Cargando…
Inter-Day Reliability of Finapres (®) Cardiovascular Measurements During Rest and Exercise
This study evaluated the inter-day test-retest reliability of the Finapres (®) finger pulse pressure measuring device during rest and exercise. Eight male participants visited the laboratory twice for evaluation of the inter-day reliability of the Finapres (®) finger-pulse pressure device to measure...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
© Georg Thieme Verlag KG
2017
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6225956/ https://www.ncbi.nlm.nih.gov/pubmed/30539112 http://dx.doi.org/10.1055/s-0043-122081 |
_version_ | 1783369880675811328 |
---|---|
author | Waldron, Mark David Patterson, Stephen Jeffries, Owen |
author_facet | Waldron, Mark David Patterson, Stephen Jeffries, Owen |
author_sort | Waldron, Mark |
collection | PubMed |
description | This study evaluated the inter-day test-retest reliability of the Finapres (®) finger pulse pressure measuring device during rest and exercise. Eight male participants visited the laboratory twice for evaluation of the inter-day reliability of the Finapres (®) finger-pulse pressure device to measure: heart rate (HR), stroke volume (SV), cardiac output (Q̇) and mean arterial pressure (MAP) at rest, and treadmill walking at 3 km/h on 1% and 5% inclines. There were no systematic biases for any of the variables between days. The coefficient of variation (CV%) and 95% limits of agreement (95% LoA) was smallest for MAP (CV%=1.6–3.2%; LoA total error=4.6–12 mmHg) and HR (CV%=3.2–3.9%; LoA total error=6.8–11.9 b/min), increasing with exercise intensity (gradient). The pattern of error was different for Q̇, with decreasing CV% (4.8–3.8%) and LoA (4.2–5.7 L/min) from rest to 5% gradient, with the larger errors occurring for resting SV (CV=7.4%; LoA total error=21.5 ml). The device measures MAP and HR reliably between days; however, error increases at higher intensities. The measurement of SV is less reliable, probably owing to underlying algorithmic assumptions. |
format | Online Article Text |
id | pubmed-6225956 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2017 |
publisher | © Georg Thieme Verlag KG |
record_format | MEDLINE/PubMed |
spelling | pubmed-62259562018-12-11 Inter-Day Reliability of Finapres (®) Cardiovascular Measurements During Rest and Exercise Waldron, Mark David Patterson, Stephen Jeffries, Owen Sports Med Int Open This study evaluated the inter-day test-retest reliability of the Finapres (®) finger pulse pressure measuring device during rest and exercise. Eight male participants visited the laboratory twice for evaluation of the inter-day reliability of the Finapres (®) finger-pulse pressure device to measure: heart rate (HR), stroke volume (SV), cardiac output (Q̇) and mean arterial pressure (MAP) at rest, and treadmill walking at 3 km/h on 1% and 5% inclines. There were no systematic biases for any of the variables between days. The coefficient of variation (CV%) and 95% limits of agreement (95% LoA) was smallest for MAP (CV%=1.6–3.2%; LoA total error=4.6–12 mmHg) and HR (CV%=3.2–3.9%; LoA total error=6.8–11.9 b/min), increasing with exercise intensity (gradient). The pattern of error was different for Q̇, with decreasing CV% (4.8–3.8%) and LoA (4.2–5.7 L/min) from rest to 5% gradient, with the larger errors occurring for resting SV (CV=7.4%; LoA total error=21.5 ml). The device measures MAP and HR reliably between days; however, error increases at higher intensities. The measurement of SV is less reliable, probably owing to underlying algorithmic assumptions. © Georg Thieme Verlag KG 2017-11-17 /pmc/articles/PMC6225956/ /pubmed/30539112 http://dx.doi.org/10.1055/s-0043-122081 Text en https://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives License, which permits unrestricted reproduction and distribution, for non-commercial purposes only; and use and reproduction, but not distribution, of adapted material for non-commercial purposes only, provided the original work is properly cited. |
spellingShingle | Waldron, Mark David Patterson, Stephen Jeffries, Owen Inter-Day Reliability of Finapres (®) Cardiovascular Measurements During Rest and Exercise |
title |
Inter-Day Reliability of Finapres
(®)
Cardiovascular Measurements During Rest and Exercise
|
title_full |
Inter-Day Reliability of Finapres
(®)
Cardiovascular Measurements During Rest and Exercise
|
title_fullStr |
Inter-Day Reliability of Finapres
(®)
Cardiovascular Measurements During Rest and Exercise
|
title_full_unstemmed |
Inter-Day Reliability of Finapres
(®)
Cardiovascular Measurements During Rest and Exercise
|
title_short |
Inter-Day Reliability of Finapres
(®)
Cardiovascular Measurements During Rest and Exercise
|
title_sort | inter-day reliability of finapres
(®)
cardiovascular measurements during rest and exercise |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6225956/ https://www.ncbi.nlm.nih.gov/pubmed/30539112 http://dx.doi.org/10.1055/s-0043-122081 |
work_keys_str_mv | AT waldronmark interdayreliabilityoffinaprescardiovascularmeasurementsduringrestandexercise AT davidpattersonstephen interdayreliabilityoffinaprescardiovascularmeasurementsduringrestandexercise AT jeffriesowen interdayreliabilityoffinaprescardiovascularmeasurementsduringrestandexercise |