Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device
BACKGROUND: Respiratory rate is among the first vital signs to change in deteriorating patients. The aims of this study were to evaluate the accuracy of respiratory rate measurements using a specifically dedicated reflection-mode photoplethysmographic signal analysis in a pathological condition (PPG...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6340913/ https://www.ncbi.nlm.nih.gov/pubmed/30666472 http://dx.doi.org/10.1186/s13613-019-0485-z |
_version_ | 1783388854108028928 |
---|---|
author | L’Her, Erwan N’Guyen, Quang-Thang Pateau, Victoire Bodenes, Laetitia Lellouche, François |
author_facet | L’Her, Erwan N’Guyen, Quang-Thang Pateau, Victoire Bodenes, Laetitia Lellouche, François |
author_sort | L’Her, Erwan |
collection | PubMed |
description | BACKGROUND: Respiratory rate is among the first vital signs to change in deteriorating patients. The aims of this study were to evaluate the accuracy of respiratory rate measurements using a specifically dedicated reflection-mode photoplethysmographic signal analysis in a pathological condition (PPG-RR) and to validate its implementation within medical devices. METHODS: This study is derived from a data mining project, including all consecutive patients admitted to our ICU (ReaSTOC study, ClinicalTrials.gov identifier: NCT02893462). During the evaluation phase of the algorithm, PPG-RR calculations were retrospectively performed on PPG waveforms extracted from the data warehouse and compared with RR reference values. During the prospective phase, PPG-RR calculations were automatically and continuously performed using a dedicated device (FreeO(2), Oxynov, Québec, QC, Canada). In all phases, reference RR was measured continuously using electrical thoracic impedance and chronometric evaluation (Manual-RR) over a 30-s period. RESULTS: In total, 201 ICU patients’ recordings (SAPS II 51.7 ± 34.6) were analysed during the retrospective evaluation phase, most of them being admitted for a respiratory failure and requiring invasive mechanical ventilation. PPG-RR determination was available in 95.5% cases, similar to reference (22 ± 4 vs. 22 ± 5 c/min, respectively; p = 1), and well correlated with reference values (R = 0.952; p < 0.0001), with a low bias (0.1 b/min) and deviation (± 3.5 b/min). Prospective estimation of the PPG-RR on 30 ICU patients’ recordings was well correlated with the reference method (Manual-RR; r = 0.78; p < 0.001). Comparison of the methods depicted a low bias (0.5 b/min) and acceptable deviation (< ± 5.5 b/min). CONCLUSION: According to our results, PPG-RR is an interesting approach for ventilation monitoring, as this technique would make simultaneous monitoring of respiratory rate and arterial oxygen saturation possible, thus minimizing the number of sensors attached to the patient. Trial registry number ClinicalTrials.gov identifier NCT02893462 |
format | Online Article Text |
id | pubmed-6340913 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-63409132019-02-07 Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device L’Her, Erwan N’Guyen, Quang-Thang Pateau, Victoire Bodenes, Laetitia Lellouche, François Ann Intensive Care Research BACKGROUND: Respiratory rate is among the first vital signs to change in deteriorating patients. The aims of this study were to evaluate the accuracy of respiratory rate measurements using a specifically dedicated reflection-mode photoplethysmographic signal analysis in a pathological condition (PPG-RR) and to validate its implementation within medical devices. METHODS: This study is derived from a data mining project, including all consecutive patients admitted to our ICU (ReaSTOC study, ClinicalTrials.gov identifier: NCT02893462). During the evaluation phase of the algorithm, PPG-RR calculations were retrospectively performed on PPG waveforms extracted from the data warehouse and compared with RR reference values. During the prospective phase, PPG-RR calculations were automatically and continuously performed using a dedicated device (FreeO(2), Oxynov, Québec, QC, Canada). In all phases, reference RR was measured continuously using electrical thoracic impedance and chronometric evaluation (Manual-RR) over a 30-s period. RESULTS: In total, 201 ICU patients’ recordings (SAPS II 51.7 ± 34.6) were analysed during the retrospective evaluation phase, most of them being admitted for a respiratory failure and requiring invasive mechanical ventilation. PPG-RR determination was available in 95.5% cases, similar to reference (22 ± 4 vs. 22 ± 5 c/min, respectively; p = 1), and well correlated with reference values (R = 0.952; p < 0.0001), with a low bias (0.1 b/min) and deviation (± 3.5 b/min). Prospective estimation of the PPG-RR on 30 ICU patients’ recordings was well correlated with the reference method (Manual-RR; r = 0.78; p < 0.001). Comparison of the methods depicted a low bias (0.5 b/min) and acceptable deviation (< ± 5.5 b/min). CONCLUSION: According to our results, PPG-RR is an interesting approach for ventilation monitoring, as this technique would make simultaneous monitoring of respiratory rate and arterial oxygen saturation possible, thus minimizing the number of sensors attached to the patient. Trial registry number ClinicalTrials.gov identifier NCT02893462 Springer International Publishing 2019-01-21 /pmc/articles/PMC6340913/ /pubmed/30666472 http://dx.doi.org/10.1186/s13613-019-0485-z Text en © The Author(s) 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Research L’Her, Erwan N’Guyen, Quang-Thang Pateau, Victoire Bodenes, Laetitia Lellouche, François Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device |
title | Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device |
title_full | Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device |
title_fullStr | Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device |
title_full_unstemmed | Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device |
title_short | Photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device |
title_sort | photoplethysmographic determination of the respiratory rate in acutely ill patients: validation of a new algorithm and implementation into a biomedical device |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6340913/ https://www.ncbi.nlm.nih.gov/pubmed/30666472 http://dx.doi.org/10.1186/s13613-019-0485-z |
work_keys_str_mv | AT lhererwan photoplethysmographicdeterminationoftherespiratoryrateinacutelyillpatientsvalidationofanewalgorithmandimplementationintoabiomedicaldevice AT nguyenquangthang photoplethysmographicdeterminationoftherespiratoryrateinacutelyillpatientsvalidationofanewalgorithmandimplementationintoabiomedicaldevice AT pateauvictoire photoplethysmographicdeterminationoftherespiratoryrateinacutelyillpatientsvalidationofanewalgorithmandimplementationintoabiomedicaldevice AT bodeneslaetitia photoplethysmographicdeterminationoftherespiratoryrateinacutelyillpatientsvalidationofanewalgorithmandimplementationintoabiomedicaldevice AT lellouchefrancois photoplethysmographicdeterminationoftherespiratoryrateinacutelyillpatientsvalidationofanewalgorithmandimplementationintoabiomedicaldevice |