Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics

In this paper, we proposed cancer cell acoustic stimulation by shunt-diode pre-linearizer scheme using a very high frequency (≥100 MHz) piezoelectric transducer. To verify the concept of our proposed scheme, we performed pulse-echo detection, and accessed therapeutic effects of human cervical cancer...

Descripción completa

Detalles Bibliográficos
Autores principales: Choi, Hojong, Choe, Se-woon
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6359139/
https://www.ncbi.nlm.nih.gov/pubmed/30654599
http://dx.doi.org/10.3390/s19020357
_version_ 1783392161523302400
author Choi, Hojong
Choe, Se-woon
author_facet Choi, Hojong
Choe, Se-woon
author_sort Choi, Hojong
collection PubMed
description In this paper, we proposed cancer cell acoustic stimulation by shunt-diode pre-linearizer scheme using a very high frequency (≥100 MHz) piezoelectric transducer. To verify the concept of our proposed scheme, we performed pulse-echo detection, and accessed therapeutic effects of human cervical cancer cells exposed to acoustic stimulation experiments using 100 MHz focused piezoelectric transducer triggered by PA with and without the proposed shunt-diode pre-linearizer scheme. In the pulse-echo detection responses, the peak-to-peak voltage of the echo signal when using the PA with shunt-diode pre-linearizer (49.79 mV) was higher than that when using the PA alone (29.87 mV). In the experimental results, the cell densities of cancer cells on Day 4 when using no acoustic stimulation (control group), the very high-frequency piezoelectric transducer triggered by PA only and PA combined with proposed pre-linearizer schemes (1 V and 5 V DC bias voltages) showed 100%, 92.8 ± 4.2%, 84.2 ± 4.6%, and 78 ± 2.9%, respectively. Therefore, we confirmed that the shunt-diode pre-linearizer could improve the performances of the pulse signals of the PA, thus, enabling better therapeutic stimulation performances for cancer cell suppression.
format Online
Article
Text
id pubmed-6359139
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-63591392019-02-06 Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics Choi, Hojong Choe, Se-woon Sensors (Basel) Article In this paper, we proposed cancer cell acoustic stimulation by shunt-diode pre-linearizer scheme using a very high frequency (≥100 MHz) piezoelectric transducer. To verify the concept of our proposed scheme, we performed pulse-echo detection, and accessed therapeutic effects of human cervical cancer cells exposed to acoustic stimulation experiments using 100 MHz focused piezoelectric transducer triggered by PA with and without the proposed shunt-diode pre-linearizer scheme. In the pulse-echo detection responses, the peak-to-peak voltage of the echo signal when using the PA with shunt-diode pre-linearizer (49.79 mV) was higher than that when using the PA alone (29.87 mV). In the experimental results, the cell densities of cancer cells on Day 4 when using no acoustic stimulation (control group), the very high-frequency piezoelectric transducer triggered by PA only and PA combined with proposed pre-linearizer schemes (1 V and 5 V DC bias voltages) showed 100%, 92.8 ± 4.2%, 84.2 ± 4.6%, and 78 ± 2.9%, respectively. Therefore, we confirmed that the shunt-diode pre-linearizer could improve the performances of the pulse signals of the PA, thus, enabling better therapeutic stimulation performances for cancer cell suppression. MDPI 2019-01-16 /pmc/articles/PMC6359139/ /pubmed/30654599 http://dx.doi.org/10.3390/s19020357 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Choi, Hojong
Choe, Se-woon
Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics
title Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics
title_full Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics
title_fullStr Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics
title_full_unstemmed Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics
title_short Acoustic Stimulation by Shunt-Diode Pre-Linearizer Using Very High Frequency Piezoelectric Transducer for Cancer Therapeutics
title_sort acoustic stimulation by shunt-diode pre-linearizer using very high frequency piezoelectric transducer for cancer therapeutics
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6359139/
https://www.ncbi.nlm.nih.gov/pubmed/30654599
http://dx.doi.org/10.3390/s19020357
work_keys_str_mv AT choihojong acousticstimulationbyshuntdiodeprelinearizerusingveryhighfrequencypiezoelectrictransducerforcancertherapeutics
AT choesewoon acousticstimulationbyshuntdiodeprelinearizerusingveryhighfrequencypiezoelectrictransducerforcancertherapeutics