Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions

Measurement of pH in IVF-media using the blood gas analyzer (BGA) requires validation, because IVF-media is outside the intended scope of the BGA. To determine whether the Siemens Rapidpoint 500 BGA is suitable for pH measurements in IVF-media this study will validate the BGA and assess its accuracy...

Descripción completa

Detalles Bibliográficos
Autores principales: Diaz de Pool, Juan D. N., Van Den Berg, Sjoerd A. A., Pilgram, Gonneke S. K., Ballieux, Bartholomeus E. P. B., Van Der Westerlaken, Lucia A. J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6231605/
https://www.ncbi.nlm.nih.gov/pubmed/30418977
http://dx.doi.org/10.1371/journal.pone.0206707
_version_ 1783370256632250368
author Diaz de Pool, Juan D. N.
Van Den Berg, Sjoerd A. A.
Pilgram, Gonneke S. K.
Ballieux, Bartholomeus E. P. B.
Van Der Westerlaken, Lucia A. J.
author_facet Diaz de Pool, Juan D. N.
Van Den Berg, Sjoerd A. A.
Pilgram, Gonneke S. K.
Ballieux, Bartholomeus E. P. B.
Van Der Westerlaken, Lucia A. J.
author_sort Diaz de Pool, Juan D. N.
collection PubMed
description Measurement of pH in IVF-media using the blood gas analyzer (BGA) requires validation, because IVF-media is outside the intended scope of the BGA. To determine whether the Siemens Rapidpoint 500 BGA is suitable for pH measurements in IVF-media this study will validate the BGA and assess its accuracy. In this method comparison study, the pH of over three hundred IVF-media samples was measured with the BGA and a pH electrode (Hanna pH checker). The precision of both the BGA and the pH electrode were excellent (coefficient variation <1.4%). However, the closeness of agreement between measured values of both devices were not equivalent to each other in the tested IVF-media, showing 15% to 85% accordance between devices. The pH measured with the blood gas analyzer was also significantly higher in the tested media, compared to that measured by the pH electrode. One of the tested media did not reach its target pH when it was measured with the BGA, even at 9% CO(2). The results show that the validated blood gas analyzer produces excellent results in terms of precision but not in terms of accuracy. Inaccurate measurement may lead to misinterpretation of results and consequently to suboptimal culture conditions. Therefore, each laboratory is encouraged to perform a validation of their BGA.
format Online
Article
Text
id pubmed-6231605
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-62316052018-11-19 Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions Diaz de Pool, Juan D. N. Van Den Berg, Sjoerd A. A. Pilgram, Gonneke S. K. Ballieux, Bartholomeus E. P. B. Van Der Westerlaken, Lucia A. J. PLoS One Research Article Measurement of pH in IVF-media using the blood gas analyzer (BGA) requires validation, because IVF-media is outside the intended scope of the BGA. To determine whether the Siemens Rapidpoint 500 BGA is suitable for pH measurements in IVF-media this study will validate the BGA and assess its accuracy. In this method comparison study, the pH of over three hundred IVF-media samples was measured with the BGA and a pH electrode (Hanna pH checker). The precision of both the BGA and the pH electrode were excellent (coefficient variation <1.4%). However, the closeness of agreement between measured values of both devices were not equivalent to each other in the tested IVF-media, showing 15% to 85% accordance between devices. The pH measured with the blood gas analyzer was also significantly higher in the tested media, compared to that measured by the pH electrode. One of the tested media did not reach its target pH when it was measured with the BGA, even at 9% CO(2). The results show that the validated blood gas analyzer produces excellent results in terms of precision but not in terms of accuracy. Inaccurate measurement may lead to misinterpretation of results and consequently to suboptimal culture conditions. Therefore, each laboratory is encouraged to perform a validation of their BGA. Public Library of Science 2018-11-12 /pmc/articles/PMC6231605/ /pubmed/30418977 http://dx.doi.org/10.1371/journal.pone.0206707 Text en © 2018 Diaz de Pool et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Diaz de Pool, Juan D. N.
Van Den Berg, Sjoerd A. A.
Pilgram, Gonneke S. K.
Ballieux, Bartholomeus E. P. B.
Van Der Westerlaken, Lucia A. J.
Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
title Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
title_full Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
title_fullStr Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
title_full_unstemmed Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
title_short Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions
title_sort validation of the blood gas analyzer for ph measurements in ivf culture medium: prevent suboptimal culture conditions
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6231605/
https://www.ncbi.nlm.nih.gov/pubmed/30418977
http://dx.doi.org/10.1371/journal.pone.0206707
work_keys_str_mv AT diazdepooljuandn validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions
AT vandenbergsjoerdaa validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions
AT pilgramgonnekesk validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions
AT ballieuxbartholomeusepb validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions
AT vanderwesterlakenluciaaj validationofthebloodgasanalyzerforphmeasurementsinivfculturemediumpreventsuboptimalcultureconditions