Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
Measurement of pH in IVF-media using the blood gas analyzer (BGA) requires validation, because IVF-media is outside the intended scope of the BGA. To determine whether the Siemens Rapidpoint 500 BGA is suitable for pH measurements in IVF-media this study will validate the BGA and assess its accuracy...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6231605/ https://www.ncbi.nlm.nih.gov/pubmed/30418977 http://dx.doi.org/10.1371/journal.pone.0206707 |
_version_ | 1783370256632250368 |
---|---|
author | Diaz de Pool, Juan D. N. Van Den Berg, Sjoerd A. A. Pilgram, Gonneke S. K. Ballieux, Bartholomeus E. P. B. Van Der Westerlaken, Lucia A. J. |
author_facet | Diaz de Pool, Juan D. N. Van Den Berg, Sjoerd A. A. Pilgram, Gonneke S. K. Ballieux, Bartholomeus E. P. B. Van Der Westerlaken, Lucia A. J. |
author_sort | Diaz de Pool, Juan D. N. |
collection | PubMed |
description | Measurement of pH in IVF-media using the blood gas analyzer (BGA) requires validation, because IVF-media is outside the intended scope of the BGA. To determine whether the Siemens Rapidpoint 500 BGA is suitable for pH measurements in IVF-media this study will validate the BGA and assess its accuracy. In this method comparison study, the pH of over three hundred IVF-media samples was measured with the BGA and a pH electrode (Hanna pH checker). The precision of both the BGA and the pH electrode were excellent (coefficient variation <1.4%). However, the closeness of agreement between measured values of both devices were not equivalent to each other in the tested IVF-media, showing 15% to 85% accordance between devices. The pH measured with the blood gas analyzer was also significantly higher in the tested media, compared to that measured by the pH electrode. One of the tested media did not reach its target pH when it was measured with the BGA, even at 9% CO(2). The results show that the validated blood gas analyzer produces excellent results in terms of precision but not in terms of accuracy. Inaccurate measurement may lead to misinterpretation of results and consequently to suboptimal culture conditions. Therefore, each laboratory is encouraged to perform a validation of their BGA. |
format | Online Article Text |
id | pubmed-6231605 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-62316052018-11-19 Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions Diaz de Pool, Juan D. N. Van Den Berg, Sjoerd A. A. Pilgram, Gonneke S. K. Ballieux, Bartholomeus E. P. B. Van Der Westerlaken, Lucia A. J. PLoS One Research Article Measurement of pH in IVF-media using the blood gas analyzer (BGA) requires validation, because IVF-media is outside the intended scope of the BGA. To determine whether the Siemens Rapidpoint 500 BGA is suitable for pH measurements in IVF-media this study will validate the BGA and assess its accuracy. In this method comparison study, the pH of over three hundred IVF-media samples was measured with the BGA and a pH electrode (Hanna pH checker). The precision of both the BGA and the pH electrode were excellent (coefficient variation <1.4%). However, the closeness of agreement between measured values of both devices were not equivalent to each other in the tested IVF-media, showing 15% to 85% accordance between devices. The pH measured with the blood gas analyzer was also significantly higher in the tested media, compared to that measured by the pH electrode. One of the tested media did not reach its target pH when it was measured with the BGA, even at 9% CO(2). The results show that the validated blood gas analyzer produces excellent results in terms of precision but not in terms of accuracy. Inaccurate measurement may lead to misinterpretation of results and consequently to suboptimal culture conditions. Therefore, each laboratory is encouraged to perform a validation of their BGA. Public Library of Science 2018-11-12 /pmc/articles/PMC6231605/ /pubmed/30418977 http://dx.doi.org/10.1371/journal.pone.0206707 Text en © 2018 Diaz de Pool et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Diaz de Pool, Juan D. N. Van Den Berg, Sjoerd A. A. Pilgram, Gonneke S. K. Ballieux, Bartholomeus E. P. B. Van Der Westerlaken, Lucia A. J. Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions |
title | Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions |
title_full | Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions |
title_fullStr | Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions |
title_full_unstemmed | Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions |
title_short | Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions |
title_sort | validation of the blood gas analyzer for ph measurements in ivf culture medium: prevent suboptimal culture conditions |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6231605/ https://www.ncbi.nlm.nih.gov/pubmed/30418977 http://dx.doi.org/10.1371/journal.pone.0206707 |
work_keys_str_mv | AT diazdepooljuandn validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions AT vandenbergsjoerdaa validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions AT pilgramgonnekesk validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions AT ballieuxbartholomeusepb validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions AT vanderwesterlakenluciaaj validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions |