Cargando…

Seven-year clinical outcomes in patients undergoing percutaneous coronary intervention with biodegradable polymer coated sirolimus-eluting stent: Results from a single-center real-world experience

OBJECTIVE: The aim of the present study was to assess seven-year clinical outcomes of biodegradable polymer coated Supralimus sirolimus-eluting stent (S-SES) [Sahajanand Medical Technologies Pvt. Ltd., Surat, India] in real-world patients with coronary artery disease. METHODS: This observational, re...

Descripción completa

Detalles Bibliográficos
Autores principales: Abhyankar, Atul, Kaul, Upendra, Sinha, Santosh Kumar
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6309142/
https://www.ncbi.nlm.nih.gov/pubmed/30595275
http://dx.doi.org/10.1016/j.ihj.2018.05.014
Descripción
Sumario:OBJECTIVE: The aim of the present study was to assess seven-year clinical outcomes of biodegradable polymer coated Supralimus sirolimus-eluting stent (S-SES) [Sahajanand Medical Technologies Pvt. Ltd., Surat, India] in real-world patients with coronary artery disease. METHODS: This observational, retrospective study was carried out in all 346 consecutive enrolled patients who underwent percutaneous coronary intervention (PCI) with the S-SES, between April 2008 and December 2009, at a single center. We analyzed major adverse cardiac events (MACE) [a composite of cardiac death, myocardial infarction (MI), target lesion revascularization (TLR) and target vessel revascularization (TVR)] as primary outcomes at seven-year follow-up. RESULTS: Out of 346 patients, seven-year follow-up was obtained in 327 (94.5%) patients and hence results were analyzed for 327 patients. At seven-year, MACE occurred in 41 (12.5%) patients, consisting of 23 (7.0%) cardiac deaths, 14 (4.3%) TLR, and 4 (1.2%) TVR. The incidence of late stent thrombosis was observed in 3 (0.9%) patients. At follow-up of seven-year, the cumulative event-free survival was found to be 84.7% by Kaplan-Meier method. CONCLUSIONS: The present study demonstrated satisfactory and sustained seven-year clinical outcomes as evidenced by the low rates of MACE and ST for the biodegradable polymer coated S-SES.