Cargando…

A clinical 3D/4D CBCT‐based treatment dose monitoring system

To monitor delivered dose and trigger plan adaptation when deviation becomes unacceptable, a clinical treatment dose (Tx‐Dose) reconstruction system based on three‐dimensional (3D)/four‐dimensional (4D)‐cone beam computed tomograpy (CBCT) images was developed and evaluated on various treatment sites...

Descripción completa

Detalles Bibliográficos
Autores principales: Qin, An, Gersten, David, Liang, Jian, Liu, Qiang, Grill, Inga, Guerrero, Thomas, Stevens, Craig, Yan, Di
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6236849/
https://www.ncbi.nlm.nih.gov/pubmed/30306710
http://dx.doi.org/10.1002/acm2.12474
_version_ 1783371094028189696
author Qin, An
Gersten, David
Liang, Jian
Liu, Qiang
Grill, Inga
Guerrero, Thomas
Stevens, Craig
Yan, Di
author_facet Qin, An
Gersten, David
Liang, Jian
Liu, Qiang
Grill, Inga
Guerrero, Thomas
Stevens, Craig
Yan, Di
author_sort Qin, An
collection PubMed
description To monitor delivered dose and trigger plan adaptation when deviation becomes unacceptable, a clinical treatment dose (Tx‐Dose) reconstruction system based on three‐dimensional (3D)/four‐dimensional (4D)‐cone beam computed tomograpy (CBCT) images was developed and evaluated on various treatment sites, particularly for lung cancer patient treated by stereotactic body radiation therapy (SBRT). This system integrates with our treatment planning system (TPS), Linacs recording and verification system (R&V), and CBCT imaging system, consisting of three modules: Treatment Schedule Monitoring module (TSM), pseudo‐CT Generating module (PCG), and Treatment Dose Reconstruction/evaluation module (TDR). TSM watches the treatment progress in the R&V system and triggers the PCG module when new CBCT is available. PCG retrieves the CBCTs and performs planning CT to CBCT deformable registration (DIR) to generate pseudo‐CT. The 4D‐CBCT images are taken for target localization and correction in lung cancer patient before treatment. To take full advantage of the valuable information carried by 4D‐CBCT, a novel phase‐matching DIR scheme was developed to generate 4D pseudo‐CT images for 4D dose reconstruction. Finally, TDR module creates TPS scripts to automate Tx‐Dose calculation on the pseudo‐CT images. Both initial quantitative commissioning and patient‐specific qualitative quality assurance of the DIR tool were utilized to ensure the DIR quality. The treatment doses of ten patients (six SBRT‐lung, two head and neck (HN), one breast and one prostate cancer patients) were retrospectively constructed and evaluated. The target registration error (mean ± STD: 1.05 ± 1.13 mm) of the DIR tool is comparable to the interobserver uncertainty (0.88 ± 1.31 mm) evaluated by a publically available lung‐landmarks dataset. For lung SBRT patients, the D(99) of the final cumulative Tx‐Dose of GTV is 93.8 ± 5.5% (83.7–100.1%) of the originally planned D(99). CTV D(99) decreases by 3% and mean ipsilateral parotid dose increases by 11.5% for one of the two HN patients. In conclusion, we have demonstrated the feasibility and effectiveness of a treatment dose verification system in our clinical setting.
format Online
Article
Text
id pubmed-6236849
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-62368492018-11-20 A clinical 3D/4D CBCT‐based treatment dose monitoring system Qin, An Gersten, David Liang, Jian Liu, Qiang Grill, Inga Guerrero, Thomas Stevens, Craig Yan, Di J Appl Clin Med Phys Radiation Oncology Physics To monitor delivered dose and trigger plan adaptation when deviation becomes unacceptable, a clinical treatment dose (Tx‐Dose) reconstruction system based on three‐dimensional (3D)/four‐dimensional (4D)‐cone beam computed tomograpy (CBCT) images was developed and evaluated on various treatment sites, particularly for lung cancer patient treated by stereotactic body radiation therapy (SBRT). This system integrates with our treatment planning system (TPS), Linacs recording and verification system (R&V), and CBCT imaging system, consisting of three modules: Treatment Schedule Monitoring module (TSM), pseudo‐CT Generating module (PCG), and Treatment Dose Reconstruction/evaluation module (TDR). TSM watches the treatment progress in the R&V system and triggers the PCG module when new CBCT is available. PCG retrieves the CBCTs and performs planning CT to CBCT deformable registration (DIR) to generate pseudo‐CT. The 4D‐CBCT images are taken for target localization and correction in lung cancer patient before treatment. To take full advantage of the valuable information carried by 4D‐CBCT, a novel phase‐matching DIR scheme was developed to generate 4D pseudo‐CT images for 4D dose reconstruction. Finally, TDR module creates TPS scripts to automate Tx‐Dose calculation on the pseudo‐CT images. Both initial quantitative commissioning and patient‐specific qualitative quality assurance of the DIR tool were utilized to ensure the DIR quality. The treatment doses of ten patients (six SBRT‐lung, two head and neck (HN), one breast and one prostate cancer patients) were retrospectively constructed and evaluated. The target registration error (mean ± STD: 1.05 ± 1.13 mm) of the DIR tool is comparable to the interobserver uncertainty (0.88 ± 1.31 mm) evaluated by a publically available lung‐landmarks dataset. For lung SBRT patients, the D(99) of the final cumulative Tx‐Dose of GTV is 93.8 ± 5.5% (83.7–100.1%) of the originally planned D(99). CTV D(99) decreases by 3% and mean ipsilateral parotid dose increases by 11.5% for one of the two HN patients. In conclusion, we have demonstrated the feasibility and effectiveness of a treatment dose verification system in our clinical setting. John Wiley and Sons Inc. 2018-10-10 /pmc/articles/PMC6236849/ /pubmed/30306710 http://dx.doi.org/10.1002/acm2.12474 Text en © 2018 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
Qin, An
Gersten, David
Liang, Jian
Liu, Qiang
Grill, Inga
Guerrero, Thomas
Stevens, Craig
Yan, Di
A clinical 3D/4D CBCT‐based treatment dose monitoring system
title A clinical 3D/4D CBCT‐based treatment dose monitoring system
title_full A clinical 3D/4D CBCT‐based treatment dose monitoring system
title_fullStr A clinical 3D/4D CBCT‐based treatment dose monitoring system
title_full_unstemmed A clinical 3D/4D CBCT‐based treatment dose monitoring system
title_short A clinical 3D/4D CBCT‐based treatment dose monitoring system
title_sort clinical 3d/4d cbct‐based treatment dose monitoring system
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6236849/
https://www.ncbi.nlm.nih.gov/pubmed/30306710
http://dx.doi.org/10.1002/acm2.12474
work_keys_str_mv AT qinan aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT gerstendavid aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT liangjian aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT liuqiang aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT grillinga aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT guerrerothomas aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT stevenscraig aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT yandi aclinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT qinan clinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT gerstendavid clinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT liangjian clinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT liuqiang clinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT grillinga clinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT guerrerothomas clinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT stevenscraig clinical3d4dcbctbasedtreatmentdosemonitoringsystem
AT yandi clinical3d4dcbctbasedtreatmentdosemonitoringsystem