Standardization of the HbA(2) Assay
BACKGROUND: A project for the standardization of HbA(2) was launched by the IFCC back in 2004. MATERIALS AND METHODS: In this work we report on the state-of-the-art of the project on standardization of HbA(2). Data obtained from various EQAS studies, and from previous experimental evaluations, are p...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Communications and Publications Division (CPD) of the IFCC
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6295589/ https://www.ncbi.nlm.nih.gov/pubmed/30574041 |
Sumario: | BACKGROUND: A project for the standardization of HbA(2) was launched by the IFCC back in 2004. MATERIALS AND METHODS: In this work we report on the state-of-the-art of the project on standardization of HbA(2). Data obtained from various EQAS studies, and from previous experimental evaluations, are presented. RESULTS: We have proven that biases between various commercial methods are still currently significant. We have also shown that calibration by commutable control materials may halve the inter-method variability. CONCLUSIONS: The foundation of the reference system for HbA(2), together with a brief preliminary presentation of the proposed primary reference measurement procedure based on ID-MS are outlined. |
---|