Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel

INTRODUCTION: Operating rooms in hospitals are facilitated with different types of electronic systems, which produce electromagnetic waves. High intensities of magnetic waves may have harmful effects on biological environments. This study aims to evaluate the electromagnetic field intensity at diffe...

Descripción completa

Detalles Bibliográficos
Autores principales: Ghazikhanlou-Sani, Karim, Rahimi, Azizollah, Poorkaveh, Maryam, Eynali, Samira, Koosha, Fereshteh, Shoja, Mohsen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Akadémiai Kiadó 2018
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6343582/
https://www.ncbi.nlm.nih.gov/pubmed/30713749
http://dx.doi.org/10.1556/1646.10.2018.18
_version_ 1783389311384682496
author Ghazikhanlou-Sani, Karim
Rahimi, Azizollah
Poorkaveh, Maryam
Eynali, Samira
Koosha, Fereshteh
Shoja, Mohsen
author_facet Ghazikhanlou-Sani, Karim
Rahimi, Azizollah
Poorkaveh, Maryam
Eynali, Samira
Koosha, Fereshteh
Shoja, Mohsen
author_sort Ghazikhanlou-Sani, Karim
collection PubMed
description INTRODUCTION: Operating rooms in hospitals are facilitated with different types of electronic systems, which produce electromagnetic waves. High intensities of magnetic waves may have harmful effects on biological environments. This study aims to evaluate the electromagnetic field intensity at different parts of operating rooms at the first stage and estimate the occupational exposure to operating room personnel at the next phase. MATERIALS AND METHODS: At this cross-sectional study, the magnetic field intensity was evaluated using teslameter at several parts of operating rooms, during operating procedures, while electrical instruments were working. Background electromagnetic field intensity was measured when all the electrical systems were idle. Statistical analysis was performed using SPSS software. The results were compared with ICNIRP standards. RESULTS: The maximum intensity of magnetic field was measured around high-voltage systems at the distance of 50 cm in the personnel’s standing area at DCR and PCNL operating procedures were 5.9 and 5.6, respectively. The number of on-mode electrical systems was inconsistent with the intensity of electromagnetic fields at the standing area of operating room personnel’s. The intensity of magnetic fields around high-voltage systems, which was about 46.75 mG at the distance of 10 cm, was the highest among measured electromagnetic fields. CONCLUSIONS: The highest magnetic field intensity measured in this study was related to high-voltage systems and is lower than advised intensity by ICNIRP for occupational exposure. Based on this study, it can be concluded that there are no considerable risks of electromagnetic exposure for operating room personnels.
format Online
Article
Text
id pubmed-6343582
institution National Center for Biotechnology Information
language English
publishDate 2018
publisher Akadémiai Kiadó
record_format MEDLINE/PubMed
spelling pubmed-63435822019-02-01 Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel Ghazikhanlou-Sani, Karim Rahimi, Azizollah Poorkaveh, Maryam Eynali, Samira Koosha, Fereshteh Shoja, Mohsen Interv Med Appl Sci Original Paper INTRODUCTION: Operating rooms in hospitals are facilitated with different types of electronic systems, which produce electromagnetic waves. High intensities of magnetic waves may have harmful effects on biological environments. This study aims to evaluate the electromagnetic field intensity at different parts of operating rooms at the first stage and estimate the occupational exposure to operating room personnel at the next phase. MATERIALS AND METHODS: At this cross-sectional study, the magnetic field intensity was evaluated using teslameter at several parts of operating rooms, during operating procedures, while electrical instruments were working. Background electromagnetic field intensity was measured when all the electrical systems were idle. Statistical analysis was performed using SPSS software. The results were compared with ICNIRP standards. RESULTS: The maximum intensity of magnetic field was measured around high-voltage systems at the distance of 50 cm in the personnel’s standing area at DCR and PCNL operating procedures were 5.9 and 5.6, respectively. The number of on-mode electrical systems was inconsistent with the intensity of electromagnetic fields at the standing area of operating room personnel’s. The intensity of magnetic fields around high-voltage systems, which was about 46.75 mG at the distance of 10 cm, was the highest among measured electromagnetic fields. CONCLUSIONS: The highest magnetic field intensity measured in this study was related to high-voltage systems and is lower than advised intensity by ICNIRP for occupational exposure. Based on this study, it can be concluded that there are no considerable risks of electromagnetic exposure for operating room personnels. Akadémiai Kiadó 2018-07-25 2018-09 /pmc/articles/PMC6343582/ /pubmed/30713749 http://dx.doi.org/10.1556/1646.10.2018.18 Text en © 2018 The Author(s) http://creativecommons.org/licenses/by-nc/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://creativecommons.org/licenses/by-nc/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium for non-commercial purposes, provided the original author and source are credited, a link to the CC License is provided, and changes – if any – are indicated.
spellingShingle Original Paper
Ghazikhanlou-Sani, Karim
Rahimi, Azizollah
Poorkaveh, Maryam
Eynali, Samira
Koosha, Fereshteh
Shoja, Mohsen
Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel
title Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel
title_full Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel
title_fullStr Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel
title_full_unstemmed Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel
title_short Evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel
title_sort evaluation of the electromagnetic field intensity in operating rooms and estimation of occupational exposures of personnel
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6343582/
https://www.ncbi.nlm.nih.gov/pubmed/30713749
http://dx.doi.org/10.1556/1646.10.2018.18
work_keys_str_mv AT ghazikhanlousanikarim evaluationoftheelectromagneticfieldintensityinoperatingroomsandestimationofoccupationalexposuresofpersonnel
AT rahimiazizollah evaluationoftheelectromagneticfieldintensityinoperatingroomsandestimationofoccupationalexposuresofpersonnel
AT poorkavehmaryam evaluationoftheelectromagneticfieldintensityinoperatingroomsandestimationofoccupationalexposuresofpersonnel
AT eynalisamira evaluationoftheelectromagneticfieldintensityinoperatingroomsandestimationofoccupationalexposuresofpersonnel
AT kooshafereshteh evaluationoftheelectromagneticfieldintensityinoperatingroomsandestimationofoccupationalexposuresofpersonnel
AT shojamohsen evaluationoftheelectromagneticfieldintensityinoperatingroomsandestimationofoccupationalexposuresofpersonnel