Cargando…

Apparent tunneling barrier height and local work function of atomic arrays

Spatially resolved measurements of the apparent tunneling barrier height Φ(app) in scanning tunneling microscopy have been used to estimate variations of the local work function Φ of surface structures. We experimentally show that Φ(app) can fail as a measure of Φ. The discrepancies are attributed t...

Descripción completa

Detalles Bibliográficos
Autores principales: Noei, Neda, Weismann, Alexander, Berndt, Richard
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Beilstein-Institut 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6317414/
https://www.ncbi.nlm.nih.gov/pubmed/30643703
http://dx.doi.org/10.3762/bjnano.9.283

Ejemplares similares