Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro

Purpose: To investigate in an in vitro model if secondary endobag filling can reduce gutter size during chimney endovascular aneurysm sealing (chEVAS). Materials and Methods: Nellix EVAS systems were deployed in 2 silicone juxtarenal aneurysm models with suprarenal aortic diameters of 19 and 24 mm....

Descripción completa

Detalles Bibliográficos
Autores principales: van Schaik, Theodorus G., Meekel, Jorn P., Jongkind, Vincent, Lely, Rutger J., Truijers, Maarten, Hoksbergen, Arjan W. J., Wisselink, Willem, Blankensteijn, Jan D., Yeung, Kak Khee
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6330694/
https://www.ncbi.nlm.nih.gov/pubmed/30572773
http://dx.doi.org/10.1177/1526602818819494
_version_ 1783387015089225728
author van Schaik, Theodorus G.
Meekel, Jorn P.
Jongkind, Vincent
Lely, Rutger J.
Truijers, Maarten
Hoksbergen, Arjan W. J.
Wisselink, Willem
Blankensteijn, Jan D.
Yeung, Kak Khee
author_facet van Schaik, Theodorus G.
Meekel, Jorn P.
Jongkind, Vincent
Lely, Rutger J.
Truijers, Maarten
Hoksbergen, Arjan W. J.
Wisselink, Willem
Blankensteijn, Jan D.
Yeung, Kak Khee
author_sort van Schaik, Theodorus G.
collection PubMed
description Purpose: To investigate in an in vitro model if secondary endobag filling can reduce gutter size during chimney endovascular aneurysm sealing (chEVAS). Materials and Methods: Nellix EVAS systems were deployed in 2 silicone juxtarenal aneurysm models with suprarenal aortic diameters of 19 and 24 mm. Four configurations were tested: EVAS with 6-mm balloon-expandable (BE) or self-expanding (SE) chimney grafts (CGs) in the renal branches of both models. Balloons were inflated simultaneously in the CGs and main endografts during primary and secondary endobag filling and polymer curing. Computed tomography (CT) was performed immediately after the primary and secondary fills. Cross-sectional lumen areas were measured on the CT images to calculate gutter volumes and percent change. CG compression was calculated as the reduction in lumen surface area measured perpendicular to the central lumen line. The largest gutter volume and highest compression were presented per CG configuration per model. Results: Secondary endobag filling reduced the largest gutter volumes from 99.4 to 73.1 mm(3) (13.2% change) and 84.2 to 72.0 mm(3) (27.6% change) in the BECG configurations and from 67.2 to 44.0 mm(3) (34.5% change) and 92.7 to 82.3 mm(3) (11.2% change) in the SECG configurations in the 19- and 24-mm models, respectively. Secondary endobag filling increased CG compression in 6 of 8 configurations. BECG compression changed by −0.2% and 5.4% and by −1.0% and 0.4% in the 19- and 24-mm models, respectively. SECG compression changed by 10.2% and 16.0% and by 7.2% and 7.3% in the 19- and 24-mm models, respectively. Conclusion: Secondary endobag filling reduced paragraft gutters; however, this technique did not obliterate them. Increased CG compression and prolonged renal ischemia time should be considered if secondary endobag filling is used.
format Online
Article
Text
id pubmed-6330694
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-63306942019-01-29 Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro van Schaik, Theodorus G. Meekel, Jorn P. Jongkind, Vincent Lely, Rutger J. Truijers, Maarten Hoksbergen, Arjan W. J. Wisselink, Willem Blankensteijn, Jan D. Yeung, Kak Khee J Endovasc Ther Endovascular Aneurysm Sealing Purpose: To investigate in an in vitro model if secondary endobag filling can reduce gutter size during chimney endovascular aneurysm sealing (chEVAS). Materials and Methods: Nellix EVAS systems were deployed in 2 silicone juxtarenal aneurysm models with suprarenal aortic diameters of 19 and 24 mm. Four configurations were tested: EVAS with 6-mm balloon-expandable (BE) or self-expanding (SE) chimney grafts (CGs) in the renal branches of both models. Balloons were inflated simultaneously in the CGs and main endografts during primary and secondary endobag filling and polymer curing. Computed tomography (CT) was performed immediately after the primary and secondary fills. Cross-sectional lumen areas were measured on the CT images to calculate gutter volumes and percent change. CG compression was calculated as the reduction in lumen surface area measured perpendicular to the central lumen line. The largest gutter volume and highest compression were presented per CG configuration per model. Results: Secondary endobag filling reduced the largest gutter volumes from 99.4 to 73.1 mm(3) (13.2% change) and 84.2 to 72.0 mm(3) (27.6% change) in the BECG configurations and from 67.2 to 44.0 mm(3) (34.5% change) and 92.7 to 82.3 mm(3) (11.2% change) in the SECG configurations in the 19- and 24-mm models, respectively. Secondary endobag filling increased CG compression in 6 of 8 configurations. BECG compression changed by −0.2% and 5.4% and by −1.0% and 0.4% in the 19- and 24-mm models, respectively. SECG compression changed by 10.2% and 16.0% and by 7.2% and 7.3% in the 19- and 24-mm models, respectively. Conclusion: Secondary endobag filling reduced paragraft gutters; however, this technique did not obliterate them. Increased CG compression and prolonged renal ischemia time should be considered if secondary endobag filling is used. SAGE Publications 2018-12-21 2019-02 /pmc/articles/PMC6330694/ /pubmed/30572773 http://dx.doi.org/10.1177/1526602818819494 Text en © The Author(s) 2018 http://www.creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Endovascular Aneurysm Sealing
van Schaik, Theodorus G.
Meekel, Jorn P.
Jongkind, Vincent
Lely, Rutger J.
Truijers, Maarten
Hoksbergen, Arjan W. J.
Wisselink, Willem
Blankensteijn, Jan D.
Yeung, Kak Khee
Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro
title Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro
title_full Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro
title_fullStr Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro
title_full_unstemmed Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro
title_short Secondary Fill Minimizes Gutter Size in Chimney EVAS Configurations In Vitro
title_sort secondary fill minimizes gutter size in chimney evas configurations in vitro
topic Endovascular Aneurysm Sealing
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6330694/
https://www.ncbi.nlm.nih.gov/pubmed/30572773
http://dx.doi.org/10.1177/1526602818819494
work_keys_str_mv AT vanschaiktheodorusg secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT meekeljornp secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT jongkindvincent secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT lelyrutgerj secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT truijersmaarten secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT hoksbergenarjanwj secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT wisselinkwillem secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT blankensteijnjand secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro
AT yeungkakkhee secondaryfillminimizesguttersizeinchimneyevasconfigurationsinvitro