Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions

PURPOSE/OBJECTIVE: Reproducible patient positioning remains one of the major challenges in modern radiation therapy. Recently, optical surface scanners have been introduced into clinical practice in addition to well-established positioning systems, such as room laser and skin marks. The aim of this...

Descripción completa

Detalles Bibliográficos
Autores principales: Carl, G, Reitz, D, Schönecker, S, Pazos, M, Freislederer, P, Reiner, M, Alongi, F, Niyazi, M, Ganswindt, U, Belka, C, Corradini, S
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6243634/
https://www.ncbi.nlm.nih.gov/pubmed/30453842
http://dx.doi.org/10.1177/1533033818806002
_version_ 1783371989322301440
author Carl, G
Reitz, D
Schönecker, S
Pazos, M
Freislederer, P
Reiner, M
Alongi, F
Niyazi, M
Ganswindt, U
Belka, C
Corradini, S
author_facet Carl, G
Reitz, D
Schönecker, S
Pazos, M
Freislederer, P
Reiner, M
Alongi, F
Niyazi, M
Ganswindt, U
Belka, C
Corradini, S
author_sort Carl, G
collection PubMed
description PURPOSE/OBJECTIVE: Reproducible patient positioning remains one of the major challenges in modern radiation therapy. Recently, optical surface scanners have been introduced into clinical practice in addition to well-established positioning systems, such as room laser and skin marks. The aim of this prospective study was to evaluate setup errors of the optical surface scanner Catalyst HD (C-RAD AB) in different anatomic regions. MATERIAL/METHODS: Between October 2016 and June 2017 a total of 1902 treatment sessions in 110 patients were evaluated. The workflow of this study included conventional setup procedures using laser-based positioning with skin marks and an additional registration of the 3-dimensional (3D) deviations detected by the Catalyst system. The deviations of the surface-based method were then compared to the corrections of cone beam computed tomography alignment which was considered as gold standard. A practical Catalyst setup error was calculated between the translational deviations of the surface scanner and the laser positioning. Two one-sided t tests for equivalence were used for statistical analysis. RESULTS: Data analysis revealed total deviations of 0.09 mm ± 2.03 mm for the lateral axis, 0.07 mm ± 3.21 mm for the longitudinal axis, and 0.44 mm ± 3.08 mm vertical axis for the Catalyst system, compared to −0.06 ± 3.54 mm lateral, 0.53 ± 3.47 mm longitudinal, and 0.19 ± 3.49 mm vertical for the laser positioning compared to cone beam computed tomography. The lowest positional deviations were found in the cranial region, and larger deviations occurred in the thoracic and abdominal sites. A statistical comparison using 2 one-sided t tests showed a general concordance of the 2 methods (P ≤ 0.036), excluding the vertical direction of the abdominal region (P = 0.198). CONCLUSION: The optical surface scanner Catalyst HD is a reliable and feasible patient positioning system without any additional radiation exposure. From the head to the thoracic and abdominal region, a decrease in accuracy was observed within a comparable range for Catalyst and laser-assisted positioning.
format Online
Article
Text
id pubmed-6243634
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-62436342018-11-26 Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions Carl, G Reitz, D Schönecker, S Pazos, M Freislederer, P Reiner, M Alongi, F Niyazi, M Ganswindt, U Belka, C Corradini, S Technol Cancer Res Treat Original Article PURPOSE/OBJECTIVE: Reproducible patient positioning remains one of the major challenges in modern radiation therapy. Recently, optical surface scanners have been introduced into clinical practice in addition to well-established positioning systems, such as room laser and skin marks. The aim of this prospective study was to evaluate setup errors of the optical surface scanner Catalyst HD (C-RAD AB) in different anatomic regions. MATERIAL/METHODS: Between October 2016 and June 2017 a total of 1902 treatment sessions in 110 patients were evaluated. The workflow of this study included conventional setup procedures using laser-based positioning with skin marks and an additional registration of the 3-dimensional (3D) deviations detected by the Catalyst system. The deviations of the surface-based method were then compared to the corrections of cone beam computed tomography alignment which was considered as gold standard. A practical Catalyst setup error was calculated between the translational deviations of the surface scanner and the laser positioning. Two one-sided t tests for equivalence were used for statistical analysis. RESULTS: Data analysis revealed total deviations of 0.09 mm ± 2.03 mm for the lateral axis, 0.07 mm ± 3.21 mm for the longitudinal axis, and 0.44 mm ± 3.08 mm vertical axis for the Catalyst system, compared to −0.06 ± 3.54 mm lateral, 0.53 ± 3.47 mm longitudinal, and 0.19 ± 3.49 mm vertical for the laser positioning compared to cone beam computed tomography. The lowest positional deviations were found in the cranial region, and larger deviations occurred in the thoracic and abdominal sites. A statistical comparison using 2 one-sided t tests showed a general concordance of the 2 methods (P ≤ 0.036), excluding the vertical direction of the abdominal region (P = 0.198). CONCLUSION: The optical surface scanner Catalyst HD is a reliable and feasible patient positioning system without any additional radiation exposure. From the head to the thoracic and abdominal region, a decrease in accuracy was observed within a comparable range for Catalyst and laser-assisted positioning. SAGE Publications 2018-11-19 /pmc/articles/PMC6243634/ /pubmed/30453842 http://dx.doi.org/10.1177/1533033818806002 Text en © The Author(s) 2018 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Article
Carl, G
Reitz, D
Schönecker, S
Pazos, M
Freislederer, P
Reiner, M
Alongi, F
Niyazi, M
Ganswindt, U
Belka, C
Corradini, S
Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions
title Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions
title_full Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions
title_fullStr Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions
title_full_unstemmed Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions
title_short Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions
title_sort optical surface scanning for patient positioning in radiation therapy: a prospective analysis of 1902 fractions
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6243634/
https://www.ncbi.nlm.nih.gov/pubmed/30453842
http://dx.doi.org/10.1177/1533033818806002
work_keys_str_mv AT carlg opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT reitzd opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT schoneckers opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT pazosm opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT freisledererp opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT reinerm opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT alongif opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT niyazim opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT ganswindtu opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT belkac opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions
AT corradinis opticalsurfacescanningforpatientpositioninginradiationtherapyaprospectiveanalysisof1902fractions