Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system

BACKGROUND: PoleStar m660 is a newly developed clinical PET/CT system with time-of-flight (TOF) capability. The aim of this study is to characterize the performance of the new system. Spatial resolution, sensitivity, scatter fraction, and noise equivalent count rate (NECR) were measured on the scann...

Descripción completa

Detalles Bibliográficos
Autores principales: Huo, Li, Li, Nan, Wu, Heyu, Zhu, Wenjia, Xing, Haiqun, Ba, Jiantao, Wang, Tong, Li, Fang, Zhang, Hui
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6269439/
https://www.ncbi.nlm.nih.gov/pubmed/30506174
http://dx.doi.org/10.1186/s40658-018-0229-4
_version_ 1783376469944172544
author Huo, Li
Li, Nan
Wu, Heyu
Zhu, Wenjia
Xing, Haiqun
Ba, Jiantao
Wang, Tong
Li, Fang
Zhang, Hui
author_facet Huo, Li
Li, Nan
Wu, Heyu
Zhu, Wenjia
Xing, Haiqun
Ba, Jiantao
Wang, Tong
Li, Fang
Zhang, Hui
author_sort Huo, Li
collection PubMed
description BACKGROUND: PoleStar m660 is a newly developed clinical PET/CT system with time-of-flight (TOF) capability. The aim of this study is to characterize the performance of the new system. Spatial resolution, sensitivity, scatter fraction, and noise equivalent count rate (NECR) were measured on the scanner according to the NEMA NU 2-2012 protocol. The timing resolution was measured using a rotating line source that orbited around the center of field of view (CFOV) at a radius of 20 cm. The image quality phantom was also imaged to quantify the percent contrast, percent background variability, and residual error. The impacts of data acquisition time and bed overlap on the PET image quality were also evaluated using phantom and patient studies. RESULTS: The transverse (axial) spatial resolutions were 3.59 mm (3.67), 4.08 mm (4.65), and 5.32 mm (6.48) full width at half maximum (FWHM) at 1 cm, 10 cm, and 20 cm, respectively, off the CFOV. The measured sensitivity was 10.7 cps/kBq at the CFOV and 10.4 cps/kBq at 10 cm off the CFOV. The peak NECR was 216.7 kcps at an activity concentration of 29.1 kBq/ml, and the scatter fraction was 38.2%. An average of 435 ps FWHM timing resolution was measured. For the image quality phantom, the contrast recovery ratios ranged from 33.9 to 76.4%, while the background variability ranged from 4.7 to 2.0%. In the preliminary clinical study, no noticeable difference in the image quality was observed when the scan time for the whole body and brain was reduced to 1 min/bed and 3 min, respectively. The tested 21% bed overlap showed no significant difference in the image quality compared with the default 38% bed overlap setting. CONCLUSIONS: The physical performances of the PoleStar m660 PET/CT system showed good sensitivity, count rate performance, and timing resolution. The improved performance could help to reduce the acquisition time and bed overlap in the clinical application without detectable compromise in the image quality.
format Online
Article
Text
id pubmed-6269439
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-62694392018-12-18 Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system Huo, Li Li, Nan Wu, Heyu Zhu, Wenjia Xing, Haiqun Ba, Jiantao Wang, Tong Li, Fang Zhang, Hui EJNMMI Phys Original Research BACKGROUND: PoleStar m660 is a newly developed clinical PET/CT system with time-of-flight (TOF) capability. The aim of this study is to characterize the performance of the new system. Spatial resolution, sensitivity, scatter fraction, and noise equivalent count rate (NECR) were measured on the scanner according to the NEMA NU 2-2012 protocol. The timing resolution was measured using a rotating line source that orbited around the center of field of view (CFOV) at a radius of 20 cm. The image quality phantom was also imaged to quantify the percent contrast, percent background variability, and residual error. The impacts of data acquisition time and bed overlap on the PET image quality were also evaluated using phantom and patient studies. RESULTS: The transverse (axial) spatial resolutions were 3.59 mm (3.67), 4.08 mm (4.65), and 5.32 mm (6.48) full width at half maximum (FWHM) at 1 cm, 10 cm, and 20 cm, respectively, off the CFOV. The measured sensitivity was 10.7 cps/kBq at the CFOV and 10.4 cps/kBq at 10 cm off the CFOV. The peak NECR was 216.7 kcps at an activity concentration of 29.1 kBq/ml, and the scatter fraction was 38.2%. An average of 435 ps FWHM timing resolution was measured. For the image quality phantom, the contrast recovery ratios ranged from 33.9 to 76.4%, while the background variability ranged from 4.7 to 2.0%. In the preliminary clinical study, no noticeable difference in the image quality was observed when the scan time for the whole body and brain was reduced to 1 min/bed and 3 min, respectively. The tested 21% bed overlap showed no significant difference in the image quality compared with the default 38% bed overlap setting. CONCLUSIONS: The physical performances of the PoleStar m660 PET/CT system showed good sensitivity, count rate performance, and timing resolution. The improved performance could help to reduce the acquisition time and bed overlap in the clinical application without detectable compromise in the image quality. Springer International Publishing 2018-12-01 /pmc/articles/PMC6269439/ /pubmed/30506174 http://dx.doi.org/10.1186/s40658-018-0229-4 Text en © The Author(s). 2018 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Huo, Li
Li, Nan
Wu, Heyu
Zhu, Wenjia
Xing, Haiqun
Ba, Jiantao
Wang, Tong
Li, Fang
Zhang, Hui
Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
title Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
title_full Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
title_fullStr Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
title_full_unstemmed Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
title_short Performance evaluation of a new high-sensitivity time-of-flight clinical PET/CT system
title_sort performance evaluation of a new high-sensitivity time-of-flight clinical pet/ct system
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6269439/
https://www.ncbi.nlm.nih.gov/pubmed/30506174
http://dx.doi.org/10.1186/s40658-018-0229-4
work_keys_str_mv AT huoli performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT linan performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT wuheyu performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT zhuwenjia performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT xinghaiqun performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT bajiantao performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT wangtong performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT lifang performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem
AT zhanghui performanceevaluationofanewhighsensitivitytimeofflightclinicalpetctsystem