Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response

PURPOSE: The purpose of this study was to investigate the effect of interstimulus frequency on the photopic negative response (PhNR) in the clinical electroretinogram (ERG) in glaucoma and healthy eyes. METHODS: Participants with open angle glaucoma (n = 15) and age-matched controls (n = 20) were re...

Descripción completa

Detalles Bibliográficos
Autores principales: Hui, Flora, Tang, Jessica, Hadoux, Xavier, Coote, Michael, Crowston, Jonathan G.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6314062/
https://www.ncbi.nlm.nih.gov/pubmed/30619646
http://dx.doi.org/10.1167/tvst.7.6.26
_version_ 1783384065378877440
author Hui, Flora
Tang, Jessica
Hadoux, Xavier
Coote, Michael
Crowston, Jonathan G.
author_facet Hui, Flora
Tang, Jessica
Hadoux, Xavier
Coote, Michael
Crowston, Jonathan G.
author_sort Hui, Flora
collection PubMed
description PURPOSE: The purpose of this study was to investigate the effect of interstimulus frequency on the photopic negative response (PhNR) in the clinical electroretinogram (ERG) in glaucoma and healthy eyes. METHODS: Participants with open angle glaucoma (n = 15) and age-matched controls (n = 20) were recruited. Photopic ERGs were recorded in one eye using five frequencies (1–5 Hz) delivered in random order. ERGs were analyzed for changes to amplitude and timing between groups and interstimulus frequency. Coefficient of variation (CoV) was used to examine variability within recordings for each frequency. RESULTS: While the a-wave and b-wave showed minimal alteration, the PhNR was highly sensitive to changes in interstimulus frequency. The PhNR signal was largest at 1 Hz and steadily diminished with higher frequencies in both control and glaucoma groups. Significant differences in PhNR amplitude were found between controls and glaucoma groups at 2 and 3 Hz. While 1 Hz delivered the largest PhNR, it also showed a significantly greater CoV compared to other frequencies. CONCLUSIONS: An interstimulus frequency of 2 Hz was optimal for recording the PhNR, creating a good balance between testing time and signal quality. A higher frequency could be used to further shorten clinical testing times; however, this may compromise its clinical utility by dampening the PhNR. TRANSLATIONAL RELEVANCE: Here we show the importance of considering flash interstimulus frequency when designing ERG protocols for recording the PhNR as while higher frequencies can shorten test times, they also have considerable effects on the PhNR.
format Online
Article
Text
id pubmed-6314062
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-63140622019-01-07 Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response Hui, Flora Tang, Jessica Hadoux, Xavier Coote, Michael Crowston, Jonathan G. Transl Vis Sci Technol Articles PURPOSE: The purpose of this study was to investigate the effect of interstimulus frequency on the photopic negative response (PhNR) in the clinical electroretinogram (ERG) in glaucoma and healthy eyes. METHODS: Participants with open angle glaucoma (n = 15) and age-matched controls (n = 20) were recruited. Photopic ERGs were recorded in one eye using five frequencies (1–5 Hz) delivered in random order. ERGs were analyzed for changes to amplitude and timing between groups and interstimulus frequency. Coefficient of variation (CoV) was used to examine variability within recordings for each frequency. RESULTS: While the a-wave and b-wave showed minimal alteration, the PhNR was highly sensitive to changes in interstimulus frequency. The PhNR signal was largest at 1 Hz and steadily diminished with higher frequencies in both control and glaucoma groups. Significant differences in PhNR amplitude were found between controls and glaucoma groups at 2 and 3 Hz. While 1 Hz delivered the largest PhNR, it also showed a significantly greater CoV compared to other frequencies. CONCLUSIONS: An interstimulus frequency of 2 Hz was optimal for recording the PhNR, creating a good balance between testing time and signal quality. A higher frequency could be used to further shorten clinical testing times; however, this may compromise its clinical utility by dampening the PhNR. TRANSLATIONAL RELEVANCE: Here we show the importance of considering flash interstimulus frequency when designing ERG protocols for recording the PhNR as while higher frequencies can shorten test times, they also have considerable effects on the PhNR. The Association for Research in Vision and Ophthalmology 2018-12-21 /pmc/articles/PMC6314062/ /pubmed/30619646 http://dx.doi.org/10.1167/tvst.7.6.26 Text en Copyright 2018 The Authors http://creativecommons.org/licenses/by-nc-nd/4.0/ This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
spellingShingle Articles
Hui, Flora
Tang, Jessica
Hadoux, Xavier
Coote, Michael
Crowston, Jonathan G.
Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response
title Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response
title_full Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response
title_fullStr Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response
title_full_unstemmed Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response
title_short Optimizing a Portable ERG Device for Glaucoma Clinic: The Effect of Interstimulus Frequency on the Photopic Negative Response
title_sort optimizing a portable erg device for glaucoma clinic: the effect of interstimulus frequency on the photopic negative response
topic Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6314062/
https://www.ncbi.nlm.nih.gov/pubmed/30619646
http://dx.doi.org/10.1167/tvst.7.6.26
work_keys_str_mv AT huiflora optimizingaportableergdeviceforglaucomaclinictheeffectofinterstimulusfrequencyonthephotopicnegativeresponse
AT tangjessica optimizingaportableergdeviceforglaucomaclinictheeffectofinterstimulusfrequencyonthephotopicnegativeresponse
AT hadouxxavier optimizingaportableergdeviceforglaucomaclinictheeffectofinterstimulusfrequencyonthephotopicnegativeresponse
AT cootemichael optimizingaportableergdeviceforglaucomaclinictheeffectofinterstimulusfrequencyonthephotopicnegativeresponse
AT crowstonjonathang optimizingaportableergdeviceforglaucomaclinictheeffectofinterstimulusfrequencyonthephotopicnegativeresponse