Measuring the rate of manual transcription error in outpatient point-of-care testing
Many point-of-care laboratory tests are manually entered into the electronic health record by ambulatory clinic staff, but the rate of manual transcription error for this testing is poorly characterized. Using a dataset arising from a duplicated workflow that created a set of paired interfaced and m...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Oxford University Press
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6351970/ https://www.ncbi.nlm.nih.gov/pubmed/30649499 http://dx.doi.org/10.1093/jamia/ocy170 |
Sumario: | Many point-of-care laboratory tests are manually entered into the electronic health record by ambulatory clinic staff, but the rate of manual transcription error for this testing is poorly characterized. Using a dataset arising from a duplicated workflow that created a set of paired interfaced and manually entered point-of-care glucose measurements, we found that 260 of 6930 (3.7%) manual entries were discrepant from their interfaced result. Thirty-seven of the 260 (14.2%) errors were discrepant by more than 20% and included potentially dangerous mistranscriptions. An additional 37 (14.2%) errors were due to inclusion of non-numeric characters. Staff-entered result flags deviated from the result flag generated in the laboratory information system in 5121 of 6930 (73.9%) pairs. These data demonstrate that clinically significant discrepancies for clinic-entered point of care results occurred at a rate of approximately 5 per 1000 results and they underline the importance of interfacing instruments when feasible. |
---|