Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing
OBJECTIVE: The Welch Allyn Pro BP 2000, an automated sphygmomanometer, was subjected to ANSI/AAMI/ISO testing. PATIENTS AND METHODS: The protocol used was the same-arm sequential procedure described in the ANSI/AAMI/ISO 2013 Standard. Eight-eight (53 adults and 35 children aged 3–12 years old) patie...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6250254/ https://www.ncbi.nlm.nih.gov/pubmed/30312180 http://dx.doi.org/10.1097/MBP.0000000000000350 |
_version_ | 1783372898605465600 |
---|---|
author | Alpert, Bruce S. |
author_facet | Alpert, Bruce S. |
author_sort | Alpert, Bruce S. |
collection | PubMed |
description | OBJECTIVE: The Welch Allyn Pro BP 2000, an automated sphygmomanometer, was subjected to ANSI/AAMI/ISO testing. PATIENTS AND METHODS: The protocol used was the same-arm sequential procedure described in the ANSI/AAMI/ISO 2013 Standard. Eight-eight (53 adults and 35 children aged 3–12 years old) patients completed testing successfully. Arm circumferences ranged from 15 to 49.5 cm. Seven different cuff sizes were used, including two long cuffs. RESULTS: All requirements for age, sex, blood pressure (BP), and cuff sizes were fulfilled. The mean±SD was −2.8±6.37 mmHg for systolic BP for criterion 1 and −3.6±6.14 mmHg for diastolic BP. The SD for criterion 2 were 5.29 for systolic BP and 5.75 for diastolic BP. All data passed the Standard’s requirements. CONCLUSION: The Pro BP 2000 uses an accurate inflationary algorithm. The time for each inflation/deflation cycle is short, thus improving patient comfort. There is also arrhythmia detection to caution the use of BP values obtained during irregular heart rhythms. The algorithm for the professional-grade Pro BP 2000 is also contained in the Home BP 1700, targeted for home use. This encourages out-of-office self-measurement. The healthcare professional can be confident that the values obtained at different sites are comparable. |
format | Online Article Text |
id | pubmed-6250254 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Lippincott Williams & Wilkins |
record_format | MEDLINE/PubMed |
spelling | pubmed-62502542018-12-10 Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing Alpert, Bruce S. Blood Press Monit Devices and Technology OBJECTIVE: The Welch Allyn Pro BP 2000, an automated sphygmomanometer, was subjected to ANSI/AAMI/ISO testing. PATIENTS AND METHODS: The protocol used was the same-arm sequential procedure described in the ANSI/AAMI/ISO 2013 Standard. Eight-eight (53 adults and 35 children aged 3–12 years old) patients completed testing successfully. Arm circumferences ranged from 15 to 49.5 cm. Seven different cuff sizes were used, including two long cuffs. RESULTS: All requirements for age, sex, blood pressure (BP), and cuff sizes were fulfilled. The mean±SD was −2.8±6.37 mmHg for systolic BP for criterion 1 and −3.6±6.14 mmHg for diastolic BP. The SD for criterion 2 were 5.29 for systolic BP and 5.75 for diastolic BP. All data passed the Standard’s requirements. CONCLUSION: The Pro BP 2000 uses an accurate inflationary algorithm. The time for each inflation/deflation cycle is short, thus improving patient comfort. There is also arrhythmia detection to caution the use of BP values obtained during irregular heart rhythms. The algorithm for the professional-grade Pro BP 2000 is also contained in the Home BP 1700, targeted for home use. This encourages out-of-office self-measurement. The healthcare professional can be confident that the values obtained at different sites are comparable. Lippincott Williams & Wilkins 2018-12 2018-10-11 /pmc/articles/PMC6250254/ /pubmed/30312180 http://dx.doi.org/10.1097/MBP.0000000000000350 Text en Copyright © 2018 The Author(s). Published by Wolters Kluwer Health, Inc. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (http://creativecommons.org/licenses/by-nc-nd/4.0/) (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal. http://creativecommons.org/licenses/by-nc-nd/4.0/ |
spellingShingle | Devices and Technology Alpert, Bruce S. Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing |
title | Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing |
title_full | Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing |
title_fullStr | Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing |
title_full_unstemmed | Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing |
title_short | Validation of the Welch Allyn Pro BP 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ANSI/AAMI/ISO standard testing |
title_sort | validation of the welch allyn pro bp 2000, a professional-grade inflation-based automated sphygmomanometer with arrhythmia detection in a combined pediatric and adult population by ansi/aami/iso standard testing |
topic | Devices and Technology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6250254/ https://www.ncbi.nlm.nih.gov/pubmed/30312180 http://dx.doi.org/10.1097/MBP.0000000000000350 |