Cargando…
Methods of Increasing the Performance of Radionuclide Generators Used in Nuclear Medicine: Daughter Nuclide Build-Up Optimisation, Elution-Purification-Concentration Integration, and Effective Control of Radionuclidic Purity
Methods of increasing the performance of radionuclide generators used in nuclear medicine radiotherapy and SPECT/PET imaging were developed and detailed for (99)Mo/(99m)Tc and (68)Ge/(68)Ga radionuclide generators as the cases. Optimisation methods of the daughter nuclide build-up versus stand-by ti...
Autores principales: | Le, Van So, Do, Zoe Phuc-Hien, Le, Minh Khoi, Le, Vicki, Le, Natalie Nha-Truc |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6271982/ https://www.ncbi.nlm.nih.gov/pubmed/24918542 http://dx.doi.org/10.3390/molecules19067714 |