Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy

INTRODUCTION: Variability among manufacturers of urine dipsticks, respective to their accuracy and measurement range, may lead to diagnostic errors and thus create a serious risk for the patient. Our aims were to determine the level of agreement between 12 most commonly used urine dipsticks in Croat...

Descripción completa

Detalles Bibliográficos
Autores principales: Vuljanić, Dora, Dojder, Ana, Špoljarić, Valentina, Saračević, Andrea, Dukić, Lora, Leniček-Krleža, Jasna, Vlašić-Tanasković, Jelena, Maradin, Ivana, Grzunov, Ana, Vogrinc, Željka, Šimundić, Ana-Maria
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Croatian Society of Medical Biochemistry and Laboratory Medicine 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6366948/
https://www.ncbi.nlm.nih.gov/pubmed/30799977
http://dx.doi.org/10.11613/BM.2019.010708
_version_ 1783393696039829504
author Vuljanić, Dora
Dojder, Ana
Špoljarić, Valentina
Saračević, Andrea
Dukić, Lora
Leniček-Krleža, Jasna
Vlašić-Tanasković, Jelena
Maradin, Ivana
Grzunov, Ana
Vogrinc, Željka
Šimundić, Ana-Maria
author_facet Vuljanić, Dora
Dojder, Ana
Špoljarić, Valentina
Saračević, Andrea
Dukić, Lora
Leniček-Krleža, Jasna
Vlašić-Tanasković, Jelena
Maradin, Ivana
Grzunov, Ana
Vogrinc, Željka
Šimundić, Ana-Maria
author_sort Vuljanić, Dora
collection PubMed
description INTRODUCTION: Variability among manufacturers of urine dipsticks, respective to their accuracy and measurement range, may lead to diagnostic errors and thus create a serious risk for the patient. Our aims were to determine the level of agreement between 12 most commonly used urine dipsticks in Croatia, examine their accuracy for glucose and total protein and to test their repeatability. MATERIALS AND METHODS: A total of 75 urine samples were used to examine comparability and accuracy of 12 dipstick brands (Combur 10 TestM, ChoiceLine 10, Combur 10 TestUX, ComboStik 10M, ComboStik 11M, CombiScreen 11SYS, CombiScreen 10SL, Combina 13, Combina 11S, Combina 10M, UriGnost 11, Multistix 10SG). Agreement between each dipstick and the reference (Combur 10 TestM) was expressed as kappa coefficient (acceptable κ ≥ 0.80). Accuracy for glucose and total protein was tested by comparison with quantitative measurements on analysers: AU400 (Beckman Coulter, USA), Cobas 6000 c501 (Roche Diagnostics, Germany) and Architect plus c4000 (Abbott, USA). Repeatability was assessed on 20 replicates (acceptable > 90%). RESULTS: Best agreement was achieved for glucose, total protein and nitrite (11/11, k > 0.80) and the lowest for bilirubin (5/5, k < 0.60). Sensitivities for total protein were 41-75% (AU400) and 56-92% (Cobas and Architect); while specificities were 41-75% (AU400, Cobas, Architect). Dipsticks’ sensitivity and specificity for glucose were 68-98%. Most of the dipsticks showed unacceptable repeatability (6/12, < 90%) for one parameter, most prominently for pH (3/12, < 90%). CONCLUSIONS: Most commonly used dipsticks in Croatia showed low level of agreement between each other. Moreover, their repeatability varies among manufacturers and their accuracy for glucose and proteins is poor.
format Online
Article
Text
id pubmed-6366948
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Croatian Society of Medical Biochemistry and Laboratory Medicine
record_format MEDLINE/PubMed
spelling pubmed-63669482019-02-22 Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy Vuljanić, Dora Dojder, Ana Špoljarić, Valentina Saračević, Andrea Dukić, Lora Leniček-Krleža, Jasna Vlašić-Tanasković, Jelena Maradin, Ivana Grzunov, Ana Vogrinc, Željka Šimundić, Ana-Maria Biochem Med (Zagreb) Original Articles INTRODUCTION: Variability among manufacturers of urine dipsticks, respective to their accuracy and measurement range, may lead to diagnostic errors and thus create a serious risk for the patient. Our aims were to determine the level of agreement between 12 most commonly used urine dipsticks in Croatia, examine their accuracy for glucose and total protein and to test their repeatability. MATERIALS AND METHODS: A total of 75 urine samples were used to examine comparability and accuracy of 12 dipstick brands (Combur 10 TestM, ChoiceLine 10, Combur 10 TestUX, ComboStik 10M, ComboStik 11M, CombiScreen 11SYS, CombiScreen 10SL, Combina 13, Combina 11S, Combina 10M, UriGnost 11, Multistix 10SG). Agreement between each dipstick and the reference (Combur 10 TestM) was expressed as kappa coefficient (acceptable κ ≥ 0.80). Accuracy for glucose and total protein was tested by comparison with quantitative measurements on analysers: AU400 (Beckman Coulter, USA), Cobas 6000 c501 (Roche Diagnostics, Germany) and Architect plus c4000 (Abbott, USA). Repeatability was assessed on 20 replicates (acceptable > 90%). RESULTS: Best agreement was achieved for glucose, total protein and nitrite (11/11, k > 0.80) and the lowest for bilirubin (5/5, k < 0.60). Sensitivities for total protein were 41-75% (AU400) and 56-92% (Cobas and Architect); while specificities were 41-75% (AU400, Cobas, Architect). Dipsticks’ sensitivity and specificity for glucose were 68-98%. Most of the dipsticks showed unacceptable repeatability (6/12, < 90%) for one parameter, most prominently for pH (3/12, < 90%). CONCLUSIONS: Most commonly used dipsticks in Croatia showed low level of agreement between each other. Moreover, their repeatability varies among manufacturers and their accuracy for glucose and proteins is poor. Croatian Society of Medical Biochemistry and Laboratory Medicine 2019-02-15 2019-02-15 /pmc/articles/PMC6366948/ /pubmed/30799977 http://dx.doi.org/10.11613/BM.2019.010708 Text en ©Croatian Society of Medical Biochemistry and Laboratory Medicine. This is an Open Access article distributed under the terms of the Creative Commons Attribution (http://creativecommons.org/licenses/by/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Articles
Vuljanić, Dora
Dojder, Ana
Špoljarić, Valentina
Saračević, Andrea
Dukić, Lora
Leniček-Krleža, Jasna
Vlašić-Tanasković, Jelena
Maradin, Ivana
Grzunov, Ana
Vogrinc, Željka
Šimundić, Ana-Maria
Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
title Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
title_full Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
title_fullStr Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
title_full_unstemmed Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
title_short Analytical verification of 12 most commonly used urine dipsticks in Croatia: comparability, repeatability and accuracy
title_sort analytical verification of 12 most commonly used urine dipsticks in croatia: comparability, repeatability and accuracy
topic Original Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6366948/
https://www.ncbi.nlm.nih.gov/pubmed/30799977
http://dx.doi.org/10.11613/BM.2019.010708
work_keys_str_mv AT vuljanicdora analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT dojderana analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT spoljaricvalentina analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT saracevicandrea analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT dukiclora analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT lenicekkrlezajasna analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT vlasictanaskovicjelena analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT maradinivana analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT grzunovana analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT vogrinczeljka analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy
AT simundicanamaria analyticalverificationof12mostcommonlyusedurinedipsticksincroatiacomparabilityrepeatabilityandaccuracy