Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
OBJECTIVE: The purpose of this study was to assess whether a photoplethysmography (PPG) sensor in a smart watch can accurately recognise the return of spontaneous circulation (ROSC) in cardiac arrest patients compared with carotid artery palpation. METHODS: This prospective observational study was c...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BMJ Publishing Group
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6367979/ https://www.ncbi.nlm.nih.gov/pubmed/30782884 http://dx.doi.org/10.1136/bmjopen-2018-023627 |
_version_ | 1783393911338696704 |
---|---|
author | Lee, Yoonje Shin, Hyungoo Choi, Hyuk Joong Kim, Changsun |
author_facet | Lee, Yoonje Shin, Hyungoo Choi, Hyuk Joong Kim, Changsun |
author_sort | Lee, Yoonje |
collection | PubMed |
description | OBJECTIVE: The purpose of this study was to assess whether a photoplethysmography (PPG) sensor in a smart watch can accurately recognise the return of spontaneous circulation (ROSC) in cardiac arrest patients compared with carotid artery palpation. METHODS: This prospective observational study was conducted on 50 out-of-hospital cardiac arrest patients who visited the emergency department (ED) of one tertiary hospital. As soon as the patient arrived at the ED, advanced cardiac life support was carried out immediately. At this time, three smart watches were attached to the carotid artery, forehead and wrist and were checked for pulse measurements every 2 min. In the case of ROSC, blood pressure, heart rate and heart rate regularity were confirmed, and pulse was simultaneously measured at three sites with smart watches. In the case of no ROSC, only the pulse was measured at three sites with the smart watches. RESULTS: There were 33 males (66%) and the mean age was 68±11.57 years. In 14 patients (28%), spontaneous circulation was recovered through cardiopulmonary resuscitation, and all survived. The sensitivity and specificity of manual palpation were 78.6% and 90.4%, respectively. False-positive and false-negative rates were 9.6% and 21.4%, respectively. Smart watches at all three sites had the same or higher sensitivity than manual palpation. The sensitivity of the smart watch was the highest, at 100%, in the carotid region and the lowest, at 78.6%, in the wrist region. The specificity of the smart watch was the highest, at 100%, in the wrist region and the lowest, at 78.7%, in the carotid region. CONCLUSION: Compared with manual pulse check, the PPG sensor embedded in the smart watch showed the same sensitivity and a higher specificity for recognising ROSC when measured at the wrist. |
format | Online Article Text |
id | pubmed-6367979 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | BMJ Publishing Group |
record_format | MEDLINE/PubMed |
spelling | pubmed-63679792019-03-10 Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study Lee, Yoonje Shin, Hyungoo Choi, Hyuk Joong Kim, Changsun BMJ Open Emergency Medicine OBJECTIVE: The purpose of this study was to assess whether a photoplethysmography (PPG) sensor in a smart watch can accurately recognise the return of spontaneous circulation (ROSC) in cardiac arrest patients compared with carotid artery palpation. METHODS: This prospective observational study was conducted on 50 out-of-hospital cardiac arrest patients who visited the emergency department (ED) of one tertiary hospital. As soon as the patient arrived at the ED, advanced cardiac life support was carried out immediately. At this time, three smart watches were attached to the carotid artery, forehead and wrist and were checked for pulse measurements every 2 min. In the case of ROSC, blood pressure, heart rate and heart rate regularity were confirmed, and pulse was simultaneously measured at three sites with smart watches. In the case of no ROSC, only the pulse was measured at three sites with the smart watches. RESULTS: There were 33 males (66%) and the mean age was 68±11.57 years. In 14 patients (28%), spontaneous circulation was recovered through cardiopulmonary resuscitation, and all survived. The sensitivity and specificity of manual palpation were 78.6% and 90.4%, respectively. False-positive and false-negative rates were 9.6% and 21.4%, respectively. Smart watches at all three sites had the same or higher sensitivity than manual palpation. The sensitivity of the smart watch was the highest, at 100%, in the carotid region and the lowest, at 78.6%, in the wrist region. The specificity of the smart watch was the highest, at 100%, in the wrist region and the lowest, at 78.7%, in the carotid region. CONCLUSION: Compared with manual pulse check, the PPG sensor embedded in the smart watch showed the same sensitivity and a higher specificity for recognising ROSC when measured at the wrist. BMJ Publishing Group 2019-02-03 /pmc/articles/PMC6367979/ /pubmed/30782884 http://dx.doi.org/10.1136/bmjopen-2018-023627 Text en © Author(s) (or their employer(s)) 2019. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ. This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/. |
spellingShingle | Emergency Medicine Lee, Yoonje Shin, Hyungoo Choi, Hyuk Joong Kim, Changsun Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study |
title | Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study |
title_full | Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study |
title_fullStr | Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study |
title_full_unstemmed | Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study |
title_short | Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study |
title_sort | can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study |
topic | Emergency Medicine |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6367979/ https://www.ncbi.nlm.nih.gov/pubmed/30782884 http://dx.doi.org/10.1136/bmjopen-2018-023627 |
work_keys_str_mv | AT leeyoonje canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy AT shinhyungoo canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy AT choihyukjoong canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy AT kimchangsun canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy |