Validation of a bioelectrical impedance analysis system for body composition assessment in patients with COPD
OBJECTIVE: To investigate the validity of an eight-contact electrode bioelectrical impedance analysis (BIA) system within a household scale for assessing whole body composition in COPD patients. METHODS: Seventeen patients with COPD (mean age = 67 ± 8 years; mean FEV(1) = 38.6 ± 16.1% of predicted;...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Sociedade Brasileira de Pneumologia e Tisiologia
2018
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6326721/ https://www.ncbi.nlm.nih.gov/pubmed/30328930 http://dx.doi.org/10.1590/S1806-37562017000000121 |
Sumario: | OBJECTIVE: To investigate the validity of an eight-contact electrode bioelectrical impedance analysis (BIA) system within a household scale for assessing whole body composition in COPD patients. METHODS: Seventeen patients with COPD (mean age = 67 ± 8 years; mean FEV(1) = 38.6 ± 16.1% of predicted; and mean body mass index = 24.7 ± 5.4 kg/m(2)) underwent dual-energy X-ray absorptiometry (DEXA) and an eight-contact electrode BIA system for body composition assessment. RESULTS: There was a strong inter-method correlation for fat mass (r = 0.95), fat-free mass (r = 0.93), and lean mass (r = 0.93), but the correlation was moderate for bone mineral content (r = 0.73; p < 0.01 for all). In the agreement analysis, the values between DEXA and the BIA system differed by only 0.15 kg (−6.39 to 6.70 kg), 0.26 kg (−5.96 to 6.49 kg), −0.13 kg (−0.76 to 0.50 kg), and −0.55 kg (−6.71 to 5.61 kg) for fat-free mass, lean mass, bone mineral content, and fat mass, respectively. CONCLUSIONS: The eight-contact electrode BIA system showed to be a valid tool in the assessment of whole body composition in our sample of patients with COPD. |
---|