Cargando…
Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study
PURPOSE: Altered respiratory rate (RR) has been identified as an important predictor of serious adverse events during hospitalization. Introduction of a well-tolerated continuous RR monitor could potentially reduce serious adverse events such as opioid-induced respiratory depression. The purpose of...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer US
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6244627/ https://www.ncbi.nlm.nih.gov/pubmed/30194672 http://dx.doi.org/10.1007/s12630-018-1214-z |
_version_ | 1783372088465162240 |
---|---|
author | van Loon, Kim Peelen, Linda M. van de Vlasakker, Emmy C. Kalkman, Cor J. van Wolfswinkel, Leo van Zaane, Bas |
author_facet | van Loon, Kim Peelen, Linda M. van de Vlasakker, Emmy C. Kalkman, Cor J. van Wolfswinkel, Leo van Zaane, Bas |
author_sort | van Loon, Kim |
collection | PubMed |
description | PURPOSE: Altered respiratory rate (RR) has been identified as an important predictor of serious adverse events during hospitalization. Introduction of a well-tolerated continuous RR monitor could potentially reduce serious adverse events such as opioid-induced respiratory depression. The purpose of this study was to investigate the ability of different monitor devices to detect RR in low care clinical settings. METHODS: This was a prospective method-comparison study with a cross-sectional design. Thoracic impedance pneumography (IPG), frequency modulated continuous wave radar, and an acoustic breath sounds monitor were compared with the gold standard of capnography for their ability to detect RR in breaths per minute (breaths·min(−1)) in awake postoperative patients in the postanesthesia care unit. The Bland and Altman method for repeated measurements and mixed effect modelling was used to obtain bias and limits of agreement (LoA). Furthermore, the ability of the three devices to assist with correct treatment decisions was evaluated in Clarke Error Grids. RESULTS: Twenty patients were monitored for 1,203 min, with a median [interquartile range] of 61 [60-63] min per patient. The bias (98.9% LoA) were 0.1 (−7.9 to 7.9) breaths·min(−1) for the acoustic monitor, −1.6 (−10.8 to 7.6) for the radar, and −1.9 (−13.1 to 9.2) for the IPG. The extent to which the monitors guided adequate or led to inadequate treatment decisions (determined by Clarke Error Grid analysis) differed significantly between the monitors (P = 0.011). Decisions were correct 96% of the time for acoustic, 95% of the time for radar, and 94% of the time for IPG monitoring devices. CONCLUSIONS: None of the studied devices (acoustic, IPG, and radar monitor) had LoA that were within our predefined (based on clinical judgement) limits of ± 2 breaths·min(−1). The acoustic breath sound monitor predicted the correct treatment more often than the IPG and the radar device. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s12630-018-1214-z) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-6244627 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Springer US |
record_format | MEDLINE/PubMed |
spelling | pubmed-62446272018-12-04 Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study van Loon, Kim Peelen, Linda M. van de Vlasakker, Emmy C. Kalkman, Cor J. van Wolfswinkel, Leo van Zaane, Bas Can J Anaesth Reports of Original Investigations PURPOSE: Altered respiratory rate (RR) has been identified as an important predictor of serious adverse events during hospitalization. Introduction of a well-tolerated continuous RR monitor could potentially reduce serious adverse events such as opioid-induced respiratory depression. The purpose of this study was to investigate the ability of different monitor devices to detect RR in low care clinical settings. METHODS: This was a prospective method-comparison study with a cross-sectional design. Thoracic impedance pneumography (IPG), frequency modulated continuous wave radar, and an acoustic breath sounds monitor were compared with the gold standard of capnography for their ability to detect RR in breaths per minute (breaths·min(−1)) in awake postoperative patients in the postanesthesia care unit. The Bland and Altman method for repeated measurements and mixed effect modelling was used to obtain bias and limits of agreement (LoA). Furthermore, the ability of the three devices to assist with correct treatment decisions was evaluated in Clarke Error Grids. RESULTS: Twenty patients were monitored for 1,203 min, with a median [interquartile range] of 61 [60-63] min per patient. The bias (98.9% LoA) were 0.1 (−7.9 to 7.9) breaths·min(−1) for the acoustic monitor, −1.6 (−10.8 to 7.6) for the radar, and −1.9 (−13.1 to 9.2) for the IPG. The extent to which the monitors guided adequate or led to inadequate treatment decisions (determined by Clarke Error Grid analysis) differed significantly between the monitors (P = 0.011). Decisions were correct 96% of the time for acoustic, 95% of the time for radar, and 94% of the time for IPG monitoring devices. CONCLUSIONS: None of the studied devices (acoustic, IPG, and radar monitor) had LoA that were within our predefined (based on clinical judgement) limits of ± 2 breaths·min(−1). The acoustic breath sound monitor predicted the correct treatment more often than the IPG and the radar device. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (10.1007/s12630-018-1214-z) contains supplementary material, which is available to authorized users. Springer US 2018-09-07 2018 /pmc/articles/PMC6244627/ /pubmed/30194672 http://dx.doi.org/10.1007/s12630-018-1214-z Text en © The Author(s) 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Reports of Original Investigations van Loon, Kim Peelen, Linda M. van de Vlasakker, Emmy C. Kalkman, Cor J. van Wolfswinkel, Leo van Zaane, Bas Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study |
title | Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study |
title_full | Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study |
title_fullStr | Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study |
title_full_unstemmed | Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study |
title_short | Accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study |
title_sort | accuracy of remote continuous respiratory rate monitoring technologies intended for low care clinical settings: a prospective observational study |
topic | Reports of Original Investigations |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6244627/ https://www.ncbi.nlm.nih.gov/pubmed/30194672 http://dx.doi.org/10.1007/s12630-018-1214-z |
work_keys_str_mv | AT vanloonkim accuracyofremotecontinuousrespiratoryratemonitoringtechnologiesintendedforlowcareclinicalsettingsaprospectiveobservationalstudy AT peelenlindam accuracyofremotecontinuousrespiratoryratemonitoringtechnologiesintendedforlowcareclinicalsettingsaprospectiveobservationalstudy AT vandevlasakkeremmyc accuracyofremotecontinuousrespiratoryratemonitoringtechnologiesintendedforlowcareclinicalsettingsaprospectiveobservationalstudy AT kalkmancorj accuracyofremotecontinuousrespiratoryratemonitoringtechnologiesintendedforlowcareclinicalsettingsaprospectiveobservationalstudy AT vanwolfswinkelleo accuracyofremotecontinuousrespiratoryratemonitoringtechnologiesintendedforlowcareclinicalsettingsaprospectiveobservationalstudy AT vanzaanebas accuracyofremotecontinuousrespiratoryratemonitoringtechnologiesintendedforlowcareclinicalsettingsaprospectiveobservationalstudy |