Blood pressure measurements in research: suitability of auscultatory, beat-to-beat, and ambulatory blood pressure measurements

OBJECTIVE: The objective of this study was to validate the accuracy of beat-to-beat measurements with those taken with an aneroid sphygmomanometer by auscultatory method. A secondary aim was to explore differences between auscultatory and beat-to-beat blood pressure (BP) with daytime ambulatory BP m...

Descripción completa

Detalles Bibliográficos
Autores principales: Carlson, Debra J., Dieberg, Gudrun, Mcfarlane, James R., Smart, Neil A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Lippincott Williams & Wilkins 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6314505/
https://www.ncbi.nlm.nih.gov/pubmed/30489333
http://dx.doi.org/10.1097/MBP.0000000000000355
Descripción
Sumario:OBJECTIVE: The objective of this study was to validate the accuracy of beat-to-beat measurements with those taken with an aneroid sphygmomanometer by auscultatory method. A secondary aim was to explore differences between auscultatory and beat-to-beat blood pressure (BP) with daytime ambulatory BP measurements. PARTICIPANTS AND METHODS: A total of 46 participants, comprising 21 males, aged 47±13 years, height 171±8.5 cm and weight 82±16.8 kg attended the Exercise Physiology Laboratory at the University of New England (Armidale, New South Wales, Australia). During the visit, participants had their BP – systolic BP (SBP) and diastolic BP (DBP) – measured using auscultatory methods and a Finometer. An ambulatory BP monitor was fitted during the same visit and worn for a minimum of 12 h. RESULTS: Auscultatory measurements were slightly higher than beat-to-beat for both SBP and DBP. There was no difference between auscultatory and beat-to-beat SBP with a mean difference of 0.23 mmHg (P=0.87). There were disparities between auscultatory and beat-to-beat DBP, with a mean difference of 4.82 mmHg (P<0.01). Daytime ambulatory BP was higher than both auscultatory and beat-to-beat measurements for both SBP and DBP, with P less than 0.001 for all measures. CONCLUSION: There was a high level of reliability in the beat-to-beat SBP with that seen by auscultatory; however, there were disparities in DBP measurements using the same devices, which raise concerns over the accuracy of beat-to-beat DBP. Ambulatory systolic and diastolic measures were higher than beat-to-beat and auscultatory; however, they may be more suitable for monitoring diurnal changes in BP, depending upon the research model.