High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers

PURPOSE: Objective of this study was to evaluate the effectiveness of the iterative reconstruction of high-pitch dual-source chest CT (IR-HP-CT) scanned with low radiation exposure compared with low dose chest CT (LDCT). MATERIALS AND METHODS: This study was approved by the institutional review boar...

Descripción completa

Detalles Bibliográficos
Autores principales: Lim, Hyun Kyung, Ha, Hong Il, Hwang, Hye Jeon, Lee, Kwanseop
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6345490/
https://www.ncbi.nlm.nih.gov/pubmed/30677082
http://dx.doi.org/10.1371/journal.pone.0211097
_version_ 1783389579826429952
author Lim, Hyun Kyung
Ha, Hong Il
Hwang, Hye Jeon
Lee, Kwanseop
author_facet Lim, Hyun Kyung
Ha, Hong Il
Hwang, Hye Jeon
Lee, Kwanseop
author_sort Lim, Hyun Kyung
collection PubMed
description PURPOSE: Objective of this study was to evaluate the effectiveness of the iterative reconstruction of high-pitch dual-source chest CT (IR-HP-CT) scanned with low radiation exposure compared with low dose chest CT (LDCT). MATERIALS AND METHODS: This study was approved by the institutional review board. Thirty healthy adult volunteers (mean age 44 years) were enrolled in this study. All volunteers underwent both IR-HP-CT and LDCT. IR-HP-CT was scanned with 120 kVp tube voltage, 30 mAs tube current and pitch 3.2 and reconstructed with sinogram affirmed iterative reconstruction. LDCT was scanned with 120 kVp tube voltage, 40 mAs tube current and pitch 0.8 and reconstructed with B50 filtered back projection. Image noise, and signal to noise ratio (SNR) of the infraspinatus muscle, subcutaneous fat and lung parenchyma were calculated. Cardiac motion artifact, overall image quality and artifacts was rated by two blinded readers using 4-point scale. The dose-length product (DLP) (mGy∙cm) were obtained from each CT dosimetry table. Scan length was calculated from the DLP results. The DLP parameter was a metric of radiation output, not of patient dose. Size-specific dose estimation (SSDE, mGy) was calculated using the sum of the anteroposterior and lateral dimensions and effective radiation dose (ED, mSv) were calculated using CT dosimetry index. RESULTS: Approximately, mean 40% of SSDE (2.1 ± 0.2 mGy vs. 3.5 ± 0.3 mGy) and 34% of ED (1.0 ± 0.1 mSv vs. 1.5 ± 0.1 mSv) was reduced in IR-HP-CT compared to LDCT (P < 0.0001). Image noise was reduced in the IR-HP-CT (16.8 ± 2.8 vs. 19.8 ± 3.4, P = 0.0001). SNR of lung and aorta of IR-HP-CT showed better results compared with that of LDCT (22.2 ± 5.9 vs. 33.0 ± 7.8, 1.9 ± 0.4 vs 1.1 ± 0.3, P < 0.0001). The score of cardiac pulsation artifacts were significantly reduced on IR-HP-CT (3.8 ± 0.4, 95% confidence interval, 3.7‒4.0) compared with LDCT (1.6 ± 0.6, 95% confidence interval, 1.3‒1.8) (P < 0.0001). SNR of muscle and fat, beam hardening artifact and overall subjective image quality of the mediastinum, lung and chest wall were comparable on both scans (P ≥ 0.05). CONCLUSION: IR-HP-CT with 120 kVp and 30 mAs tube setting in addition to an iterative reconstruction reduced cardiac motion artifact and radiation exposure while representing similar image quality compared with LDCT.
format Online
Article
Text
id pubmed-6345490
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-63454902019-02-02 High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers Lim, Hyun Kyung Ha, Hong Il Hwang, Hye Jeon Lee, Kwanseop PLoS One Research Article PURPOSE: Objective of this study was to evaluate the effectiveness of the iterative reconstruction of high-pitch dual-source chest CT (IR-HP-CT) scanned with low radiation exposure compared with low dose chest CT (LDCT). MATERIALS AND METHODS: This study was approved by the institutional review board. Thirty healthy adult volunteers (mean age 44 years) were enrolled in this study. All volunteers underwent both IR-HP-CT and LDCT. IR-HP-CT was scanned with 120 kVp tube voltage, 30 mAs tube current and pitch 3.2 and reconstructed with sinogram affirmed iterative reconstruction. LDCT was scanned with 120 kVp tube voltage, 40 mAs tube current and pitch 0.8 and reconstructed with B50 filtered back projection. Image noise, and signal to noise ratio (SNR) of the infraspinatus muscle, subcutaneous fat and lung parenchyma were calculated. Cardiac motion artifact, overall image quality and artifacts was rated by two blinded readers using 4-point scale. The dose-length product (DLP) (mGy∙cm) were obtained from each CT dosimetry table. Scan length was calculated from the DLP results. The DLP parameter was a metric of radiation output, not of patient dose. Size-specific dose estimation (SSDE, mGy) was calculated using the sum of the anteroposterior and lateral dimensions and effective radiation dose (ED, mSv) were calculated using CT dosimetry index. RESULTS: Approximately, mean 40% of SSDE (2.1 ± 0.2 mGy vs. 3.5 ± 0.3 mGy) and 34% of ED (1.0 ± 0.1 mSv vs. 1.5 ± 0.1 mSv) was reduced in IR-HP-CT compared to LDCT (P < 0.0001). Image noise was reduced in the IR-HP-CT (16.8 ± 2.8 vs. 19.8 ± 3.4, P = 0.0001). SNR of lung and aorta of IR-HP-CT showed better results compared with that of LDCT (22.2 ± 5.9 vs. 33.0 ± 7.8, 1.9 ± 0.4 vs 1.1 ± 0.3, P < 0.0001). The score of cardiac pulsation artifacts were significantly reduced on IR-HP-CT (3.8 ± 0.4, 95% confidence interval, 3.7‒4.0) compared with LDCT (1.6 ± 0.6, 95% confidence interval, 1.3‒1.8) (P < 0.0001). SNR of muscle and fat, beam hardening artifact and overall subjective image quality of the mediastinum, lung and chest wall were comparable on both scans (P ≥ 0.05). CONCLUSION: IR-HP-CT with 120 kVp and 30 mAs tube setting in addition to an iterative reconstruction reduced cardiac motion artifact and radiation exposure while representing similar image quality compared with LDCT. Public Library of Science 2019-01-24 /pmc/articles/PMC6345490/ /pubmed/30677082 http://dx.doi.org/10.1371/journal.pone.0211097 Text en © 2019 Lim et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Lim, Hyun Kyung
Ha, Hong Il
Hwang, Hye Jeon
Lee, Kwanseop
High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers
title High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers
title_full High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers
title_fullStr High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers
title_full_unstemmed High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers
title_short High-pitch, 120 kVp/30 mAs, low-dose dual-source chest CT with iterative reconstruction: Prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest CT in healthy adult volunteers
title_sort high-pitch, 120 kvp/30 mas, low-dose dual-source chest ct with iterative reconstruction: prospective evaluation of radiation dose reduction and image quality compared with those of standard-pitch low-dose chest ct in healthy adult volunteers
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6345490/
https://www.ncbi.nlm.nih.gov/pubmed/30677082
http://dx.doi.org/10.1371/journal.pone.0211097
work_keys_str_mv AT limhyunkyung highpitch120kvp30maslowdosedualsourcechestctwithiterativereconstructionprospectiveevaluationofradiationdosereductionandimagequalitycomparedwiththoseofstandardpitchlowdosechestctinhealthyadultvolunteers
AT hahongil highpitch120kvp30maslowdosedualsourcechestctwithiterativereconstructionprospectiveevaluationofradiationdosereductionandimagequalitycomparedwiththoseofstandardpitchlowdosechestctinhealthyadultvolunteers
AT hwanghyejeon highpitch120kvp30maslowdosedualsourcechestctwithiterativereconstructionprospectiveevaluationofradiationdosereductionandimagequalitycomparedwiththoseofstandardpitchlowdosechestctinhealthyadultvolunteers
AT leekwanseop highpitch120kvp30maslowdosedualsourcechestctwithiterativereconstructionprospectiveevaluationofradiationdosereductionandimagequalitycomparedwiththoseofstandardpitchlowdosechestctinhealthyadultvolunteers