Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation
BACKGROUND: The current technology for remote assessment of fine-needle aspiration-rapid on-site evaluation (FNA-ROSE) is limited. Recent advances may provide solutions. This study compared the performance of VisionTek digital microscope (VDM) (Sakura, Japan) and Hamamatsu NanoZoomer C9600-12 single...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6319035/ https://www.ncbi.nlm.nih.gov/pubmed/30662795 http://dx.doi.org/10.4103/jpi.jpi_75_18 |
_version_ | 1783384998292750336 |
---|---|
author | Yao, Keluo Shen, Rulong Parwani, Anil Li, Zaibo |
author_facet | Yao, Keluo Shen, Rulong Parwani, Anil Li, Zaibo |
author_sort | Yao, Keluo |
collection | PubMed |
description | BACKGROUND: The current technology for remote assessment of fine-needle aspiration-rapid on-site evaluation (FNA-ROSE) is limited. Recent advances may provide solutions. This study compared the performance of VisionTek digital microscope (VDM) (Sakura, Japan) and Hamamatsu NanoZoomer C9600-12 single Z-stack digital scan (SZDS) to conventional light microscopy (CLM) for FNA-ROSE. METHODS: We assembled sixty FNA cases from the thyroid (n = 16), lymph node (n = 16), pancreas (n = 9), head and neck (n = 9), salivary gland (n = 5), lung (n = 4), and rectum (n = 1) based on a single institution's routine workflow. One Diff-Quik-stained slide was selected for each case. Two board-certified cytopathologists independently evaluated the cases using VDM, SZDS, and CLM. A “washout” period of at least 14 days was placed between the reviews. The results were categorized into satisfactory versus unsatisfactory for adequacy assessment (AA) and unsatisfactory, benign, atypical, suspicious, and malignant for preliminary diagnosis (PD). RESULTS: For AA, the Cohen's kappa statistics (CKS) scores of intermodality agreement (IMA) were 0.74–0.94 (CLM vs. VDM) and 0.86–1 (CLM vs. SZDS). The discordant rates of IMA were 3.3% (4/120) for VDM versus CLM, and 1.7% (2/120) for SZDS versus CLM. For PD, the CKS scores of IMA ranged 0.7–0.93. The overall discordant rates of IMA were 15% (18/120) for CLM versus VDM and 10.8% (13/120) for CLM versus SZDS. The discordant rates of IMA with 2 or higher degrees were 5.8% (7/120) for CLM versus VDM and 1.7% (2/120) for CLM versus SZDS. The average time spent per slide was 270 s for VDM, significantly longer than that for CLM (113 s) or for SZDS (122 s). CONCLUSIONS: Our data demonstrate that both VDM and SZDS are suitable to provide AA and reasonable PD evaluation. VDM, however, has a significantly longer turnaround time and worse diagnostic performance. The study demonstrates both the potentials and challenges of using VDM and SZDS for FNA-ROSE. |
format | Online Article Text |
id | pubmed-6319035 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2018 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-63190352019-01-18 Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation Yao, Keluo Shen, Rulong Parwani, Anil Li, Zaibo J Pathol Inform Research Article BACKGROUND: The current technology for remote assessment of fine-needle aspiration-rapid on-site evaluation (FNA-ROSE) is limited. Recent advances may provide solutions. This study compared the performance of VisionTek digital microscope (VDM) (Sakura, Japan) and Hamamatsu NanoZoomer C9600-12 single Z-stack digital scan (SZDS) to conventional light microscopy (CLM) for FNA-ROSE. METHODS: We assembled sixty FNA cases from the thyroid (n = 16), lymph node (n = 16), pancreas (n = 9), head and neck (n = 9), salivary gland (n = 5), lung (n = 4), and rectum (n = 1) based on a single institution's routine workflow. One Diff-Quik-stained slide was selected for each case. Two board-certified cytopathologists independently evaluated the cases using VDM, SZDS, and CLM. A “washout” period of at least 14 days was placed between the reviews. The results were categorized into satisfactory versus unsatisfactory for adequacy assessment (AA) and unsatisfactory, benign, atypical, suspicious, and malignant for preliminary diagnosis (PD). RESULTS: For AA, the Cohen's kappa statistics (CKS) scores of intermodality agreement (IMA) were 0.74–0.94 (CLM vs. VDM) and 0.86–1 (CLM vs. SZDS). The discordant rates of IMA were 3.3% (4/120) for VDM versus CLM, and 1.7% (2/120) for SZDS versus CLM. For PD, the CKS scores of IMA ranged 0.7–0.93. The overall discordant rates of IMA were 15% (18/120) for CLM versus VDM and 10.8% (13/120) for CLM versus SZDS. The discordant rates of IMA with 2 or higher degrees were 5.8% (7/120) for CLM versus VDM and 1.7% (2/120) for CLM versus SZDS. The average time spent per slide was 270 s for VDM, significantly longer than that for CLM (113 s) or for SZDS (122 s). CONCLUSIONS: Our data demonstrate that both VDM and SZDS are suitable to provide AA and reasonable PD evaluation. VDM, however, has a significantly longer turnaround time and worse diagnostic performance. The study demonstrates both the potentials and challenges of using VDM and SZDS for FNA-ROSE. Medknow Publications & Media Pvt Ltd 2018-12-24 /pmc/articles/PMC6319035/ /pubmed/30662795 http://dx.doi.org/10.4103/jpi.jpi_75_18 Text en Copyright: © 2018 Journal of Pathology Informatics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. |
spellingShingle | Research Article Yao, Keluo Shen, Rulong Parwani, Anil Li, Zaibo Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation |
title | Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation |
title_full | Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation |
title_fullStr | Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation |
title_full_unstemmed | Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation |
title_short | Comprehensive Study of Telecytology Using Robotic Digital Microscope and Single Z-Stack Digital Scan for Fine-Needle Aspiration-Rapid On-Site Evaluation |
title_sort | comprehensive study of telecytology using robotic digital microscope and single z-stack digital scan for fine-needle aspiration-rapid on-site evaluation |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6319035/ https://www.ncbi.nlm.nih.gov/pubmed/30662795 http://dx.doi.org/10.4103/jpi.jpi_75_18 |
work_keys_str_mv | AT yaokeluo comprehensivestudyoftelecytologyusingroboticdigitalmicroscopeandsinglezstackdigitalscanforfineneedleaspirationrapidonsiteevaluation AT shenrulong comprehensivestudyoftelecytologyusingroboticdigitalmicroscopeandsinglezstackdigitalscanforfineneedleaspirationrapidonsiteevaluation AT parwanianil comprehensivestudyoftelecytologyusingroboticdigitalmicroscopeandsinglezstackdigitalscanforfineneedleaspirationrapidonsiteevaluation AT lizaibo comprehensivestudyoftelecytologyusingroboticdigitalmicroscopeandsinglezstackdigitalscanforfineneedleaspirationrapidonsiteevaluation |