Cargando…
Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study
BACKGROUND: Despite the high survival rate of dental implants, screw loosening is frequently reported. Screw loosening can cause a misfit of the implant-abutment connection leading to peri-implantitis or abutment screw fracture. Therefore, studies about related factors and mechanism of screw looseni...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6330554/ https://www.ncbi.nlm.nih.gov/pubmed/30637611 http://dx.doi.org/10.1186/s40729-018-0154-2 |
_version_ | 1783387001104367616 |
---|---|
author | Benjaboonyazit, Kanyarin Chaijareenont, Pisaisit Khongkhunthian, Pathawee |
author_facet | Benjaboonyazit, Kanyarin Chaijareenont, Pisaisit Khongkhunthian, Pathawee |
author_sort | Benjaboonyazit, Kanyarin |
collection | PubMed |
description | BACKGROUND: Despite the high survival rate of dental implants, screw loosening is frequently reported. Screw loosening can cause a misfit of the implant-abutment connection leading to peri-implantitis or abutment screw fracture. Therefore, studies about related factors and mechanism of screw loosening are needed. The aim of this study was to evaluate the decreasing pattern of removal torque values (RTVs) of a combined cone and octalobule index implant-abutment connection under different numbers of mechanical loading cycles. MATERIALS AND METHODS: The study was performed in accordance with ISO 14801:2007. Eighty-four implants with the combined cone and octalobule index implant-abutment connection (PW Plus dental implant system, PW Plus Company) were used. All abutment screws were tightened 30 N cm twice with a 10-min interval. The control group was without cyclic loading and the experimental groups underwent different numbers of loading cycles until 2,000,000 cycles. Then, the abutment screws of all samples were untightened to measure the RTVs. The data were analyzed using ANOVA and Tukey’s HSD test. RESULTS: The RTVs of the control group decreased 7.78% compared to the insertion torque. All experimental groups from 50,000 to 2,000,000 cycles showed significant decreases in RTVs compared to the control group (P < 0.05). RTVs in the group of 50,000 cycles to 1,800,000 cycles did not change significantly, but there was a significant reduction of RTVs in the group of 2,000,000 cycles when compared to the group of 50,000 cycles (P < 0.05). CONCLUSIONS: According to the setting condition for the fatigue test complied to ISO 14801:2007, the RTVs of the combined cone and octalobule index implant-abutment connection reduced significantly after 50,000 cycles and did not change significantly until 2,000,000 cycles. |
format | Online Article Text |
id | pubmed-6330554 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-63305542019-01-27 Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study Benjaboonyazit, Kanyarin Chaijareenont, Pisaisit Khongkhunthian, Pathawee Int J Implant Dent Research BACKGROUND: Despite the high survival rate of dental implants, screw loosening is frequently reported. Screw loosening can cause a misfit of the implant-abutment connection leading to peri-implantitis or abutment screw fracture. Therefore, studies about related factors and mechanism of screw loosening are needed. The aim of this study was to evaluate the decreasing pattern of removal torque values (RTVs) of a combined cone and octalobule index implant-abutment connection under different numbers of mechanical loading cycles. MATERIALS AND METHODS: The study was performed in accordance with ISO 14801:2007. Eighty-four implants with the combined cone and octalobule index implant-abutment connection (PW Plus dental implant system, PW Plus Company) were used. All abutment screws were tightened 30 N cm twice with a 10-min interval. The control group was without cyclic loading and the experimental groups underwent different numbers of loading cycles until 2,000,000 cycles. Then, the abutment screws of all samples were untightened to measure the RTVs. The data were analyzed using ANOVA and Tukey’s HSD test. RESULTS: The RTVs of the control group decreased 7.78% compared to the insertion torque. All experimental groups from 50,000 to 2,000,000 cycles showed significant decreases in RTVs compared to the control group (P < 0.05). RTVs in the group of 50,000 cycles to 1,800,000 cycles did not change significantly, but there was a significant reduction of RTVs in the group of 2,000,000 cycles when compared to the group of 50,000 cycles (P < 0.05). CONCLUSIONS: According to the setting condition for the fatigue test complied to ISO 14801:2007, the RTVs of the combined cone and octalobule index implant-abutment connection reduced significantly after 50,000 cycles and did not change significantly until 2,000,000 cycles. Springer Berlin Heidelberg 2019-01-14 /pmc/articles/PMC6330554/ /pubmed/30637611 http://dx.doi.org/10.1186/s40729-018-0154-2 Text en © The Author(s). 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Research Benjaboonyazit, Kanyarin Chaijareenont, Pisaisit Khongkhunthian, Pathawee Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study |
title | Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study |
title_full | Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study |
title_fullStr | Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study |
title_full_unstemmed | Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study |
title_short | Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study |
title_sort | removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6330554/ https://www.ncbi.nlm.nih.gov/pubmed/30637611 http://dx.doi.org/10.1186/s40729-018-0154-2 |
work_keys_str_mv | AT benjaboonyazitkanyarin removaltorquepatternofacombinedconeandoctalobuleindeximplantabutmentconnectionatdifferentcyclicloadinganinvitroexperimentalstudy AT chaijareenontpisaisit removaltorquepatternofacombinedconeandoctalobuleindeximplantabutmentconnectionatdifferentcyclicloadinganinvitroexperimentalstudy AT khongkhunthianpathawee removaltorquepatternofacombinedconeandoctalobuleindeximplantabutmentconnectionatdifferentcyclicloadinganinvitroexperimentalstudy |