The Polar (®) OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise
PURPOSE: Traditional heart-rate monitoring through the use of electrocardiograms or chest-worn heart rate sensors can be challenging in certain sports or in field settings. New technologies, such as photoplethysmography (PPG), have enabled heart-rate monitoring at alternate sites. However, to date,...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
© Georg Thieme Verlag KG
2018
|
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6225963/ https://www.ncbi.nlm.nih.gov/pubmed/30539120 http://dx.doi.org/10.1055/a-0631-0920 |
_version_ | 1783369882137526272 |
---|---|
author | Schubert, Matthew M. Clark, Amy De La Rosa, Annie B. |
author_facet | Schubert, Matthew M. Clark, Amy De La Rosa, Annie B. |
author_sort | Schubert, Matthew M. |
collection | PubMed |
description | PURPOSE: Traditional heart-rate monitoring through the use of electrocardiograms or chest-worn heart rate sensors can be challenging in certain sports or in field settings. New technologies, such as photoplethysmography (PPG), have enabled heart-rate monitoring at alternate sites. However, to date, the accuracy and validity of various PPG sensors has not been examined in detail. The purpose of the present investigation was to determine the validity of an arm-worn PPG sensor during yoga sequences. METHODS: Fifteen college-aged men and women participated in a ~45 min power vinyasa yoga class. During the class, participants wore Polar (®) H7 chest straps and RCX3 receivers (criterion) and Polar (®) OH1 arm bands on their upper right arm (practical). Mean differences were compared via a paired t -test, heart rate during yoga using a time*device repeated measures ANOVA, and agreement assessed with Bland-Altman analysis. RESULTS: Mean heart rates during yoga were not different (mean difference=0.76, 95% CI: –0.54 to 2.06; p =0.229). Yoga created a main effect of time on heart rate ( p <0.0001), but there was no difference between devices ( p =0.86) or interaction ( p =0.90). Mean bias±95% limits of agreement was 0.76±1.30 bpm, with a typical error of 2.42±1.49 bpm and a coefficient of variation of 1.8±1.5%. CONCLUSIONS: Results of the present investigation revealed that the Polar (®) OH1 is a valid measure of heart rate during moderate-vigorous exercise. Future validation studies should consider other exercise modes and participant characteristics. |
format | Online Article Text |
id | pubmed-6225963 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | © Georg Thieme Verlag KG |
record_format | MEDLINE/PubMed |
spelling | pubmed-62259632018-12-11 The Polar (®) OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise Schubert, Matthew M. Clark, Amy De La Rosa, Annie B. Sports Med Int Open PURPOSE: Traditional heart-rate monitoring through the use of electrocardiograms or chest-worn heart rate sensors can be challenging in certain sports or in field settings. New technologies, such as photoplethysmography (PPG), have enabled heart-rate monitoring at alternate sites. However, to date, the accuracy and validity of various PPG sensors has not been examined in detail. The purpose of the present investigation was to determine the validity of an arm-worn PPG sensor during yoga sequences. METHODS: Fifteen college-aged men and women participated in a ~45 min power vinyasa yoga class. During the class, participants wore Polar (®) H7 chest straps and RCX3 receivers (criterion) and Polar (®) OH1 arm bands on their upper right arm (practical). Mean differences were compared via a paired t -test, heart rate during yoga using a time*device repeated measures ANOVA, and agreement assessed with Bland-Altman analysis. RESULTS: Mean heart rates during yoga were not different (mean difference=0.76, 95% CI: –0.54 to 2.06; p =0.229). Yoga created a main effect of time on heart rate ( p <0.0001), but there was no difference between devices ( p =0.86) or interaction ( p =0.90). Mean bias±95% limits of agreement was 0.76±1.30 bpm, with a typical error of 2.42±1.49 bpm and a coefficient of variation of 1.8±1.5%. CONCLUSIONS: Results of the present investigation revealed that the Polar (®) OH1 is a valid measure of heart rate during moderate-vigorous exercise. Future validation studies should consider other exercise modes and participant characteristics. © Georg Thieme Verlag KG 2018-06-17 /pmc/articles/PMC6225963/ /pubmed/30539120 http://dx.doi.org/10.1055/a-0631-0920 Text en https://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivatives License, which permits unrestricted reproduction and distribution, for non-commercial purposes only; and use and reproduction, but not distribution, of adapted material for non-commercial purposes only, provided the original work is properly cited. |
spellingShingle | Schubert, Matthew M. Clark, Amy De La Rosa, Annie B. The Polar (®) OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise |
title |
The Polar
(®)
OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise
|
title_full |
The Polar
(®)
OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise
|
title_fullStr |
The Polar
(®)
OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise
|
title_full_unstemmed |
The Polar
(®)
OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise
|
title_short |
The Polar
(®)
OH1 Optical Heart Rate Sensor is Valid during Moderate-Vigorous Exercise
|
title_sort | polar
(®)
oh1 optical heart rate sensor is valid during moderate-vigorous exercise |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6225963/ https://www.ncbi.nlm.nih.gov/pubmed/30539120 http://dx.doi.org/10.1055/a-0631-0920 |
work_keys_str_mv | AT schubertmatthewm thepolaroh1opticalheartratesensorisvalidduringmoderatevigorousexercise AT clarkamy thepolaroh1opticalheartratesensorisvalidduringmoderatevigorousexercise AT delarosaannieb thepolaroh1opticalheartratesensorisvalidduringmoderatevigorousexercise AT schubertmatthewm polaroh1opticalheartratesensorisvalidduringmoderatevigorousexercise AT clarkamy polaroh1opticalheartratesensorisvalidduringmoderatevigorousexercise AT delarosaannieb polaroh1opticalheartratesensorisvalidduringmoderatevigorousexercise |