Evaluation of Acceptance Angle in Iodine-131 Single Photon Emission Computed Tomography Imaging with Monte Carlo Simulation

INTRODUCTION: In iodine-131 (I-131) imaging, the image quality is degraded by scatter and penetration in a collimator. In this work, we assessed the penetrated and the scattered photon fractions in the photopeak energy window using Monte Carlo Simulation code. MATERIALS AND METHODS: The Siemens Medi...

Descripción completa

Detalles Bibliográficos
Autores principales: Asmi, Hicham, Bentayeb, Farida, Bouzekraoui, Youssef, Bonutti, Faustino
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6352644/
https://www.ncbi.nlm.nih.gov/pubmed/30713374
http://dx.doi.org/10.4103/ijnm.IJNM_109_18
Descripción
Sumario:INTRODUCTION: In iodine-131 (I-131) imaging, the image quality is degraded by scatter and penetration in a collimator. In this work, we assessed the penetrated and the scattered photon fractions in the photopeak energy window using Monte Carlo Simulation code. MATERIALS AND METHODS: The Siemens Medical System equipped with high-energy collimator was simulated. We evaluated the acceptance angle values on geometric, penetration, and scatter components in a separate file. Binary images in a data file are obtained and each one of them was imported in software ImageJ. Full-width at half-maximum (FWHM) and sensitivity were calculated and compared. RESULTS: The simulation data show that for the acceptance angle value equal to 4.845°, the geometric, scatter, and penetration components were 93.20%, 4.13%, and 2.67%, respectively. Moreover, the resolution is improved (FWHM = 7.21 mm, full width at tenth maximum = 12.36 mm) for a point source at 12 cm from the detector. CONCLUSION: The small acceptance angle has a major impact on the image quality in I-131 single photon emission computed tomography.