PRaVDA: The first solid-state system for proton computed tomography

PURPOSE: Proton CT is widely recognised as a beneficial alternative to conventional X-ray CT for treatment planning in proton beam radiotherapy. A novel proton CT imaging system, based entirely on solid-state detector technology, is presented. Compared to conventional scintillator-based calorimeters...

Descripción completa

Detalles Bibliográficos
Autores principales: Esposito, Michela, Waltham, Chris, Taylor, Jonathan T., Manger, Sam, Phoenix, Ben, Price, Tony, Poludniowski, Gavin, Green, Stuart, Evans, Philip M., Allport, Philip P., Manolopulos, Spyros, Nieto-Camero, Jaime, Symons, Julyan, Allinson, Nigel M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Istituti Editoriali e Poligrafici Internazionali 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6250048/
https://www.ncbi.nlm.nih.gov/pubmed/30420271
http://dx.doi.org/10.1016/j.ejmp.2018.10.020
Descripción
Sumario:PURPOSE: Proton CT is widely recognised as a beneficial alternative to conventional X-ray CT for treatment planning in proton beam radiotherapy. A novel proton CT imaging system, based entirely on solid-state detector technology, is presented. Compared to conventional scintillator-based calorimeters, positional sensitive detectors allow for multiple protons to be tracked per read out cycle, leading to a potential reduction in proton CT scan time. Design and characterisation of its components are discussed. An early proton CT image obtained with a fully solid-state imaging system is shown and accuracy (as defined in Section IV) in Relative Stopping Power to water (RSP) quantified. METHOD: A solid-state imaging system for proton CT, based on silicon strip detectors, has been developed by the PRaVDA collaboration. The system comprises a tracking system that infers individual proton trajectories through an imaging phantom, and a Range Telescope (RT) which records the corresponding residual energy (range) for each proton. A back-projection-then-filtering algorithm is used for CT reconstruction of an experimentally acquired proton CT scan. RESULTS: An initial experimental result for proton CT imaging with a fully solid-state system is shown for an imaging phantom, namely a 75 mm diameter PMMA sphere containing tissue substitute inserts, imaged with a passively-scattered 125 MeV beam. Accuracy in RSP is measured to be [Formula: see text] 1.6% for all the inserts shown. CONCLUSIONS: A fully solid-state imaging system for proton CT has been shown capable of imaging a phantom with protons and successfully improving RSP accuracy. These promising results, together with system the capability to cope with high proton fluences ([Formula: see text]  protons/s), suggests that this research platform could improve current standards in treatment planning for proton beam radiotherapy.