Cargando…
Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients
AIMS: To analyse the impact of device and software updates on the prevention of T‑wave oversensing (TWOS) and inappropriate shocks (IS) in subcutaneous ICD (S-ICD) patients. BACKGROUND: TWOS is a feared complication after implantation. It may lead to harmful IS. To date, specific strategies to reduc...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Bohn Stafleu van Loghum
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6288039/ https://www.ncbi.nlm.nih.gov/pubmed/30250998 http://dx.doi.org/10.1007/s12471-018-1160-y |
_version_ | 1783379722490609664 |
---|---|
author | Larbig, R. Motloch, L. J. Bettin, M. Fischer, A. Bode, N. Frommeyer, G. Loeher, A. Koebe, J. Reinke, F. Eckardt, L. |
author_facet | Larbig, R. Motloch, L. J. Bettin, M. Fischer, A. Bode, N. Frommeyer, G. Loeher, A. Koebe, J. Reinke, F. Eckardt, L. |
author_sort | Larbig, R. |
collection | PubMed |
description | AIMS: To analyse the impact of device and software updates on the prevention of T‑wave oversensing (TWOS) and inappropriate shocks (IS) in subcutaneous ICD (S-ICD) patients. BACKGROUND: TWOS is a feared complication after implantation. It may lead to harmful IS. To date, specific strategies to reduce these events are lacking. METHODS: In this retrospective single-centre trial we analysed 146 S‑ICD patients who were implanted between 2010 and 2016. In all eligible consecutive patients (n = 139), follow-up of at least 6 weeks was studied. The incidence of TWOS/IS was analysed in patients receiving a 2(nd) generation S‑ICD (Emblem-S-ICD) between 2014 and 2016 (Emblem). Their outcome was compared with a control group (SQ) treated with the SQ1010 device between 2010 and 2014, who were followed up for a maximum of 2 years. Furthermore, to test if the software update SMR8 reduces inappropriate shocks in the SQ1010-S-ICD population, the incidence of TWOS/IS was evaluated before and after update installation. RESULTS: Basic characteristics and indications for S‑ICD implantation were similar in both groups. However, the cumulative incidence of TWOS/IS was significantly decreased in Emblem vs. SQ (SQ: 15.4%, n = 14/91 vs. Emblem 4.2%, n = 2/48; p = 0.049). Furthermore, with regards to the SQ population we also observed a trend towards a significant reduction of TWOS/IS after installation of the software update SMR8 in 2014 (before update: 13.4%, n = 11/82 vs. after update: 4.6%, 3/65, p = 0.07). CONCLUSION: 2(nd) generation devices but probably also the SMR8 software update reduce the incidence of TWOS/IS in S‑ICD patients. |
format | Online Article Text |
id | pubmed-6288039 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Bohn Stafleu van Loghum |
record_format | MEDLINE/PubMed |
spelling | pubmed-62880392018-12-26 Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients Larbig, R. Motloch, L. J. Bettin, M. Fischer, A. Bode, N. Frommeyer, G. Loeher, A. Koebe, J. Reinke, F. Eckardt, L. Neth Heart J Original Article AIMS: To analyse the impact of device and software updates on the prevention of T‑wave oversensing (TWOS) and inappropriate shocks (IS) in subcutaneous ICD (S-ICD) patients. BACKGROUND: TWOS is a feared complication after implantation. It may lead to harmful IS. To date, specific strategies to reduce these events are lacking. METHODS: In this retrospective single-centre trial we analysed 146 S‑ICD patients who were implanted between 2010 and 2016. In all eligible consecutive patients (n = 139), follow-up of at least 6 weeks was studied. The incidence of TWOS/IS was analysed in patients receiving a 2(nd) generation S‑ICD (Emblem-S-ICD) between 2014 and 2016 (Emblem). Their outcome was compared with a control group (SQ) treated with the SQ1010 device between 2010 and 2014, who were followed up for a maximum of 2 years. Furthermore, to test if the software update SMR8 reduces inappropriate shocks in the SQ1010-S-ICD population, the incidence of TWOS/IS was evaluated before and after update installation. RESULTS: Basic characteristics and indications for S‑ICD implantation were similar in both groups. However, the cumulative incidence of TWOS/IS was significantly decreased in Emblem vs. SQ (SQ: 15.4%, n = 14/91 vs. Emblem 4.2%, n = 2/48; p = 0.049). Furthermore, with regards to the SQ population we also observed a trend towards a significant reduction of TWOS/IS after installation of the software update SMR8 in 2014 (before update: 13.4%, n = 11/82 vs. after update: 4.6%, 3/65, p = 0.07). CONCLUSION: 2(nd) generation devices but probably also the SMR8 software update reduce the incidence of TWOS/IS in S‑ICD patients. Bohn Stafleu van Loghum 2018-09-24 2018-12 /pmc/articles/PMC6288039/ /pubmed/30250998 http://dx.doi.org/10.1007/s12471-018-1160-y Text en © The Author(s) 2018 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Original Article Larbig, R. Motloch, L. J. Bettin, M. Fischer, A. Bode, N. Frommeyer, G. Loeher, A. Koebe, J. Reinke, F. Eckardt, L. Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients |
title | Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients |
title_full | Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients |
title_fullStr | Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients |
title_full_unstemmed | Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients |
title_short | Device updates successfully reduce T‑wave oversensing and inappropriate shocks in subcutaneous ICD patients |
title_sort | device updates successfully reduce t‑wave oversensing and inappropriate shocks in subcutaneous icd patients |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6288039/ https://www.ncbi.nlm.nih.gov/pubmed/30250998 http://dx.doi.org/10.1007/s12471-018-1160-y |
work_keys_str_mv | AT larbigr deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT motlochlj deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT bettinm deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT fischera deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT boden deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT frommeyerg deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT loehera deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT koebej deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT reinkef deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients AT eckardtl deviceupdatessuccessfullyreducetwaveoversensingandinappropriateshocksinsubcutaneousicdpatients |