Cargando…

Comparison of Color Light-Emitting Diode Corneal Topographer and Dual Rotating Scheimpflug–Placido Topographer

PURPOSE: To compare measurements of anterior and posterior keratometric values, using a color light-emitting diode corneal topographer and a dual rotating Scheimpflug–Placido topographer. METHODS: Anterior and posterior corneal index measurements were performed using a color light-emitting diode cor...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Jae Hyuck, Lee, Yong Woo, Lee, Jong Soo, Knorz, Michael C., Auffarth, Gerd U., Choi, Chul Young
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6327256/
https://www.ncbi.nlm.nih.gov/pubmed/30687546
http://dx.doi.org/10.1155/2018/6764805
Descripción
Sumario:PURPOSE: To compare measurements of anterior and posterior keratometric values, using a color light-emitting diode corneal topographer and a dual rotating Scheimpflug–Placido topographer. METHODS: Anterior and posterior corneal index measurements were performed using a color light-emitting diode corneal topographer (Cassini) and a dual rotating Scheimpflug–Placido topographer (Galilei G4) and then compared. The paired t-test, intraclass correlation coefficients (ICCs), and Bland–Altman plots were used to evaluate the agreement between measurements. RESULTS: Sixty postrefractive surgery eyes and 60 normal eyes were evaluated. Both the color light-emitting diode corneal topographer and the dual rotating Scheimpflug–Placido topographer provided highly repeatable corneal measurements (ICC > 0.969). The agreement levels between the 2 devices for anterior corneal power, astigmatism magnitude, and J0 and J45 values were ICC > 0.906 for the total group. However, the ICC values for posterior corneal power, astigmatism magnitude, and J0 and J45 values were lower than 0.681 for the total group. CONCLUSIONS: The anterior keratometric values obtained by the color light-emitting diode corneal topographer and the dual rotating Scheimpflug–Placido topographer showed high agreement levels, but the posterior keratometric values showed lower agreement levels.