Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA

BACKGROUND: Aseptic loosening of the tibial component remains a major cause of failure in unicompartmental knee arthroplasty (UKA) and may be related to micro-motion at the cement-bone interface due to insufficient cement penetration depth. Cement penetration is therefore taken as an indicator of so...

Descripción completa

Detalles Bibliográficos
Autores principales: Scheele, Christian B., Müller, Peter E., Schröder, Christian, Grupp, Thomas, Jansson, Volkmar, Pietschmann, Matthias F.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6341644/
https://www.ncbi.nlm.nih.gov/pubmed/30665370
http://dx.doi.org/10.1186/s12880-019-0312-x
_version_ 1783388983313563648
author Scheele, Christian B.
Müller, Peter E.
Schröder, Christian
Grupp, Thomas
Jansson, Volkmar
Pietschmann, Matthias F.
author_facet Scheele, Christian B.
Müller, Peter E.
Schröder, Christian
Grupp, Thomas
Jansson, Volkmar
Pietschmann, Matthias F.
author_sort Scheele, Christian B.
collection PubMed
description BACKGROUND: Aseptic loosening of the tibial component remains a major cause of failure in unicompartmental knee arthroplasty (UKA) and may be related to micro-motion at the cement-bone interface due to insufficient cement penetration depth. Cement penetration is therefore taken as an indicator of solid fixation strength and primary stability. However, its non-invasive clinical assessment remains difficult in vivo as conventional x-ray is prone to distortion and CT-scans (computed tomography) are difficult to assess due to metal artifacts. The purpose of this study was to develop and validate a reliable in vivo measuring technique of cement penetration depth in human tibial UKA. METHODS: In an experimental setting, twelve UKA were implanted in fresh-frozen human cadaver knees using a minimal-invasive medial approach. Cement penetration depth was then measured via 1) virtual 3D-models based on metal artifact reduced CT-scans and 2) histological evaluation of nine serial cross-section cuts through the implant-cement-bone-interface. Subsequently, a concordance analysis between the two measuring techniques was conducted. RESULTS: The average cement penetration depth was 1) 2.20 mm (SD 0.30 mm) measured on metal artifact reduced CT-scans and 2) 2.21 mm (SD = 0.42) measured on serial cuts (p = 0.956). The mean difference between both techniques was 0.01 mm (SD 0.31 mm) and the Person correlation coefficient was r = 0.686 (p = 0.014). All differences were within the upper and lower limit of agreement. There was no evidence of any significant proportional bias between both techniques (p = 0.182). CONCLUSIONS: CT-based non-invasive measurement of cement penetration depth delivers reliable results in measuring the penetration depth in tibial UKA. Thereby, it enables clinicians and researchers to assess the cement penetration for in vivo diagnostics in the clinical setting as well as in vitro biomechanical research with subsequent application of load to failure on the implant-cement-bone-interface.
format Online
Article
Text
id pubmed-6341644
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-63416442019-01-24 Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA Scheele, Christian B. Müller, Peter E. Schröder, Christian Grupp, Thomas Jansson, Volkmar Pietschmann, Matthias F. BMC Med Imaging Research Article BACKGROUND: Aseptic loosening of the tibial component remains a major cause of failure in unicompartmental knee arthroplasty (UKA) and may be related to micro-motion at the cement-bone interface due to insufficient cement penetration depth. Cement penetration is therefore taken as an indicator of solid fixation strength and primary stability. However, its non-invasive clinical assessment remains difficult in vivo as conventional x-ray is prone to distortion and CT-scans (computed tomography) are difficult to assess due to metal artifacts. The purpose of this study was to develop and validate a reliable in vivo measuring technique of cement penetration depth in human tibial UKA. METHODS: In an experimental setting, twelve UKA were implanted in fresh-frozen human cadaver knees using a minimal-invasive medial approach. Cement penetration depth was then measured via 1) virtual 3D-models based on metal artifact reduced CT-scans and 2) histological evaluation of nine serial cross-section cuts through the implant-cement-bone-interface. Subsequently, a concordance analysis between the two measuring techniques was conducted. RESULTS: The average cement penetration depth was 1) 2.20 mm (SD 0.30 mm) measured on metal artifact reduced CT-scans and 2) 2.21 mm (SD = 0.42) measured on serial cuts (p = 0.956). The mean difference between both techniques was 0.01 mm (SD 0.31 mm) and the Person correlation coefficient was r = 0.686 (p = 0.014). All differences were within the upper and lower limit of agreement. There was no evidence of any significant proportional bias between both techniques (p = 0.182). CONCLUSIONS: CT-based non-invasive measurement of cement penetration depth delivers reliable results in measuring the penetration depth in tibial UKA. Thereby, it enables clinicians and researchers to assess the cement penetration for in vivo diagnostics in the clinical setting as well as in vitro biomechanical research with subsequent application of load to failure on the implant-cement-bone-interface. BioMed Central 2019-01-21 /pmc/articles/PMC6341644/ /pubmed/30665370 http://dx.doi.org/10.1186/s12880-019-0312-x Text en © The Author(s). 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Scheele, Christian B.
Müller, Peter E.
Schröder, Christian
Grupp, Thomas
Jansson, Volkmar
Pietschmann, Matthias F.
Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA
title Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA
title_full Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA
title_fullStr Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA
title_full_unstemmed Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA
title_short Accuracy of a non-invasive CT-based measuring technique for cement penetration depth in human tibial UKA
title_sort accuracy of a non-invasive ct-based measuring technique for cement penetration depth in human tibial uka
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6341644/
https://www.ncbi.nlm.nih.gov/pubmed/30665370
http://dx.doi.org/10.1186/s12880-019-0312-x
work_keys_str_mv AT scheelechristianb accuracyofanoninvasivectbasedmeasuringtechniqueforcementpenetrationdepthinhumantibialuka
AT mullerpetere accuracyofanoninvasivectbasedmeasuringtechniqueforcementpenetrationdepthinhumantibialuka
AT schroderchristian accuracyofanoninvasivectbasedmeasuringtechniqueforcementpenetrationdepthinhumantibialuka
AT gruppthomas accuracyofanoninvasivectbasedmeasuringtechniqueforcementpenetrationdepthinhumantibialuka
AT janssonvolkmar accuracyofanoninvasivectbasedmeasuringtechniqueforcementpenetrationdepthinhumantibialuka
AT pietschmannmatthiasf accuracyofanoninvasivectbasedmeasuringtechniqueforcementpenetrationdepthinhumantibialuka