Cargando…
Loop formation by an aortic occlusion balloon catheter during resuscitative endovascular balloon occlusion of the aorta (REBOA)
A 77-year-old man was transferred to our hospital for endoscopically uncontrollable active bleeding from a duodenal ulcer. Soon after his arrival, he became hemodynamically unstable and resuscitative endovascular balloon occlusion of the aorta was performed using a 7-F aortic occlusion balloon cathe...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6226620/ https://www.ncbi.nlm.nih.gov/pubmed/30425770 http://dx.doi.org/10.1016/j.radcr.2018.10.018 |
Sumario: | A 77-year-old man was transferred to our hospital for endoscopically uncontrollable active bleeding from a duodenal ulcer. Soon after his arrival, he became hemodynamically unstable and resuscitative endovascular balloon occlusion of the aorta was performed using a 7-F aortic occlusion balloon catheter (Rescue Balloon; Tokai Medical Products, Aichi, Japan). He became hemodynamically stable and was transferred to the CT room. CT demonstrated that the distal part of the catheter shaft had made a loop in the aorta and the balloon was located at the level of the upper abdomen. We consider the low-profile occlusion balloon catheter to be less rigid than large ones, and care should be taken to prevent balloon migration and catheter shaft bending. |
---|