Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images

Background: Infiltrations of 18F-fluorodeoxyglucose (FDG) injections affect positron emission tomography/computed tomography (PET/CT) image quality and quantification. A device using scintillation sensors (Lucerno Dynamics, Cary, NC) provides dynamic measurements acquired during FDG uptake to identi...

Descripción completa

Detalles Bibliográficos
Autores principales: Lattanze, Ronald K., Osman, Medhat M., Ryan, Kelley A., Frye, Sarah, Townsend, David W.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6221940/
https://www.ncbi.nlm.nih.gov/pubmed/30443549
http://dx.doi.org/10.3389/fmed.2018.00303
_version_ 1783369111724621824
author Lattanze, Ronald K.
Osman, Medhat M.
Ryan, Kelley A.
Frye, Sarah
Townsend, David W.
author_facet Lattanze, Ronald K.
Osman, Medhat M.
Ryan, Kelley A.
Frye, Sarah
Townsend, David W.
author_sort Lattanze, Ronald K.
collection PubMed
description Background: Infiltrations of 18F-fluorodeoxyglucose (FDG) injections affect positron emission tomography/computed tomography (PET/CT) image quality and quantification. A device using scintillation sensors (Lucerno Dynamics, Cary, NC) provides dynamic measurements acquired during FDG uptake to identify and characterize radioactivity near the injection site prior to patient imaging. Our aim was to compare sensor measurements against dynamic PET image acquisition, our proposed reference in assessing injection quality during the uptake period. Methods: Subjects undergoing routine FDG PET/CT imaging were eligible for this Institutional Review Board approved prospective study. After providing informed consent, subjects had sensors topically placed on their arms. FDG was injected into subjects' veins directly on the PET imaging table. Dynamic images of the injection site were acquired during 45 min of the uptake period. These dynamic image acquisitions and subjects' routine standard static images were evaluated by nuclear medicine physicians for abnormal FDG accumulation near the injection site. Sensor measurements were interpreted independently by Lucerno staff. Dynamic image acquisition interpretation results were compared to the sensor measurement interpretations and to static image interpretations. Results: Twenty-four subjects were consented and enrolled. Data from 21 subjects were gathered. During dynamic image acquisition review, physicians interpreted 4 subjects with no FDG accumulation at the injection site, whereas 17 showed evidence of accumulation. In 10 of the 17 cases that showed FDG accumulation, the FDG presence at the injection site resolved completely during uptake corresponding to venous stasis, the temporary sequestration of blood from circulation. Static image interpretation agreed with dynamic images interpretation in 11/21 (52%) subjects. Sensor measurement interpretations agreed with the dynamic images interpretations in 18/21 (86%) subjects. Conclusions: Sensor measurements can be an effective way to identify and characterize infiltrations and venous stasis. Comparable to an infiltration, venous stasis may produce spurious and clinically meaningful measurement bias and possibly even scan misinterpretation. Since the quality and quantification of PET/CT studies are of clinical importance, sensor measurements acquired during the FDG uptake may prove to be a useful quality control measure to reduce infiltration rates and potentially improve patient care. Registration: Clinicaltrials.gov, Identifier: NCT03041090
format Online
Article
Text
id pubmed-6221940
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-62219402018-11-15 Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images Lattanze, Ronald K. Osman, Medhat M. Ryan, Kelley A. Frye, Sarah Townsend, David W. Front Med (Lausanne) Medicine Background: Infiltrations of 18F-fluorodeoxyglucose (FDG) injections affect positron emission tomography/computed tomography (PET/CT) image quality and quantification. A device using scintillation sensors (Lucerno Dynamics, Cary, NC) provides dynamic measurements acquired during FDG uptake to identify and characterize radioactivity near the injection site prior to patient imaging. Our aim was to compare sensor measurements against dynamic PET image acquisition, our proposed reference in assessing injection quality during the uptake period. Methods: Subjects undergoing routine FDG PET/CT imaging were eligible for this Institutional Review Board approved prospective study. After providing informed consent, subjects had sensors topically placed on their arms. FDG was injected into subjects' veins directly on the PET imaging table. Dynamic images of the injection site were acquired during 45 min of the uptake period. These dynamic image acquisitions and subjects' routine standard static images were evaluated by nuclear medicine physicians for abnormal FDG accumulation near the injection site. Sensor measurements were interpreted independently by Lucerno staff. Dynamic image acquisition interpretation results were compared to the sensor measurement interpretations and to static image interpretations. Results: Twenty-four subjects were consented and enrolled. Data from 21 subjects were gathered. During dynamic image acquisition review, physicians interpreted 4 subjects with no FDG accumulation at the injection site, whereas 17 showed evidence of accumulation. In 10 of the 17 cases that showed FDG accumulation, the FDG presence at the injection site resolved completely during uptake corresponding to venous stasis, the temporary sequestration of blood from circulation. Static image interpretation agreed with dynamic images interpretation in 11/21 (52%) subjects. Sensor measurement interpretations agreed with the dynamic images interpretations in 18/21 (86%) subjects. Conclusions: Sensor measurements can be an effective way to identify and characterize infiltrations and venous stasis. Comparable to an infiltration, venous stasis may produce spurious and clinically meaningful measurement bias and possibly even scan misinterpretation. Since the quality and quantification of PET/CT studies are of clinical importance, sensor measurements acquired during the FDG uptake may prove to be a useful quality control measure to reduce infiltration rates and potentially improve patient care. Registration: Clinicaltrials.gov, Identifier: NCT03041090 Frontiers Media S.A. 2018-11-01 /pmc/articles/PMC6221940/ /pubmed/30443549 http://dx.doi.org/10.3389/fmed.2018.00303 Text en Copyright © 2018 Lattanze, Osman, Ryan, Frye and Townsend. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Medicine
Lattanze, Ronald K.
Osman, Medhat M.
Ryan, Kelley A.
Frye, Sarah
Townsend, David W.
Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images
title Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images
title_full Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images
title_fullStr Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images
title_full_unstemmed Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images
title_short Usefulness of Topically Applied Sensors to Assess the Quality of 18F-FDG Injections and Validation Against Dynamic Positron Emission Tomography (PET) Images
title_sort usefulness of topically applied sensors to assess the quality of 18f-fdg injections and validation against dynamic positron emission tomography (pet) images
topic Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6221940/
https://www.ncbi.nlm.nih.gov/pubmed/30443549
http://dx.doi.org/10.3389/fmed.2018.00303
work_keys_str_mv AT lattanzeronaldk usefulnessoftopicallyappliedsensorstoassessthequalityof18ffdginjectionsandvalidationagainstdynamicpositronemissiontomographypetimages
AT osmanmedhatm usefulnessoftopicallyappliedsensorstoassessthequalityof18ffdginjectionsandvalidationagainstdynamicpositronemissiontomographypetimages
AT ryankelleya usefulnessoftopicallyappliedsensorstoassessthequalityof18ffdginjectionsandvalidationagainstdynamicpositronemissiontomographypetimages
AT fryesarah usefulnessoftopicallyappliedsensorstoassessthequalityof18ffdginjectionsandvalidationagainstdynamicpositronemissiontomographypetimages
AT townsenddavidw usefulnessoftopicallyappliedsensorstoassessthequalityof18ffdginjectionsandvalidationagainstdynamicpositronemissiontomographypetimages