Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes
BACKGROUND: Secretions accumulate in endotracheal tubes' (ETT) lumens upon their placement in patients. The secretions impact airway resistance and pressure. Secretions potentiate prolonged mechanical ventilation and ventilator-associated pneumonia. Our primary objective in this study was to ev...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6311789/ https://www.ncbi.nlm.nih.gov/pubmed/30652032 http://dx.doi.org/10.1155/2018/3258396 |
_version_ | 1783383669253079040 |
---|---|
author | Waters, Christopher Wiener, R. Constance Motlagh, Hamed M. |
author_facet | Waters, Christopher Wiener, R. Constance Motlagh, Hamed M. |
author_sort | Waters, Christopher |
collection | PubMed |
description | BACKGROUND: Secretions accumulate in endotracheal tubes' (ETT) lumens upon their placement in patients. The secretions impact airway resistance and pressure. Secretions potentiate prolonged mechanical ventilation and ventilator-associated pneumonia. Our primary objective in this study was to evaluate an ETT-clearing device (ETT-CD) in its ability to remove secretions from ex vivo ETT lumens. METHODS: Forty ETTs, obtained from intensive care patients at extubation, were individually placed into a ventilator field performance testing simulator at 37°C. The pressure drop through the ETTs was measured at a flow rate of 60 L/min before and after cleaning with the ETT-CD and compared with unused, similarly sized controls tubes. The ETT-CD was inserted into an ETT until the tip reached Murphy's eye (hole in the side) of the ETT. The wiper, set back from the tip, was expanded by ETT-CD handle activation. As the ETT-CD was removed, the distal wiper extracted secretions from the ETT lumen. RESULTS: Forty ETTs were tested with nonparametric Wilcoxon signed-rank tests. Before being cleared with the ETT-CD, the median pressure drop in the extubated 7.5 mm ETTs was 17.8 cm H(2)O; after ETT-CD use, it was 12.3. The cleared ETTs were significantly improved over the ETTs before being cleared (p < 0.001); however, there remained a significant difference between the cleared ETTs and the control tubes (p=0.005), indicating the clearing was not to the level of an unused ETT. Similar results were determined for the 8.0 mm ETTs. CONCLUSIONS: For the 7.5 mm and the 8.0 mm EETs, the ETT-CD improved effective patency of the ETTs over the uncleared ETTs, independent of occlusion location, tube size, or length of tube. However, there remained a significant difference between the cleared tubes and controls. |
format | Online Article Text |
id | pubmed-6311789 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-63117892019-01-16 Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes Waters, Christopher Wiener, R. Constance Motlagh, Hamed M. Crit Care Res Pract Research Article BACKGROUND: Secretions accumulate in endotracheal tubes' (ETT) lumens upon their placement in patients. The secretions impact airway resistance and pressure. Secretions potentiate prolonged mechanical ventilation and ventilator-associated pneumonia. Our primary objective in this study was to evaluate an ETT-clearing device (ETT-CD) in its ability to remove secretions from ex vivo ETT lumens. METHODS: Forty ETTs, obtained from intensive care patients at extubation, were individually placed into a ventilator field performance testing simulator at 37°C. The pressure drop through the ETTs was measured at a flow rate of 60 L/min before and after cleaning with the ETT-CD and compared with unused, similarly sized controls tubes. The ETT-CD was inserted into an ETT until the tip reached Murphy's eye (hole in the side) of the ETT. The wiper, set back from the tip, was expanded by ETT-CD handle activation. As the ETT-CD was removed, the distal wiper extracted secretions from the ETT lumen. RESULTS: Forty ETTs were tested with nonparametric Wilcoxon signed-rank tests. Before being cleared with the ETT-CD, the median pressure drop in the extubated 7.5 mm ETTs was 17.8 cm H(2)O; after ETT-CD use, it was 12.3. The cleared ETTs were significantly improved over the ETTs before being cleared (p < 0.001); however, there remained a significant difference between the cleared ETTs and the control tubes (p=0.005), indicating the clearing was not to the level of an unused ETT. Similar results were determined for the 8.0 mm ETTs. CONCLUSIONS: For the 7.5 mm and the 8.0 mm EETs, the ETT-CD improved effective patency of the ETTs over the uncleared ETTs, independent of occlusion location, tube size, or length of tube. However, there remained a significant difference between the cleared tubes and controls. Hindawi 2018-12-10 /pmc/articles/PMC6311789/ /pubmed/30652032 http://dx.doi.org/10.1155/2018/3258396 Text en Copyright © 2018 Christopher Waters et al. https://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Waters, Christopher Wiener, R. Constance Motlagh, Hamed M. Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes |
title |
Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes |
title_full |
Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes |
title_fullStr |
Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes |
title_full_unstemmed |
Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes |
title_short |
Ex Vivo Evaluation of Secretion-Clearing Device in Reducing Airway Resistance within Endotracheal Tubes |
title_sort | ex vivo evaluation of secretion-clearing device in reducing airway resistance within endotracheal tubes |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6311789/ https://www.ncbi.nlm.nih.gov/pubmed/30652032 http://dx.doi.org/10.1155/2018/3258396 |
work_keys_str_mv | AT waterschristopher exvivoevaluationofsecretionclearingdeviceinreducingairwayresistancewithinendotrachealtubes AT wienerrconstance exvivoevaluationofsecretionclearingdeviceinreducingairwayresistancewithinendotrachealtubes AT motlaghhamedm exvivoevaluationofsecretionclearingdeviceinreducingairwayresistancewithinendotrachealtubes |