Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs
BACKGROUND: Despite an abundance of literature regarding construct strength for a myriad of anchors and anchor configurations in the shoulder, there remains a paucity of biomechanical studies detailing the efficacy of these implants for proximal hamstring repair. PURPOSE: To biomechanically evaluate...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6365995/ https://www.ncbi.nlm.nih.gov/pubmed/30783606 http://dx.doi.org/10.1177/2325967118824149 |
_version_ | 1783393519347433472 |
---|---|
author | Gerhardt, Michael B. Assenmacher, Benjamin S. Chahla, Jorge |
author_facet | Gerhardt, Michael B. Assenmacher, Benjamin S. Chahla, Jorge |
author_sort | Gerhardt, Michael B. |
collection | PubMed |
description | BACKGROUND: Despite an abundance of literature regarding construct strength for a myriad of anchors and anchor configurations in the shoulder, there remains a paucity of biomechanical studies detailing the efficacy of these implants for proximal hamstring repair. PURPOSE: To biomechanically evaluate the ultimate failure load and failure mechanism of knotless and knotted anchor configurations for hamstring repair. STUDY DESIGN: Controlled laboratory study. METHODS: A total of 17 cadaveric specimens divided into 3 groups composed of intact hamstring tendons as well as 2 different anchor configurations (all-knotted and all-knotless) underwent first cyclic loading and subsequent maximal loading to failure. This protocol entailed a 10-N preload, followed by 100 cycles incrementally applied from 20 to 200 N at a frequency of 0.5 Hz, and ultimately followed by a load to failure with a loading rate of 33 mm/s. The ultimate failure load and mechanism of failure were recorded for each specimen, as was the maximal displacement of each bone-tendon interface subsequent to maximal loading. Analysis of variance was employed to calculate differences in the maximal load to failure as well as the maximal displacement between the 3 study groups. Holm-Sidak post hoc analysis was applied when necessary. RESULTS: The all-knotless suture anchor construct failed at the highest maximal load of the 3 groups (767.18 ± 93.50 N), including that for the intact tendon group (750.58 ± 172.22 N). There was no statistically significant difference between the all-knotless and intact tendon groups; however, there was a statistically significant difference in load to failure when the all-knotless construct was compared with the all-knotted technique (549.56 ± 20.74 N) (P = .024). The most common mode of failure in both repair groups was at the suture-tendon interface, whereas the intact tendon group most frequently failed via avulsion of the tendon from its insertion site. CONCLUSION: Under biomechanical laboratory testing conditions, proximal hamstring repair using all-knotless suture anchors outperformed the all-knotted suture anchor configuration with regard to elongation during cyclic loading and maximal load to failure. Failure in the all-knotted repair group was at the suture-tendon interface in most cases, whereas the all-knotless construct failed most frequently at the musculotendinous junction. CLINICAL RELEVANCE: No biomechanical studies have clearly identified the optimal anchor configuration to avert proximal hamstring repair failure. Delineating this ideal suture anchor construct and its strength compared with an intact hamstring tendon may alter the current standards for postoperative rehabilitation, which remain extremely conservative and onerous for these patients. |
format | Online Article Text |
id | pubmed-6365995 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-63659952019-02-19 Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs Gerhardt, Michael B. Assenmacher, Benjamin S. Chahla, Jorge Orthop J Sports Med Article BACKGROUND: Despite an abundance of literature regarding construct strength for a myriad of anchors and anchor configurations in the shoulder, there remains a paucity of biomechanical studies detailing the efficacy of these implants for proximal hamstring repair. PURPOSE: To biomechanically evaluate the ultimate failure load and failure mechanism of knotless and knotted anchor configurations for hamstring repair. STUDY DESIGN: Controlled laboratory study. METHODS: A total of 17 cadaveric specimens divided into 3 groups composed of intact hamstring tendons as well as 2 different anchor configurations (all-knotted and all-knotless) underwent first cyclic loading and subsequent maximal loading to failure. This protocol entailed a 10-N preload, followed by 100 cycles incrementally applied from 20 to 200 N at a frequency of 0.5 Hz, and ultimately followed by a load to failure with a loading rate of 33 mm/s. The ultimate failure load and mechanism of failure were recorded for each specimen, as was the maximal displacement of each bone-tendon interface subsequent to maximal loading. Analysis of variance was employed to calculate differences in the maximal load to failure as well as the maximal displacement between the 3 study groups. Holm-Sidak post hoc analysis was applied when necessary. RESULTS: The all-knotless suture anchor construct failed at the highest maximal load of the 3 groups (767.18 ± 93.50 N), including that for the intact tendon group (750.58 ± 172.22 N). There was no statistically significant difference between the all-knotless and intact tendon groups; however, there was a statistically significant difference in load to failure when the all-knotless construct was compared with the all-knotted technique (549.56 ± 20.74 N) (P = .024). The most common mode of failure in both repair groups was at the suture-tendon interface, whereas the intact tendon group most frequently failed via avulsion of the tendon from its insertion site. CONCLUSION: Under biomechanical laboratory testing conditions, proximal hamstring repair using all-knotless suture anchors outperformed the all-knotted suture anchor configuration with regard to elongation during cyclic loading and maximal load to failure. Failure in the all-knotted repair group was at the suture-tendon interface in most cases, whereas the all-knotless construct failed most frequently at the musculotendinous junction. CLINICAL RELEVANCE: No biomechanical studies have clearly identified the optimal anchor configuration to avert proximal hamstring repair failure. Delineating this ideal suture anchor construct and its strength compared with an intact hamstring tendon may alter the current standards for postoperative rehabilitation, which remain extremely conservative and onerous for these patients. SAGE Publications 2019-02-05 /pmc/articles/PMC6365995/ /pubmed/30783606 http://dx.doi.org/10.1177/2325967118824149 Text en © The Author(s) 2019 http://creativecommons.org/licenses/by-nc-nd/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 License (http://www.creativecommons.org/licenses/by-nc-nd/4.0/) which permits non-commercial use, reproduction and distribution of the work as published without adaptation or alteration, without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Article Gerhardt, Michael B. Assenmacher, Benjamin S. Chahla, Jorge Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs |
title | Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs |
title_full | Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs |
title_fullStr | Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs |
title_full_unstemmed | Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs |
title_short | Proximal Hamstring Repair: A Biomechanical Analysis of Variable Suture Anchor Constructs |
title_sort | proximal hamstring repair: a biomechanical analysis of variable suture anchor constructs |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6365995/ https://www.ncbi.nlm.nih.gov/pubmed/30783606 http://dx.doi.org/10.1177/2325967118824149 |
work_keys_str_mv | AT gerhardtmichaelb proximalhamstringrepairabiomechanicalanalysisofvariablesutureanchorconstructs AT assenmacherbenjamins proximalhamstringrepairabiomechanicalanalysisofvariablesutureanchorconstructs AT chahlajorge proximalhamstringrepairabiomechanicalanalysisofvariablesutureanchorconstructs |