoxon rank-sum test. Bland–Altman plots showing the mean difference and 95% limits of agreement were graphed. Intraclass correlation coefficients (ICCs) were calculated for the mean axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and two keratometry mean values (K1 and K2) that were output from each device. RESULTS: The average time to complete biometry measurements in both eyes was significantly shorter for the IOLM700 compared with the LS900 (44.5±12.4 vs 168.8±67.2 seconds, P<0.001). The Bland–Altman analysis and ICCs showed high degrees of agreement for the mean biometry values (ICC: AL 0.9999, ACD 0.9993, LT 0.9571, K1 0.9922, K2 0.9926) generated by the two devices. CONCLUSION: There was a high level of agreement between the mean biometry output measures for IOLM700 and LS900. However, it took ~73% less time on average to acquire the images when using the IOLM700 compared with the LS900. In a busy clinic setting, use of the IOLM700 for biometry measures may save time and prove more efficient.
format Online
Article
Text
id pubmed-6219312
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Dove Medical Press
record_format MEDLINE/PubMed
spelling pubmed-62193122018-11-21 Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation Passi, Samuel F Thompson, Atalie C Gupta, Preeya K Clin Ophthalmol Original Research PURPOSE: To compare the agreement and efficiency of a swept source-optical coherence tomography biometer, IOLMaster 700 (IOLM700), and a low-coherence optical reflectometry biometer, LENSTAR LS 900 (LS900), when acquiring biometry measurements during cataract evaluation. METHODS: A retrospective chart review of biometry measurements that were performed in 64 eyes of 32 patients on the same day using the LS900 and the IOLM700. The total image acquisition time per subject was compared between the two machines using a Wilcoxon rank-sum test. Bland–Altman plots showing the mean difference and 95% limits of agreement were graphed. Intraclass correlation coefficients (ICCs) were calculated for the mean axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and two keratometry mean values (K1 and K2) that were output from each device. RESULTS: The average time to complete biometry measurements in both eyes was significantly shorter for the IOLM700 compared with the LS900 (44.5±12.4 vs 168.8±67.2 seconds, P<0.001). The Bland–Altman analysis and ICCs showed high degrees of agreement for the mean biometry values (ICC: AL 0.9999, ACD 0.9993, LT 0.9571, K1 0.9922, K2 0.9926) generated by the two devices. CONCLUSION: There was a high level of agreement between the mean biometry output measures for IOLM700 and LS900. However, it took ~73% less time on average to acquire the images when using the IOLM700 compared with the LS900. In a busy clinic setting, use of the IOLM700 for biometry measures may save time and prove more efficient. Dove Medical Press 2018-11-01 /pmc/articles/PMC6219312/ /pubmed/30464385 http://dx.doi.org/10.2147/OPTH.S182898 Text en © 2018 Passi et al. This work is published and licensed by Dove Medical Press Limited The full terms of this license are available at https://www.dovepress.com/terms.php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed.
spellingShingle Original Research
Passi, Samuel F
Thompson, Atalie C
Gupta, Preeya K
Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation
title Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation
title_full Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation
title_fullStr Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation
title_full_unstemmed Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation
title_short Comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation
title_sort comparison of agreement and efficiency of a swept source-optical coherence tomography device and an optical low-coherence reflectometry device for biometry measurements during cataract evaluation
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6219312/
https://www.ncbi.nlm.nih.gov/pubmed/30464385
http://dx.doi.org/10.2147/OPTH.S182898
work_keys_str_mv AT passisamuelf comparisonofagreementandefficiencyofasweptsourceopticalcoherencetomographydeviceandanopticallowcoherencereflectometrydeviceforbiometrymeasurementsduringcataractevaluation
AT thompsonataliec comparisonofagreementandefficiencyofasweptsourceopticalcoherencetomographydeviceandanopticallowcoherencereflectometrydeviceforbiometrymeasurementsduringcataractevaluation
AT guptapreeyak comparisonofagreementandefficiencyofasweptsourceopticalcoherencetomographydeviceandanopticallowcoherencereflectometrydeviceforbiometrymeasurementsduringcataractevaluation