Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT
PURPOSE: We have quantitatively evaluated the image quality of a new commercially available iterative cone-beam computed tomography reconstruction algorithm over standard cone-beam computed tomography image reconstruction results. METHODS: This iterative cone-beam computed tomography reconstruction...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6373994/ https://www.ncbi.nlm.nih.gov/pubmed/30803367 http://dx.doi.org/10.1177/1533033818823054 |
_version_ | 1783395087621816320 |
---|---|
author | Mao, Weihua Liu, Chang Gardner, Stephen J. Siddiqui, Farzan Snyder, Karen C. Kumarasiri, Akila Zhao, Bo Kim, Joshua Wen, Ning Winston Movsas, Benjamin Chetty, Indrin J. |
author_facet | Mao, Weihua Liu, Chang Gardner, Stephen J. Siddiqui, Farzan Snyder, Karen C. Kumarasiri, Akila Zhao, Bo Kim, Joshua Wen, Ning Winston Movsas, Benjamin Chetty, Indrin J. |
author_sort | Mao, Weihua |
collection | PubMed |
description | PURPOSE: We have quantitatively evaluated the image quality of a new commercially available iterative cone-beam computed tomography reconstruction algorithm over standard cone-beam computed tomography image reconstruction results. METHODS: This iterative cone-beam computed tomography reconstruction pipeline uses a finite element solver (AcurosCTS)-based scatter correction and a statistical (iterative) reconstruction in addition to a standard kernel-based correction followed by filtered back-projection-based Feldkamp-Davis-Kress cone-beam computed tomography reconstruction. Standard full-fan half-rotation Head, half-fan full-rotation Head, and standard Pelvis cone-beam computed tomography protocols have been investigated to scan a quality assurance phantom via the following image quality metrics: uniformity, HU constancy, spatial resolution, low contrast detection, noise level, and contrast-to-noise ratio. An anthropomorphic head phantom was scanned for verification of noise reduction. Clinical patient image data sets for 5 head/neck patients and 5 prostate patients were qualitatively evaluated. RESULTS: Quality assurance phantom study results showed that relative to filtered back-projection-based cone-beam computed tomography, noise was reduced from 28.8 ± 0.3 HU to a range between 18.3 ± 0.2 and 5.9 ± 0.2 HU for Full-Fan Head scans, from 14.4 ± 0.2 HU to a range between 12.8 ± 0.3 and 5.2 ± 0.3 HU for Half-Fan Head scans, and from 6.2 ± 0.1 HU to a range between 3.8 ± 0.1 and 2.0 ± 0.2 HU for Pelvis scans, with the iterative cone-beam computed tomography algorithm. Spatial resolution was marginally improved while results for uniformity and HU constancy were similar. For the head phantom study, noise was reduced from 43.6 HU to a range between 24.8 and 13.0 HU for a Full-Fan Head and from 35.1 HU to a range between 22.9 and 14.0 HU for a Half-Fan Head scan. The patient data study showed that artifacts due to photon starvation and streak artifacts were all reduced, and image noise in specified target regions were reduced to 62% ± 15% for 10 patients. CONCLUSION: Noise and contrast-to-noise ratio image quality characteristics were significantly improved using the iterative cone-beam computed tomography reconstruction algorithm relative to the filtered back-projection-based cone-beam computed tomography method. These improvements will enhance the accuracy of cone-beam computed tomography-based image-guided applications. |
format | Online Article Text |
id | pubmed-6373994 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-63739942019-02-20 Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT Mao, Weihua Liu, Chang Gardner, Stephen J. Siddiqui, Farzan Snyder, Karen C. Kumarasiri, Akila Zhao, Bo Kim, Joshua Wen, Ning Winston Movsas, Benjamin Chetty, Indrin J. Technol Cancer Res Treat Original Article PURPOSE: We have quantitatively evaluated the image quality of a new commercially available iterative cone-beam computed tomography reconstruction algorithm over standard cone-beam computed tomography image reconstruction results. METHODS: This iterative cone-beam computed tomography reconstruction pipeline uses a finite element solver (AcurosCTS)-based scatter correction and a statistical (iterative) reconstruction in addition to a standard kernel-based correction followed by filtered back-projection-based Feldkamp-Davis-Kress cone-beam computed tomography reconstruction. Standard full-fan half-rotation Head, half-fan full-rotation Head, and standard Pelvis cone-beam computed tomography protocols have been investigated to scan a quality assurance phantom via the following image quality metrics: uniformity, HU constancy, spatial resolution, low contrast detection, noise level, and contrast-to-noise ratio. An anthropomorphic head phantom was scanned for verification of noise reduction. Clinical patient image data sets for 5 head/neck patients and 5 prostate patients were qualitatively evaluated. RESULTS: Quality assurance phantom study results showed that relative to filtered back-projection-based cone-beam computed tomography, noise was reduced from 28.8 ± 0.3 HU to a range between 18.3 ± 0.2 and 5.9 ± 0.2 HU for Full-Fan Head scans, from 14.4 ± 0.2 HU to a range between 12.8 ± 0.3 and 5.2 ± 0.3 HU for Half-Fan Head scans, and from 6.2 ± 0.1 HU to a range between 3.8 ± 0.1 and 2.0 ± 0.2 HU for Pelvis scans, with the iterative cone-beam computed tomography algorithm. Spatial resolution was marginally improved while results for uniformity and HU constancy were similar. For the head phantom study, noise was reduced from 43.6 HU to a range between 24.8 and 13.0 HU for a Full-Fan Head and from 35.1 HU to a range between 22.9 and 14.0 HU for a Half-Fan Head scan. The patient data study showed that artifacts due to photon starvation and streak artifacts were all reduced, and image noise in specified target regions were reduced to 62% ± 15% for 10 patients. CONCLUSION: Noise and contrast-to-noise ratio image quality characteristics were significantly improved using the iterative cone-beam computed tomography reconstruction algorithm relative to the filtered back-projection-based cone-beam computed tomography method. These improvements will enhance the accuracy of cone-beam computed tomography-based image-guided applications. SAGE Publications 2019-01-08 /pmc/articles/PMC6373994/ /pubmed/30803367 http://dx.doi.org/10.1177/1533033818823054 Text en © The Author(s) 2019 http://creativecommons.org/licenses/by-nc/4.0/ This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Article Mao, Weihua Liu, Chang Gardner, Stephen J. Siddiqui, Farzan Snyder, Karen C. Kumarasiri, Akila Zhao, Bo Kim, Joshua Wen, Ning Winston Movsas, Benjamin Chetty, Indrin J. Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT |
title | Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT |
title_full | Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT |
title_fullStr | Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT |
title_full_unstemmed | Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT |
title_short | Evaluation and Clinical Application of a Commercially Available Iterative Reconstruction Algorithm for CBCT-Based IGRT |
title_sort | evaluation and clinical application of a commercially available iterative reconstruction algorithm for cbct-based igrt |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6373994/ https://www.ncbi.nlm.nih.gov/pubmed/30803367 http://dx.doi.org/10.1177/1533033818823054 |
work_keys_str_mv | AT maoweihua evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT liuchang evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT gardnerstephenj evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT siddiquifarzan evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT snyderkarenc evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT kumarasiriakila evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT zhaobo evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT kimjoshua evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT wenningwinston evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT movsasbenjamin evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt AT chettyindrinj evaluationandclinicalapplicationofacommerciallyavailableiterativereconstructionalgorithmforcbctbasedigrt |