Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study

OBJECTIVE: The purpose of this study was to assess whether a photoplethysmography (PPG) sensor in a smart watch can accurately recognise the return of spontaneous circulation (ROSC) in cardiac arrest patients compared with carotid artery palpation. METHODS: This prospective observational study was c...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Yoonje, Shin, Hyungoo, Choi, Hyuk Joong, Kim, Changsun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6367979/
https://www.ncbi.nlm.nih.gov/pubmed/30782884
http://dx.doi.org/10.1136/bmjopen-2018-023627
_version_ 1783393911338696704
author Lee, Yoonje
Shin, Hyungoo
Choi, Hyuk Joong
Kim, Changsun
author_facet Lee, Yoonje
Shin, Hyungoo
Choi, Hyuk Joong
Kim, Changsun
author_sort Lee, Yoonje
collection PubMed
description OBJECTIVE: The purpose of this study was to assess whether a photoplethysmography (PPG) sensor in a smart watch can accurately recognise the return of spontaneous circulation (ROSC) in cardiac arrest patients compared with carotid artery palpation. METHODS: This prospective observational study was conducted on 50 out-of-hospital cardiac arrest patients who visited the emergency department (ED) of one tertiary hospital. As soon as the patient arrived at the ED, advanced cardiac life support was carried out immediately. At this time, three smart watches were attached to the carotid artery, forehead and wrist and were checked for pulse measurements every 2 min. In the case of ROSC, blood pressure, heart rate and heart rate regularity were confirmed, and pulse was simultaneously measured at three sites with smart watches. In the case of no ROSC, only the pulse was measured at three sites with the smart watches. RESULTS: There were 33 males (66%) and the mean age was 68±11.57 years. In 14 patients (28%), spontaneous circulation was recovered through cardiopulmonary resuscitation, and all survived. The sensitivity and specificity of manual palpation were 78.6% and 90.4%, respectively. False-positive and false-negative rates were 9.6% and 21.4%, respectively. Smart watches at all three sites had the same or higher sensitivity than manual palpation. The sensitivity of the smart watch was the highest, at 100%, in the carotid region and the lowest, at 78.6%, in the wrist region. The specificity of the smart watch was the highest, at 100%, in the wrist region and the lowest, at 78.7%, in the carotid region. CONCLUSION: Compared with manual pulse check, the PPG sensor embedded in the smart watch showed the same sensitivity and a higher specificity for recognising ROSC when measured at the wrist.
format Online
Article
Text
id pubmed-6367979
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-63679792019-03-10 Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study Lee, Yoonje Shin, Hyungoo Choi, Hyuk Joong Kim, Changsun BMJ Open Emergency Medicine OBJECTIVE: The purpose of this study was to assess whether a photoplethysmography (PPG) sensor in a smart watch can accurately recognise the return of spontaneous circulation (ROSC) in cardiac arrest patients compared with carotid artery palpation. METHODS: This prospective observational study was conducted on 50 out-of-hospital cardiac arrest patients who visited the emergency department (ED) of one tertiary hospital. As soon as the patient arrived at the ED, advanced cardiac life support was carried out immediately. At this time, three smart watches were attached to the carotid artery, forehead and wrist and were checked for pulse measurements every 2 min. In the case of ROSC, blood pressure, heart rate and heart rate regularity were confirmed, and pulse was simultaneously measured at three sites with smart watches. In the case of no ROSC, only the pulse was measured at three sites with the smart watches. RESULTS: There were 33 males (66%) and the mean age was 68±11.57 years. In 14 patients (28%), spontaneous circulation was recovered through cardiopulmonary resuscitation, and all survived. The sensitivity and specificity of manual palpation were 78.6% and 90.4%, respectively. False-positive and false-negative rates were 9.6% and 21.4%, respectively. Smart watches at all three sites had the same or higher sensitivity than manual palpation. The sensitivity of the smart watch was the highest, at 100%, in the carotid region and the lowest, at 78.6%, in the wrist region. The specificity of the smart watch was the highest, at 100%, in the wrist region and the lowest, at 78.7%, in the carotid region. CONCLUSION: Compared with manual pulse check, the PPG sensor embedded in the smart watch showed the same sensitivity and a higher specificity for recognising ROSC when measured at the wrist. BMJ Publishing Group 2019-02-03 /pmc/articles/PMC6367979/ /pubmed/30782884 http://dx.doi.org/10.1136/bmjopen-2018-023627 Text en © Author(s) (or their employer(s)) 2019. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ. This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.
spellingShingle Emergency Medicine
Lee, Yoonje
Shin, Hyungoo
Choi, Hyuk Joong
Kim, Changsun
Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
title Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
title_full Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
title_fullStr Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
title_full_unstemmed Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
title_short Can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
title_sort can pulse check by the photoplethysmography sensor on a smart watch replace carotid artery palpation during cardiopulmonary resuscitation in cardiac arrest patients? a prospective observational diagnostic accuracy study
topic Emergency Medicine
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6367979/
https://www.ncbi.nlm.nih.gov/pubmed/30782884
http://dx.doi.org/10.1136/bmjopen-2018-023627
work_keys_str_mv AT leeyoonje canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy
AT shinhyungoo canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy
AT choihyukjoong canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy
AT kimchangsun canpulsecheckbythephotoplethysmographysensoronasmartwatchreplacecarotidarterypalpationduringcardiopulmonaryresuscitationincardiacarrestpatientsaprospectiveobservationaldiagnosticaccuracystudy