A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability

PURPOSE: This study is to compare the accuracy of four different black-blood T(2) mapping sequences in carotid vessel wall. METHODS: Four different black-blood T(2) mapping sequences were developed and tested through phantom experiments and 17 healthy volunteers. The four sequences were: 1) double i...

Descripción completa

Detalles Bibliográficos
Autores principales: Yuan, Jianmin, Patterson, Andrew J., Ruetten, Pascal P. R., Reid, Scott A., Gillard, Jonathan H., Graves, Martin J.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Japanese Society for Magnetic Resonance in Medicine 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6326764/
https://www.ncbi.nlm.nih.gov/pubmed/29515084
http://dx.doi.org/10.2463/mrms.mp.2017-0141
_version_ 1783386362088521728
author Yuan, Jianmin
Patterson, Andrew J.
Ruetten, Pascal P. R.
Reid, Scott A.
Gillard, Jonathan H.
Graves, Martin J.
author_facet Yuan, Jianmin
Patterson, Andrew J.
Ruetten, Pascal P. R.
Reid, Scott A.
Gillard, Jonathan H.
Graves, Martin J.
author_sort Yuan, Jianmin
collection PubMed
description PURPOSE: This study is to compare the accuracy of four different black-blood T(2) mapping sequences in carotid vessel wall. METHODS: Four different black-blood T(2) mapping sequences were developed and tested through phantom experiments and 17 healthy volunteers. The four sequences were: 1) double inversion-recovery (DIR) prepared 2D multi-echo spin-echo (MESE); 2) DIR-prepared 2D multi-echo fast spin-echo (MEFSE); 3) improved motion-sensitized driven-equilibrium (iMSDE) prepared 3D FSE and 4) iMSDE prepared 3D fast spoiled gradient echo (FSPGR). The concordance correlation coefficient and Bland–Altman statistics were used to compare the sequences with a gold-standard 2D MESE, without blood suppression in phantom studies. The volunteers were scanned twice to test the repeatability. Mean and standard deviation of vessel wall T(2), signal-to-noise (SNR), the coefficient of variance and interclass coefficient (ICC) of the two scans were compared. RESULTS: The phantom study demonstrated that T(2) measurements had high concordance with respect to the gold-standard (all r values >0.9). In the volunteer study, the DIR 2D MEFSE had significantly higher T(2) values than the other three sequences (P < 0.01). There was no difference in T(2) measurements obtained using the other three sequences (P > 0.05). iMSDE 3D FSE had the highest SNR (P < 0.05) compared with the other three sequences. The 2D DIR MESE has the highest repeatability (ICC: 0.96, [95% CI: 0.88–0.99]). CONCLUSION: Although accurate T(2) measurements can be achieved in phantom by the four sequences, in vivo vessel wall T(2) quantification shows significant differences. The in vivo images can be influenced by multiple factors including black-blood preparation and acquisition method. Therefore, a careful choice of acquisition methods and analysis of the confounding factors are required for accurate in vivo carotid vessel wall T(2) measurements. From the settings in this study, the iMSDE prepared 3D FSE is preferred for the future volunteer/patient scans.
format Online
Article
Text
id pubmed-6326764
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Japanese Society for Magnetic Resonance in Medicine
record_format MEDLINE/PubMed
spelling pubmed-63267642019-01-11 A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability Yuan, Jianmin Patterson, Andrew J. Ruetten, Pascal P. R. Reid, Scott A. Gillard, Jonathan H. Graves, Martin J. Magn Reson Med Sci Major Paper PURPOSE: This study is to compare the accuracy of four different black-blood T(2) mapping sequences in carotid vessel wall. METHODS: Four different black-blood T(2) mapping sequences were developed and tested through phantom experiments and 17 healthy volunteers. The four sequences were: 1) double inversion-recovery (DIR) prepared 2D multi-echo spin-echo (MESE); 2) DIR-prepared 2D multi-echo fast spin-echo (MEFSE); 3) improved motion-sensitized driven-equilibrium (iMSDE) prepared 3D FSE and 4) iMSDE prepared 3D fast spoiled gradient echo (FSPGR). The concordance correlation coefficient and Bland–Altman statistics were used to compare the sequences with a gold-standard 2D MESE, without blood suppression in phantom studies. The volunteers were scanned twice to test the repeatability. Mean and standard deviation of vessel wall T(2), signal-to-noise (SNR), the coefficient of variance and interclass coefficient (ICC) of the two scans were compared. RESULTS: The phantom study demonstrated that T(2) measurements had high concordance with respect to the gold-standard (all r values >0.9). In the volunteer study, the DIR 2D MEFSE had significantly higher T(2) values than the other three sequences (P < 0.01). There was no difference in T(2) measurements obtained using the other three sequences (P > 0.05). iMSDE 3D FSE had the highest SNR (P < 0.05) compared with the other three sequences. The 2D DIR MESE has the highest repeatability (ICC: 0.96, [95% CI: 0.88–0.99]). CONCLUSION: Although accurate T(2) measurements can be achieved in phantom by the four sequences, in vivo vessel wall T(2) quantification shows significant differences. The in vivo images can be influenced by multiple factors including black-blood preparation and acquisition method. Therefore, a careful choice of acquisition methods and analysis of the confounding factors are required for accurate in vivo carotid vessel wall T(2) measurements. From the settings in this study, the iMSDE prepared 3D FSE is preferred for the future volunteer/patient scans. Japanese Society for Magnetic Resonance in Medicine 2018-03-08 /pmc/articles/PMC6326764/ /pubmed/29515084 http://dx.doi.org/10.2463/mrms.mp.2017-0141 Text en © 2018 Japanese Society for Magnetic Resonance in Medicine This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. To view a copy of this license, visit http://creativecommons.org/licenses/by-nc-nd/4.0/
spellingShingle Major Paper
Yuan, Jianmin
Patterson, Andrew J.
Ruetten, Pascal P. R.
Reid, Scott A.
Gillard, Jonathan H.
Graves, Martin J.
A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability
title A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability
title_full A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability
title_fullStr A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability
title_full_unstemmed A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability
title_short A Comparison of Black-blood T(2) Mapping Sequences for Carotid Vessel Wall Imaging at 3T: An Assessment of Accuracy and Repeatability
title_sort comparison of black-blood t(2) mapping sequences for carotid vessel wall imaging at 3t: an assessment of accuracy and repeatability
topic Major Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6326764/
https://www.ncbi.nlm.nih.gov/pubmed/29515084
http://dx.doi.org/10.2463/mrms.mp.2017-0141
work_keys_str_mv AT yuanjianmin acomparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT pattersonandrewj acomparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT ruettenpascalpr acomparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT reidscotta acomparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT gillardjonathanh acomparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT gravesmartinj acomparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT yuanjianmin comparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT pattersonandrewj comparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT ruettenpascalpr comparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT reidscotta comparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT gillardjonathanh comparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability
AT gravesmartinj comparisonofblackbloodt2mappingsequencesforcarotidvesselwallimagingat3tanassessmentofaccuracyandrepeatability